510(k) Data Aggregation

    K Number
    K223251

    Validate with FDA (Live)

    Device Name
    Hexanium PLIF
    Date Cleared
    2022-12-16

    (56 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hexanium® PLIF (Posterior Lumbar Interbody Fusion) system is an intervertebral body fusion device indicated for use with autogenous bone graft in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two continuous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received at least 6 months of non-operative treatment prior to treatment with Hexanium PLIF system. This device has to be filled with autogenous bone graft material. This device is implanted via the posterior approach. Hexanium PLIF system must be used in combination with supplemental internal spinal fixation which has been cleared by the FDA for use in the lumbar spine.

    Device Description

    The Hexanium® PLIF (Posterior Lumbar Interbody Fusion) is a titanium alloy (Ti6Al4V ELI) interbody cage manufactured via an Additive Manufacturing method. The honeycomb structure allows for bone through-growth through the structure of the cage as well as providing lateral and vertical bone graft windows in the body of the cage. Hexanium® PLIF cages are available in heights of 7-16 mm (in 1 mm increments), width of 9 mm, lengths of 22 and 25 mm, and lordosis angles of 0°, 5°, and 8°. Hexanium® PLIF cages are provided sterile and supplied with a set of non-sterile surgical instruments.

    AI/ML Overview

    The Hexanium® PLIF (Posterior Lumbar Interbody Fusion) system is an intervertebral body fusion device. The provided text lacks the detailed "acceptance criteria" and subsequent "reported device performance" typically found in a clinical study report for AI-powered devices. The document is a 510(k) summary for a medical device (an intervertebral body fusion device, not an AI device) where equivalence to a predicate device is established through mechanical testing, not clinical performance data. Therefore, the questions related to AI device performance metrics, such as sensitivity, specificity, or human-in-the-loop performance, and associated study details are not applicable here.

    However, I can extract information related to the device's performance data that was provided to demonstrate substantial equivalence:

    1. A table of acceptance criteria and the reported device performance:

    The document states that the Hexanium® PLIF cages conform to the FDA guidance "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" dated June 12, 2007. The device was evaluated against predetermined functional and performance requirements and external standard requirements.

    Test TypeAcceptance Criteria (Implied by adherence to standards)Reported Device Performance (Implied by "meets the same... requirements")
    Mechanical TestingConformity to ASTM F2077, ASTM F2267, and draft ASTM F-04.25.02.02 for intervertebral body fusion devices.Hexanium® PLIF cages met these pre-determined functional and performance requirements.
    Static CompressionMeets force/displacement, stiffness, and permanent deformation requirements as per ASTM F2077.Performed according to ASTM F2077, implying requirements were met.
    Static Compression-ShearMeets force/displacement, stiffness, and permanent deformation requirements as per ASTM F2077.Performed according to ASTM F2077, implying requirements were met.
    Static TorsionMeets torque/angle, stiffness, and permanent deformation requirements as per ASTM F2077.Performed according to ASTM F2077, implying requirements were met.
    Dynamic CompressionWithstands specified number of cyclic loads at defined forces without fracture or excessive deformation as per ASTM F2077.Performed according to ASTM F2077, implying requirements were met.
    Dynamic Compression-ShearWithstands specified number of cyclic loads at defined forces without fracture or excessive deformation as per ASTM F2077.Performed according to ASTM F2077, implying requirements were met.
    Dynamic TorsionWithstands specified number of cyclic loads at defined torques without fracture or excessive deformation as per ASTM F2077.Performed according to ASTM F2077, implying requirements were met.
    Subsidence TestingMeets subsidence resistance requirements as per ASTM F2267.Performed according to ASTM F2267, implying requirements were met.
    Expulsion TestingMeets expulsion resistance requirements as per draft ASTM F-04.25.02.02.Performed according to draft ASTM F-04.25.02.02, implying requirements were met.
    BiocompatibilityMeets pyrogenicity limits.Bacterial Endotoxins Test performed in accordance with USP, implying pyrogen limit specifications were met.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the document for mechanical testing. These tests typically involve a specific number of device samples (e.g., typically N=6 or more) for each test configuration, as defined by the ASTM standards, but the exact number is not provided.
    • Data Provenance: The tests are laboratory-based mechanical and biocompatibility tests, not human data. The document does not specify a country of origin for the testing itself, but the submitter is based in France.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This question is not applicable. The device is not an AI device requiring expert-established ground truth for a test set. The performance is assessed against engineering standards and material properties.

    4. Adjudication method for the test set:

    This question is not applicable. There was no clinical adjudication for an AI device. Mechanical and biocompatibility tests are assessed against predefined pass/fail criteria from international standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This question is not applicable. No MRMC study was conducted as this is not an AI device. The document explicitly states: "No clinical data has been presented."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This question is not applicable. This is not an AI device.

    7. The type of ground truth used:

    For mechanical testing, the "ground truth" is defined by the objective pass/fail criteria established within the referenced ASTM international standards for intervertebral body fusion devices (e.g., F2077, F2267, F-04.25.02.02). For biocompatibility, the ground truth is the pyrogen limit specification as per USP.

    8. The sample size for the training set:

    This question is not applicable. There is no training set as this is not an AI/machine learning device. The device's design and manufacturing process are based on established engineering principles for intervertebral body fusion devices.

    9. How the ground truth for the training set was established:

    This question is not applicable. There is no training set for an AI device. The device's performance is demonstrated through in-vitro mechanical and biocompatibility testing against predefined standards.