    K Number
    K971126

    Validate with FDA (Live)

    Date Cleared
    1997-06-24

    (89 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Innovalive Technologies' Hydrocolloid & Intelligent Hydrocolloid Wound Dressings may be used for the management of wounds including.

    • Partial full thickness wounds, ie artenal, venous, diabetic ulocrs .
    • Post-operative surgical wounds
    • Donor Eiles
    • Trauma Wounds
    • Dermal Lesions
    • Profective dressing
    Device Description

    Innovative Technologies' Hydrocolloid Wound Dressings, are highly conformable, sterile, primary wound dressings-intended to provide an environment ideally suited for chronic wound management. The dressings are soft conformable hydrocolloid preparations for heavy to light exuding partial-full thickness wounds. The hydrocolloid preparations react with wound exudate to form an gelatinous mass providing for a moist healing environment. The dressing may easily be lifted away from the wound, reducing the potential for delicate peri-wound tissue damage during dressing changes. Dressings are supplied sterile in single use blister packs.

    AI/ML Overview

    The provided text is related to a 510(k) submission for "Innovative Technologies' Hydrocolloid & Intelligent Hydrocolloid Wound Dressings." This document is a regulatory submission for medical devices, specifically seeking clearance to market a device based on its substantial equivalence to a legally marketed predicate device.

    Key Observation:
    The document does not describe a study that proves the device meets specific acceptance criteria. Instead, it outlines the regulatory review process for a 510(k) submission. In a 510(k) submission, the primary goal is to demonstrate "substantial equivalence" to a predicate device, meaning the new device is as safe and effective as the predicate. This is typically achieved by comparing the new device's technological characteristics, indications for use, and performance data (if available and relevant for equivalence) to the predicate device. It does not generally involve a new clinical study with pre-defined acceptance criteria in the same way a PMA (Premarket Approval) submission would for novel devices.

    Therefore, I cannot extract the requested information as it is not present in the provided text. The document focuses on:

    • Device Description: Hydrocolloid and Intelligent Hydrocolloid Wound Dressings.
    • Intended Use: Management of various wounds (dermal ulcers, superficial wounds, burns, donor sites, post-operative wounds, protective dressings).
    • Comparison to a Predicate Device: Duoderm CGF / Extra Thin Hydrocolloid Dressings manufactured by Convatec, highlighting similarities in composition, surface, indications for use, transparency, self-adhesiveness, and sterilization method.
    • Safety Assertions: Biocompatibility testing completed.
    • Regulatory Clearance: A letter from the FDA determining substantial equivalence, along with limitations on labeling claims.

    To directly answer your request based on the provided text, while acknowledging the absence of a detailed study report:


    Acceptance Criteria and Device Performance Study (Not Applicable as a standalone study report is not provided)

    The provided document describes a 510(k) submission, which focuses on demonstrating "substantial equivalence" to a legally marketed predicate device rather than presenting a standalone study with defined acceptance criteria and performance metrics. The implicit "acceptance criteria" for a 510(k) submission is that the device is as safe and effective as the predicate device.

    1. A table of acceptance criteria and the reported device performance

    Since no specific acceptance criteria for a new study are outlined, and no new performance data comparing against such criteria is presented, this table cannot be created from the text. The document does, however, provide a comparative table between the applicant's device and the predicate device, showing their similar characteristics:

    CharacteristicsInnovative TechnologiesConvatec (Predicate Device)
    CompositionHydrocolloid & polyurethane film / Hydrocolloid, polyurethane foam & polyurethane filmHydrocolloid & polyurethane film / Hydrocolloid, polyurethane foam & polyurethane film
    SurfaceExtruded and laminatedExtruded and laminated
    Indications For UseDermal ulcers, superficial wounds, burns (1st & 2nd degree), donor sites, post-operative wounds, protective dressings, partial-full thickness wounds, trauma wounds, dermal lesionsDermal ulcers, superficial wounds, burns (1st & 2nd degree), donor sites, post-operative wounds, protective dressings
    TransparentYesYes
    Self adhesiveYesYes
    PackagingBlister PackPouch
    Sterilization MethodGamma Irradiation (in accordance with ANSI/AAMI/ISO11137-1995 and AAMI TIR8-1991 for dosimetric release with a sterility assurance level of 10^-6^)Gamma Irradiation
    BiocompatibilitySuccessfully completed per ISO/Tripartite guidelines (cytotoxicity, acute systemic toxicity, skin irritation, sensitization)(Implicitly assumed to be met given its status as a legally marketed predicate; not explicitly detailed for the predicate in this document)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The document does not describe a clinical "test set" in the context of a new study designed to meet specific performance criteria. The evaluation is based on a comparison to a predicate device and existing regulatory standards (e.g., for sterilization and biocompatibility).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No "ground truth" establishment for a test set is described as part of a clinical performance study. The expertise cited is related to quality assurance and regulatory affairs (Priscilla Whitehead, Director of QA/RA for Innovative Technologies).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No test set requiring expert adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a wound dressing, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to "human readers improving with AI" is not relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a wound dressing, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For this 510(k) submission, the "ground truth" for demonstrating safety and effectiveness relies on the established regulatory approval and safety profile of the legally marketed predicate device (Duoderm CGF / Extra Thin Hydrocolloid Dressings by Convatec), combined with the Innovative Technologies' device meeting relevant standards for sterilization and biocompatibility.

    8. The sample size for the training set

    Not applicable. There is no mention of a training set as this is not a machine learning or diagnostic algorithm device.

    9. How the ground truth for the training set was established

    Not applicable. There is no mention of a training set.