Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K192264

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-08

    (79 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

    Device Description

    EMS, Electrical Muscle Stimulator, which improves, tones, firms & strengthens muscle and relaxes stiff muscle through the skin. It is recognized as a clinically proven, effective, non-medication method of training muscle from certain causes. It manages muscle strengthen, toning and firming. It is also free from side effects when used properly, and can also be used as a simple means of self-training.

    HIVOX Spopad EMS SP series, SP-911 and SP-921 are the proposed devices for the 510(k) submission. These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in therapy or for the treatment of any medical conditions or diseases.

    SP-911 and SP-921 are 1-channel battery-operated-user-friendly muscle stimulation system specifically designed to exercise the muscles. Each device comprises namely an electronic stimulator module which generates the required stimulation signals. SP-911 comprises 2 electrodes, which connects the signals from the stimulator to the skin. SP-921 comprises 4 electrodes, which connect the signals from the stimulator to the skin. Power is supplied from one battery, CR2032, located in a compartment protected by a removable battery cover. The user cannot access the wiring or connectors.

    AI/ML Overview

    The provided text is a 510(k) Summary for the HIVOX Spopad EMS SP-911 and SP-921. It details the safety and performance testing conducted to demonstrate substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria through a clinical study for a new, complex algorithm or AI.

    The document explicitly states: "No clinical test data was used to support the decision of substantial equivalence." (Section 5.9). This means there was no clinical study conducted where the device's performance was measured against established acceptance criteria using human data, adjudicated ground truth, or expert readers.

    Instead, the submission relies on non-clinical testing to demonstrate performance and safety, specifically for a "Powered Muscle Stimulator." The acceptance criteria referenced are for electrical and physical performance characteristics of the device itself, rather than diagnostic accuracy or clinical effectiveness in the context of AI.

    Therefore, many of the requested details regarding sample size, data provenance, expert qualifications, ground truth, MRMC studies, or standalone algorithm performance are not applicable to this type of device clearance and the information provided.

    I can, however, extract the acceptance criteria for the non-clinical tests that were performed, as implied by the statement "All the test result demonstrate HIVOX Spopad EMS SP-911 and SP-921 meet the requirement of its pre-defined acceptance criteria and intended use". The document also implicitly defines acceptance criteria by noting when differences do not adversely impact safety and effectiveness, and by comparing the subject device's performance against the predicate and relevant standards.

    Here's a breakdown of the relevant information provided:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present a clean "acceptance criteria" table with "reported performance" directly. Instead, it compares the subject device's performance characteristics to those of its predicate device, and implicitly (though not explicitly numerically) against relevant international standards. The determination of "Identical" implies meeting the same performance characteristics as the cleared predicate, which serves as the de facto acceptance benchmark for substantial equivalence in many parameters. For parameters that are "Different but does not adversely impact safety and effectiveness," the standard or a justified value serves as the acceptance criteria.

    Feature / Acceptance Criteria (Implied)Subject Device Performance (SP-911 / SP-921)Predicate Device Performance (SP-910 / SP-920)Substantial Equivalence Determination / Justification of Acceptance
    General
    Intended UseIdentical to PredicateIndicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.Identical.
    Technology/Principle of OperationElectrical Muscle StimulationElectrical Muscle StimulationIdentical.
    Compliance with Voluntary StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10Identical (SP-911/921 also comply with IEC 60601-1-11, which extends compliance).
    Compliance with 21 CFR 898YesYesIdentical.
    Housing material and constructionSiliconeSiliconeIdentical.
    Electrical / Output Characteristics
    Output waveformSymmetrical biphasicSymmetrical biphasicIdentical.
    ShapeRectangularRectangularIdentical.
    Duration of primary (depolarizing) phase00Identical.
    Pulse duration (μs)400400Identical.
    Max Output Voltage (V, ±10%) at 500 ΩSP-911: 52 / SP-921: 58.4SP-910: 52 / SP-920: 58.4Identical.
    Max Output Voltage (V, ±10%) at 2k ΩSP-911: 102 / SP-921: 106SP-910: 102 / SP-920: 106Identical.
    Max Output Voltage (V, ±10%) at 10k ΩSP-911: 150 / SP-921: 154SP-910: 150 / SP-920: 154Identical.
    Max Output Current (mA, ±10%) at 500 ΩSP-911: 104 / SP-921: 117SP-910: 104 / SP-920: 117Identical.
    Max Output Current (mA, ±10%) at 2k ΩSP-911: 51 / SP-921: 53SP-910: 51 / SP-920: 53Identical.
    Max Output Current (mA, ±10%) at 10k ΩSP-911: 15 / SP-921: 15.4SP-910: 15 / SP-920: 15.4Identical.
    Frequency (Hz)SP-911: 3/4/5 / SP-921: 2/4/25SP-910: 3/4/5 / SP-920: 2/4/25Identical.
    Net charge per pulse at 500 Ω (μC)SP-911: 0.416 / SP-921: 0.468SP-910: 0.416 / SP-920: 0.468Identical.
    Maximum charge at 500 Ω (μC)SP-911: 41.6 / SP-921: 46.8SP-910: 41.6 / SP-920: 46.8Identical.
    Maximum current density at 500 Ω (mA/cm²)SP-911: 1.672 / SP-921: 1.057SP-910: 1.187 / SP-920: 1.057SP-911: Different (higher) but does not adversely impact safety and effectiveness. Justified against 0.25 W/cm² acceptance criterion for power density. SP-921: Identical.
    Maximum average power density at 500 Ω (W/cm²)SP-911: 0.0869 / SP-921: 0.0617SP-910: 0.0617 / SP-920: 0.0617SP-911: Different (higher) but does not adversely impact safety and effectiveness. Still below the acceptance criterion of 0.25 W/cm² from the Class II Special Control Guidance Document. SP-921: Identical.
    Safety Features
    Shelf-Life testingMet requirementsN/A (implied by predicate clearance)Met pre-defined acceptance criteria.
    Performance verificationMet requirementsN/A (implied by predicate clearance)Met pre-defined acceptance criteria.
    Usability validationMet requirementsN/A (implied by predicate clearance)Met pre-defined acceptance criteria.
    Method of Line Current IsolationBattery supplyBattery supplyIdentical.
    Patient Leakage Current Normal Condition (μΑ)2.02.0Identical.
    Patient Leakage Current Single Fault Condition (μΑ)2.12.1Identical.
    Method of channel isolationSingle channelSingle channelIdentical.
    Average DC current through electrodes when device is on but no pulses are being applied (μΑ)00Identical.
    Number of output modes11Identical.
    Regulated current or regulated voltage?VoltageVoltageIdentical.
    Software / Firmware / Microprocessor control?YesYesIdentical.
    Automatic overload trip?NoNoIdentical.
    Automatic no-load trip?NoNoIdentical.
    Automatic shut-off?YesYesIdentical.
    User overrides control?YesYesIdentical.
    Indicator display – On / Off StatusNoNoIdentical.
    Indicator display – Low battery?NoNoIdentical.
    Indicator display – Voltage / Current levelNoNoIdentical.
    Timer range (minutes)2020Identical.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable. The testing was non-clinical (shelf-life, performance verification, usability validation) and involves device samples, not patient data or clinical subjects. The document does not specify the number of devices tested for these non-clinical assessments.
    • Data Provenance: The testing was conducted by HIVOX Biotek Inc., based in New Taipei City, Taiwan, R.O.C. (as per "Submitter Address" in Section 5.3). The data is from the manufacturer's internal testing as part of their 510(k) submission. It is not retrospective or prospective clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. As this was a non-clinical submission, there was no "ground truth" derived from expert consensus on medical images or clinical outcomes. The "ground truth" for the device's electrical and physical parameters would be defined by engineering specifications and international standards, measured using appropriate test equipment and procedures.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable for the reasons stated above.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a powered muscle stimulator, not an AI-enabled diagnostic or assistive tool. No MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a hardware device, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests, the "ground truth" implicitly refers to engineering specifications, regulatory requirements (e.g., 21 CFR 898), and international standards (e.g., IEC 60601 series). Performance verification would involve measuring parameters against these defined acceptable ranges. Usability validation typically involves testing by representative users against predefined usability criteria.

    8. The sample size for the training set:

    • Not applicable. This device does not involve machine learning or AI that requires a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable for the reasons stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141921

    Validate with FDA (Live)

    Device Name
    HIVOX SPOPAD EMS
    Manufacturer
    Date Cleared
    2015-01-16

    (184 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

    Device Description

    These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

    SP-910 / SP-920 / SP-620 are 1-channel battery-operated user-friendly muscle stimulation systems specifically designed to exercise the muscles. Each device comprises namely an electronic stimulator module which generates the required stimulation signals. SP-910/620 comprises 2 electrodes, which connects the signals from the stimulator to the skin. SP-920 comprises 4 electrodes, which connects the signals from the stimulator to the skin. Power is supplied from one battery, CR2032, located in a compartment protected by a removable battery cover. The user cannot access the wiring or connectors.

    AI/ML Overview

    The provided document is a 510(k) summary for the HIVOX Spopad EMS SP-910, SP-920, and SP-620, which are powered muscle stimulators. This type of document focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a standalone study with defined acceptance criteria and performance results in the way a clinical trial might.

    Therefore, the information requested about acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment (as typically found in AI/ML device evaluations) is not present in this 510(k) summary. These devices are much simpler than AI-powered diagnostic tools and rely on established electrical safety and performance standards for their regulatory clearance.

    However, I can extract the relevant comparison information that serves as the basis for the declaration of substantial equivalence, which implicitly acts as the "criteria" for approval in this context.

    Here's a summary based on the provided text, aligning with your request where possible, and noting when information is unavailable:

    Acceptance Criteria and Device Performance (Implicit from Predicate Comparison)

    For a 510(k) submission, the "acceptance criteria" are generally met by demonstrating that the new device is as safe and effective as a legally marketed predicate device. The following table summarizes the comparison made by the manufacturer. The reported device performance is presented in comparison to the predicate, with the conclusion that differences are minor and do not raise new safety or effectiveness concerns.

    Feature/ParameterAcceptance Criteria (Predicate Device K092476, Model 4M)Reported Performance (HIVOX Spopad EMS SP-910, SP-920, SP-620)
    Intended UseImprovement of muscle tone/firmness, strengthening muscles in arms, abdomen, thighs, buttocks; Not for therapy or treatment of medical conditions.Improvement of muscle tone/firmness, strengthening muscles in arms, abdomen, thighs, buttocks; Not for therapy or treatment of medical conditions. (Same)
    TechnologyElectrical Muscle StimulationElectrical Muscle Stimulation (Same)
    Power Source1.5V battery * 33V Battery * 1
    Method of Line Current IsolationBattery supplyBattery Supply (Same)
    Patient Leakage Current (Normal Condition)< 3 µA2.0 µA (SP-920)
    Patient Leakage Current (Single Fault Condition)< 4 µA2.1 µA (SP-920)
    Method of channel isolationSoftware1 channel
    Average DC current through electrodes (device on, no pulses)0 µA0 µA (Same)
    Number of output modes11 (Same)
    Regulated current or voltage?VoltageVoltage (Same)
    Software/firmware/Microprocessor control?YesYes (Same)
    Automatic overload trip?NoNo (Same)
    Automatic no-load trip?NoNo (Same)
    Automatic shut-off?YesYes (Same)
    User overrides control?YesYes (Same)
    Indicator display - On/Off StatusYesNo
    Indicator display – Low battery?YesNo
    Indicator display – Voltage/CurrentNoNo (Same)
    Timer Range (minutes)N/A20
    Compliance with voluntary standards?IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 60601-1-4IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2
    Compliance with 21 CFR 898?YesYes (Same)
    Housing material and constructionABSSilicone
    Output waveformMonophasicSymmetrical biphasic
    ShapeRectangularRectangular (Same)
    Duration of primary (depolarizing) phase00 (Same)
    Pulse duration (µSec)N/A400
    Maximum output voltage (V, +/-10%) at 500 ohmsN/A52, 58.4, 60 (for SP-910, SP-920, SP-620 respectively)
    Maximum output voltage (V, +/-10%) at 2k ohmsN/A102, 106, 109
    Maximum output voltage (V, +/-10%) at 10k ohmsN/A150, 146, 140
    Maximum output current (mA, +/-10%) at 500 ohmsN/A104, 117, 120
    Maximum output current (mA, +/-10%) at 2k ohmsN/A51, 53, 54.5
    Maximum output current (mA, +/-10%) at 10k ohmsN/A15, 14.6, 14
    Frequency (Hz)N/ASP-910: 3/4/5; SP-920/SP-620: 2/4/25
    Net charge per pulse at 500 ohms (µC)N/A0.416, 0.468, 0.960
    Maximum charge at 500 ohms (µC)N/A41.6, 46.8, 48
    Maximum current density at 500 ohms (mA/cm²)< 21.328, 1.057, 1.952
    Maximum average power density at 500 ohms (W/cm²)< 0.250.0691, 0.0617, 0.117
    Burst mode (Pulse per burst)N/A25 (for SP-620)
    Burst mode (Burst per second)N/A1 (for SP-620)
    Burst mode (Burst duration (sec))N/A20 (for SP-620)
    Burst mode (Duty cycle)N/A20 (for SP-620)

    Detailed information based on your numbered points:

    1. A table of acceptance criteria and the reported device performance: See the table above. The "acceptance criteria" are implicitly defined by the characteristics of the predicate device and relevant industry standards. The reported performance of the HIVOX devices is compared against these implicit criteria.

    2. Sample sizes used for the test set and the data provenance:

      • Not Applicable. This 510(k) summary does not describe a clinical study or a test set in the conventional sense for evaluating a diagnostic or AI device. The comparison is based on the technical specifications of the device and its predicate, and compliance with performance and safety standards.
      • The document implies that the devices meet IEC electrical safety standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2) which involve specific tests, but the sample sizes for these standard compliance tests are not specified in this summary.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. No ground truth in the context of expert review for medical imaging or AI performance is mentioned. This device is a muscle stimulator, not an AI diagnostic tool.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable. No such adjudication method is mentioned or relevant for this type of device submission.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This is not an AI-assisted device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This is not an algorithm-only or AI device. The "standalone" performance refers to the device's electrical output characteristics and its compliance with safety standards, which are evaluated without human intervention in the loop of its function (though a human operates the device).
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not Applicable. For a powered muscle stimulator, "ground truth" is typically defined by adherence to engineering specifications and electrical safety standards (e.g., measured current, voltage, frequency must be within specified ranges and below safety limits). The document states that the devices (both predicate and subject) "pass medical device electric safety standard, IEC 60601-1 and standard for Nerve and Muscle Stimulator, IEC 60601-2-10 and electromagnetic compatibility standard IEC 60601-1-2." This compliance with standards serves as the "ground truth" for safety and basic performance.
    8. The sample size for the training set:

      • Not Applicable. This product is a physical electrical muscle stimulator, not an AI/ML product that undergoes training.
    9. How the ground truth for the training set was established:

      • Not Applicable. As there is no training set for an AI/ML model, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1