Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200144

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-08

    (77 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hi-Torque Steelcore™ Bare Guide Wire is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. It is not in the cerebral vasculature.

    Device Description

    The HT Steelcore Bare Guide Wire is classified under the product code DOX, with the regulation number 21 CFR 870.1330, Catheter guide wire. The HT Steelcore Bare Guide Wire is a steerable wire offered in only one configuration and is designed to be used with devices compatible with 0.018" guide wires.

    The HT Steelcore Bare Guide Wire is developed with 300 cm length and has a maximum diameter of 0.019". The distal tip has a radiopaque length of 5.0 cm. The HT Steelcore Bare Guide Wire will be available with the features as listed below:

    • Proximal Coating Single Coat Silicone Based hydrophobic coating. ●
    • Distal Coating Dual Coat Silicone Based Hydrophobic Coating. .
    AI/ML Overview

    This document describes the premarket notification (510(k)) for the HI-TORQUE Steelcore™ Bare Guide Wire. The information provided outlines the device's characteristics, its intended use, and the testing conducted to demonstrate its substantial equivalence to a legally marketed predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "performance testing passed with results meeting acceptance criteria." However, it does not provide a specific table detailing the quantitative acceptance criteria for each test and the corresponding quantitative reported device performance. It generally asserts that the device passed all tests and met acceptance criteria, but lacks specific numerical metrics for these criteria or the test results.

    The acceptance criteria are implicitly defined by the chosen performance tests and the relevant FDA guidance and standards. The reported "performance" is the qualitative statement that the device "passed" these tests.

    Performance Test CategoryAcceptance Criteria (Implicit)Reported Device Performance (Qualitative)
    BiocompatibilityMeets specified biocompatibility standards (e.g., cytotoxicity, sensitization, irritation) for patient contact.Passed
    SterilizationAchieves a Sterility Assurance Level (SAL) of 10⁻⁶ (as indicated for the predicate device, likely applies here through equivalence).Passed
    Pyrogen ValidationMeets pyrogenicity limits.Passed
    Shelf life and PackagingMaintains integrity, sterility, and performance throughout its asserted shelf life (2 years) and under specified packaging conditions.Passed
    Non-clinical bench testingMeets functional and physical specifications (e.g., diameter, length, radiopacity, coating performance, various mechanical properties relevant to guide wires). The text mentions adherence to FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling: 2019."Passed

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the performance testing (biocompatibility, sterilization, pyrogen validation, shelf life/packaging, and non-clinical bench testing).

    Regarding data provenance:

    • Country of origin: Not explicitly stated, but given the submission is to the FDA, it is implied to be relevant to the US regulatory framework.
    • Retrospective or Prospective: Not applicable in the context of bench testing. These are controlled laboratory tests, not clinical studies involving patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the "test set" here refers to physical performance characteristics of a medical device (a guide wire), not a dataset for an AI algorithm or a clinical study requiring expert consensus for ground truth. The "ground truth" is established by the specifications, standards, and validated test methods themselves.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies or AI algorithm validation where multiple human readers assess data and discrepancies need to be resolved. For bench testing of a physical device, the "adjudication" is inherent in the adherence to validated test protocols and the objective measurement of physical properties.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of diagnostic tools or AI systems, often involving human readers' interpretations of medical images. This submission is for a physical medical device (a guide wire) and relies on bench testing and substantial equivalence to a predicate device, not clinical effectiveness in a diagnostic or interpretive context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a "standalone" study in the context of an algorithm or AI was not done. This submission is for a physical medical device, not an AI algorithm.

    7. The Type of Ground Truth Used

    For the performance testing, the "ground truth" or reference standard for evaluation was based on:

    • Established engineering specifications and design requirements for guide wires.
    • Relevant FDA guidance documents: Specifically, "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling: 2019."
    • Pertinent consensus standards (though not explicitly listed, implied by "as recommended by the FDA guidance document as applicable").
    • Comparison to the predicate device: The device is considered substantially equivalent if its performance demonstrates it is "as safe and as effective" as the predicate.

    This is not ground truth derived from expert consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set

    Not Applicable. This is not an AI/ML device where a "training set" would be used to develop an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As above, there is no AI/ML training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1