510(k) Data Aggregation

    K Number
    K191425

    Validate with FDA (Live)

    Date Cleared
    2020-01-21

    (237 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DeRoyal HYDRO-TEMP® Neonatal Skin Temperature Probe Covers are used to adhere the thermistor of the temperature probe to the patient and insulate the thermistor from ambient room temperature ensuring correct temperature readings.

    Device Description

    The HYDRO-TEMP® Neonatal Skin Temperature Probe Cover is an accessory to be used with DeRoyal-branded NICU neonatal skin temperature probes used with an infant radiant warmer. They have not been evaluated for use with other manufacturer's products. It is an adhesive pad used to affix a temperature sensor to a neonate's skin to monitor the skin surface temperature. The cover also insulates the temperature sensor from ambient room temperature to ensure temperature readings are accurate.

    There are two configurations of the probe cover. Both of these configurations contain a release liner, a patient-contacting hydrogel adhesive, and a metallized polyester backing. One configuration contains an additional foam material that is placed in between the hydrogel adhesive and metallic backing. The device is non-sterile and can be packaged on its own in a foil pouch or in a resealable plastic bag with a temperature probe.

    AI/ML Overview

    The provided text describes the regulatory filing for the DeRoyal HYDRO-TEMP® Neonatal Skin Temperature Probe Cover (K191425) and its equivalence to a predicate device. The information is focused on mechanical and biocompatibility testing, rather than a clinical study evaluating the device's performance in terms of temperature reading accuracy.

    Therefore, many of the requested categories related to clinical study design, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication method, MRMC studies, and training set information, cannot be fully addressed from the provided document.

    Here's a summary of the available information structured to answer your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Test PerformedReported Device Performance (HYDRO-TEMP® Neonatal Skin Temperature Probe Cover)Predicate Device Performance (HYDRO-TEMP® NICU and Skin Surface Temperature Sensor (K925006))
    Average peel force per ASTM D903-984.42 N4.28 N
    Average force required to remove cover from surface (Internal Standard)10.1 N9.1 N
    BiocompatibilityISO 10993-1 compliant (Cytotoxicity, Sensitization, Irritation)ISO 10993-1 compliant
    Insulation TestingDid not result in a temperature difference from the predicate deviceNot explicitly stated, but implied as a baseline for comparison
    Accuracy Testing (IEC 60601-2-21)Demonstrated the tolerance range of temperature readings when the probe cover was usedNot explicitly stated for the predicate in this context

    Note: The document states "The proposed device is similar to the predicate but with a modified hydrogel material to improve the adhesive strength, which is supported by performance testing." This implies the acceptance criterion for adhesive strength was to be at least equivalent to, if not better than, the predicate. The data shows it is better.

    2. Sample size used for the test set and the data provenance

    • Sample size: Not specified for any of the individual tests (peel force, pull test, insulation, accuracy, biocompatibility).
    • Data provenance: The tests were conducted internally by DeRoyal Industries, Inc. (or contracted labs for biocompatibility) as part of their 510(k) submission. The data is retrospective for the purpose of demonstrating substantial equivalence to a legally marketed predicate device. Country of origin not explicitly stated, but DeRoyal is a U.S.-based company.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was not a clinical study involving human judgment or "ground truth" established by experts in the context of diagnoses or interpretations. The tests were objective mechanical and biological safety assessments.

    4. Adjudication method for the test set

    Not applicable. This was not a study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a passive accessory (probe cover) and does not involve AI or human "readers" making interpretations.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical accessory, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance tests would be the standardized testing methods themselves (e.g., ASTM D903-98 for peel strength, IEC 60601-2-21 for accuracy, ISO 10993 for biocompatibility) and the physical measurements obtained from those tests. There is no expert consensus, pathology, or outcomes data used as "ground truth" in the way it might be for a diagnostic device.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.