    Intended Use

    The Hercules® Syndesmosis Implant System is intended to provide fixation during the healing process following a syndesmotic trauma such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures and as an adjunct in external and intramedullary fixation systems. The Hercules® Syndesmosis Implant System is also intended to provide fixation during healing process after joint reconstruction in the midfoot and forefoot including correction of hallux valgus deformity.

    Device Description

    The In2Bones Hercules® Syndesmosis Implant System is a sterile, single-use, device intended to stabilize syndesmotic trauma of the ankle. The subject system consists of UHMWPE suture tensioned between two low profile titanium alloy buttons, designed to complement plate fixation or to allow use as a stand-alone device. This device is also packaged with various ancillary instruments to aid in insertion.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Hercules® Syndesmosis Implant System." This document is a regulatory submission to the FDA demonstrating substantial equivalence to a predicate device, rather than a study proving the device meets clinical acceptance criteria.

    Therefore, the information required to answer the prompt regarding acceptance criteria, device performance, sample sizes for test and training sets, expert qualifications, ground truth establishment, and MRMC studies, is not present in the provided text.

    The document primarily focuses on:

    • Regulatory Clearance: Announcing the FDA's determination of substantial equivalence for the Hercules® Syndesmosis Implant System (K220260) to previously cleared predicate devices.
    • Device Description: A sterile, single-use system composed of UHMWPE suture and titanium alloy buttons, intended to stabilize syndesmotic trauma of the ankle.
    • Indications for Use: Providing fixation during healing after syndesmotic trauma (e.g., Weber B and C ankle fractures) and as an adjunct in other fixation systems, as well as for fixation after joint reconstruction in the midfoot and forefoot.
    • Basis for Substantial Equivalence: Similarities in indications, materials, and geometry to predicate devices (e.g., Arthrex Tightrope, Wright Medical Gravity Syndesmosis LP, Arthrosurface KISSLoc Suture System).
    • Performance Testing (Bench Testing): Mentions "Cyclic displacement and load to failure bench testing (mechanical testing)" was performed and "The Hercules® Syndesmosis Implant System was validated per ISO 10993-1 for biocompatibility, ISO 11137-2 for gamma sterilization, ISO 11135 for ETO sterilization and ISO 11607-1 for packaging. Endotoxin testing was completed per ANSI/AAMI ST72." This refers to engineering/mechanical/biocompatibility testing, not clinical performance data or studies involving human subjects/readers.

    In summary, there is no information in this document about studies that prove the device meets acceptance criteria in a clinical or AI-assisted context, as the questions imply. The document pertains to the regulatory clearance process for a medical implant based on substantial equivalence, relying on mechanical and material testing, not diagnostic performance or human reader studies.