510(k) Data Aggregation

    K Number
    K250956

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-26

    (87 days)

    Product Code
    Regulation Number
    866.1620
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HardyDisk AST disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacterales, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Candida spp., Haemophilus spp., Neisseria gonorrhoeae, Neisseria meningitidis and Streptococcus spp., including Streptococcus pneumoniae.

    Use of HardyDisk AST Gepotidacin 10μg (GEP10) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Staphylococcus saprophyticus, and Enterococcus faecalis to gepotidacin, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

    HardyDisk AST Gepotidacin at concentration 10μg demonstrated acceptable performance to determine the zone diameter (mm) of gepotidacin against the following microorganisms:

    Enterobacterales (Citrobacter freundii complex, Escherichia coli, Klebsiella pneumoniae, Citrobacter koseri, Klebsiella aerogenes, Klebsiella oxytoca/Raoultella ornithinolytica, Morganella morganii, Proteus mirabilis, Providencia rettgeri)
    Enterococcus faecalis
    Staphylococcus saprophyticus

    Device Description

    Not Found

    AI/ML Overview

    The provided FDA 510(k) clearance letter is for the HardyDisk AST Gepotidacin 10µg (GEP10), an Antimicrobial Susceptibility Test Disc. This document primarily focuses on the regulatory clearance of a medical device and its intended use, rather than a detailed study report of an AI/ML powered device.

    Therefore, the information traditionally sought for acceptance criteria and study proving performance for AI/ML devices (such as sample sizes for test and training sets, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment) is not present in this type of FDA clearance letter for a non-AI/ML antimicrobial susceptibility test disk.

    The provided document describes:

    • Device Name: HardyDisk AST Gepotidacin 10µg (GEP10)
    • Intended Use: Semi-quantitative in vitro susceptibility testing by the agar diffusion (Kirby-Bauer) method.
    • Acceptable Performance Demonstrated For: Determining the zone diameter (mm) of gepotidacin against specific microorganisms (Enterobacterales, Staphylococcus saprophyticus, and Enterococcus faecalis).

    Since this is a traditional in vitro diagnostic device (an antimicrobial susceptibility test disk), its "acceptance criteria" relate to its ability to accurately determine susceptibility based on zone diameters, which is typically validated through comparison to a reference method (e.g., broth microdilution) and established interpretive criteria (STIC). The performance metrics would involve agreement rates (Essential Agreement and Categorical Agreement) with the reference method.

    Based on the provided text, I cannot extract the specific details requested regarding study methodology, sample sizes, expert involvement, and ground truth establishment in the context of an AI/ML device. The letter only states that the device "demonstrated acceptable performance."

    To answer your request, I will indicate where the information would typically be found in a study report for an AI/ML device and then explain why it's not present in this specific document.


    Hypothetical Description of Acceptance Criteria and Study (if this were an AI/ML device, as requested by the prompt, but adapted to the limited information for this non-AI/ML device):

    Given that the provided document is a 510(k) clearance letter for an Antimicrobial Susceptibility Test Disc (a traditional IVD, not an AI/ML device), the detailed study information typically associated with AI/ML algorithm validation (e.g., number of experts, adjudication methods, MRMC studies) is not included in this type of regulatory correspondence.

    However, for a non-AI/ML device like an AST disk, "acceptance criteria" and "proof of meeting criteria" would typically involve demonstrating:

    1. Essential Agreement (EA): The percentage of isolates for which the test device's zone diameter measurement is within a certain agreement range (e.g., ±3 mm) of the reference method's MIC or zone diameter.
    2. Categorical Agreement (CA): The percentage of isolates for which the test device's interpretive result (Susceptible, Intermediate, Resistant) matches the reference method's interpretive result based on established breakpoints.
    3. Very Major Errors (VME): False Susceptible (device says susceptible, reference says resistant).
    4. Major Errors (ME): False Resistant (device says resistant, reference says susceptible).
    5. Minor Errors (mE): Device says intermediate, but reference says susceptible or resistant, or vice versa.

    The "study that proves the device meets the acceptance criteria" for an AST disk typically involves:

    • Testing a large number of clinical isolates representing the target organisms (e.g., Enterobacterales, Staphylococcus saprophyticus, Enterococcus faecalis).
    • Comparing the zone diameters and interpretive results of the HardyDisk AST Gepotidacin 10µg (GEP10) against an FDA-recognized reference method (e.g., broth microdilution or agar dilution) for gepotidacin.
    • Analyzing the EA, CA, VME, ME, and mE rates according to FDA guidance for AST devices.

    Since this is NOT an AI/ML device, many of the requested fields (e.g., MRMC studies, number of experts for ground truth beyond standard AST expert review) are not applicable.


    Addressing Your Specific Questions based on the Provided Document:

    Here's how the provided information aligns (or doesn't align) with your specific questions:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria (Generalized for AST): For AST devices, typical acceptance criteria involve high rates of Essential Agreement and Categorical Agreement (e.g., >90%), and low rates of Major Errors and Very Major Errors (e.g., <3% for VME, <1.5% for ME).
      • Reported Device Performance: The document only states: "HardyDisk AST Gepotidacin at concentration 10μg demonstrated acceptable performance to determine the zone diameter (mm) of gepotidacin against the following microorganisms: Enterobacterales (...), Enterococcus faecalis, Staphylococcus saprophyticus." No specific performance metrics (e.g., actual EA, CA, VME, ME percentages) are provided in this clearance letter.
    2. Sample sizes used for the test set and the data provenance:

      • Sample Size: Not specified in the provided text. For AST devices, typically hundreds to thousands of isolates are tested.
      • Data Provenance: Not specified. For AST devices, data typically comes from clinical laboratories, often across multiple sites and potentially involving retrospective and prospective collections of isolates, or challenge sets from culture collections.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the AI/ML sense. For an AST device, the "ground truth" is established by a reference method (e.g., FDA-recognized broth microdilution or agar dilution standard). The interpretation of these reference method results against established Susceptibility Test Interpretive Criteria (STIC) provides the "true" susceptibility. While experts (e.g., clinical microbiologists, laboratory technicians) perform and interpret these reference tests, it's not the "expert consensus" model seen in AI/ML imaging studies.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication is typically for resolving discrepancies among human readers or ground truth establishment in complex interpretations. For AST, the reference method provides a definitive result.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is an AST disk, not an AI/ML assistant for human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Applicable in a different context. The "standalone performance" of this device would be its ability to produce accurate zone diameters and susceptibility interpretations independently when used by a trained laboratory professional according to its validated procedure (i.e., its performance in relation to the reference standard). The letter states it demonstrated "acceptable performance."
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Reference Method: The ground truth for AST devices is established by an FDA-recognized reference antimicrobial susceptibility testing method (e.g., broth microdilution, agar dilution) with results interpreted using FDA-recognized Susceptibility Test Interpretive Criteria (STIC) for gepotidacin.
    8. The sample size for the training set:

      • Not applicable. This is a physical test disk, not a machine learning model that undergoes training in the AI/ML sense. Its "performance" is inherent to its manufacturing and chemical properties and is validated, not "trained."
    9. How the ground truth for the training set was established:

      • Not applicable. (See point 8).

    In summary, the provided FDA clearance letter is for a traditional diagnostic device and therefore does not contain the detailed study methodology information typically associated with AI/ML device clearances. The "acceptable performance" mentioned refers to the device meeting the established regulatory requirements for AST disk clearance, which involves comparison to a gold standard reference method.