Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K242444

    Validate with FDA (Live)

    Date Cleared
    2024-11-27

    (103 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan+™ Mode, Combined modes, Multi-Image modes (Dual, Quad), 3D/4D modes.

    Device Description

    The HERA W9/ HERA W10 are general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as Bmode, M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S- Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes.

    The HERA W9/HERA W10 also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W9/HERA W10 have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    The provided FDA 510(k) summary (K242444) describes the acceptance criteria and study proving the device meets these criteria for three AI-powered features: HeartAssist, BiometryAssist, and ViewAssist, as well as a non-AI feature, SonoSync.

    Here's a breakdown of the requested information for each AI feature:

    HeartAssist

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    View recognition accuracy89%96.07%
    Segmentation dice-score0.80.88
    Size measurement (area) error rate8% or less8% or less
    Size measurement (angle) error rate4% or less4% or less
    Size measurement (circumference) error rate11% or less11% or less
    Size measurement (diameter) error rate11% or less11% or less

    2. Sample size used for the test set and data provenance

    • Individuals: 69
    • Static Images: 315 (at least 1 static image per view location per individual)
    • Provenance: Data collected at two hospitals in the United States and South Korea. Mixed retrospective and prospective data collection.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    • Number of Experts: Three active participating experts for initial classification and manual drawing, supervised by one additional expert.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (primary classification/drawing).
      • Two sonographers with more than 10 years of experience in fetal cardiology (primary classification/drawing).
      • One obstetrician with more than 25 years of experience (supervising the entire process).

    4. Adjudication method for the test set

    Not explicitly stated. The process mentions "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the images." It doesn't specify if there was a consensus mechanism or independent review and adjudication if the experts disagreed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the provided data describes the standalone performance of the HeartAssist algorithm against established ground truth.

    7. The type of ground truth used

    Expert consensus based on manual classification of views and manual drawing of corresponding anatomy areas.

    8. The sample size for the training set

    Not explicitly stated for HeartAssist, but it is mentioned that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three."

    9. How the ground truth for the training set was established

    Not explicitly detailed for the training set, but it is implied to be the same method as for the validation set: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the images."


    BiometryAssist

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Segmentation dice-score0.80.91
    Size measurement (circumference) error rate8% or less8% or less
    Size measurement (distance) error rate4% or less4% or less
    Size measurement (NT, NB, IT) error rate1mm or less1mm or less

    2. Sample size used for the test set and data provenance

    • Individuals: 33
    • Static Images: 360 (at least 1 static image per view location per individual)
    • Provenance: Data collected at two hospitals in South Korea and the United States. Mixed retrospective and prospective data collection.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    • Number of Experts: Three active participating experts for initial classification and manual drawing, supervised by one additional expert.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (primary classification/drawing).
      • Two sonographers with more than 10 years of experience in fetal cardiology (primary classification/drawing).
      • One obstetrician with more than 25 years of experience (supervising the entire process).

    4. Adjudication method for the test set

    Not explicitly stated. The process mentions "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image." It doesn't specify if there was a consensus mechanism or independent review and adjudication if the experts disagreed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the provided data describes the standalone performance of the BiometryAssist algorithm against established ground truth.

    7. The type of ground truth used

    Expert consensus based on manual classification of views and manual drawing of corresponding anatomy areas.

    8. The sample size for the training set

    Not explicitly stated for BiometryAssist, but it is mentioned that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three."

    9. How the ground truth for the training set was established

    Not explicitly detailed for the training set, but it is implied to be the same method as for the validation set: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image."


    ViewAssist

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    View recognition accuracy89%94.92%
    Anatomy annotation (segmentation) dice-score0.80.89

    2. Sample size used for the test set and data provenance

    • Individuals: 98
    • Static Images: 1,485 (at least 1 static image per view location per individual)
    • Provenance: Data collected at two hospitals in South Korea and the United States. Mixed retrospective and prospective data collection.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    • Number of Experts: Three active participating experts for initial classification and manual drawing, supervised by one additional expert.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (primary classification/drawing).
      • Two sonographers with more than 10 years of experience in fetal cardiology (primary classification/drawing).
      • One obstetrician with more than 25 years of experience (supervising the entire process).

    4. Adjudication method for the test set

    Not explicitly stated. The process mentions "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image." It doesn't specify if there was a consensus mechanism or independent review and adjudication if the experts disagreed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the provided data describes the standalone performance of the ViewAssist algorithm against established ground truth.

    7. The type of ground truth used

    Expert consensus based on manual classification of views and manual drawing of corresponding anatomy areas.

    8. The sample size for the training set

    Not explicitly stated for ViewAssist, but it is mentioned that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three."

    9. How the ground truth for the training set was established

    Not explicitly detailed for the training set, but it is implied to be the same method as for the validation set: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image."


    General Notes applicable to all AI features:

    • No Multi-Reader Multi-Case (MRMC) Study: The document explicitly states that "The subject of this premarket submission, HERA W9/ HERA W10, did not require clinical studies to demonstrate the substantial equivalence." The studies described are technical performance evaluations of the AI algorithms, not comparative effectiveness studies with human readers.
    • Ground Truth Consistency: For all three AI features, the ground truth establishment process is described identically, relying on a small panel of experienced experts.
    • Independence of Data: For all three AI features, it is stated that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three."
    • Demographics: For all three AI features, the demographic distribution of the validation dataset indicates Female Gender, Reproductive age (specific age not collected), and Ethnicity/Country as Not Available / United States and South Korea. ICUOG and AIUM guidelines were used to divide fetal ultrasound images into views. BMI and Gestational Age distributions are also provided.
    • Equipment: Data was acquired with SAMSUNG MEDISON's ultrasound systems (HERA W9/HERA W10) to secure diversity.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230084

    Validate with FDA (Live)

    Date Cleared
    2023-04-21

    (100 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan+TM Mode, Combined modes, Multi-Image modes (Dual, Quad), 3D/4D modes.

    Device Description

    The HERA W9/ HERA W10 are general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as Bmode, M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S- Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes.

    The HERA W9/HERA W10 also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W9/HERA W10 have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the BiometryAssist and ViewAssist features of the HERA W9/HERA W10 Diagnostic Ultrasound System, based on the provided document:


    BiometryAssist

    1. Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Performance
    Segmentation TestAverage Dice-score ≥ 0.80.91 (threshold 0.8)
    Size MeasurementCircumference error rate ≤ 8%8% or less
    Distance measured value error rate ≤ 4%4% or less
    NT measured value error rate ≤ 1mm1mm or less

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 320 fetal biometry images.
    • Data Provenance: Images collected retrospectively and prospectively at two hospitals in South Korea and the United States.

    3. Number of Experts and Qualifications

    • Experts for Test Set Ground Truth: Three participating experts and one supervising obstetrician.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (supervisor).
      • Two sonographers with more than 10 years of experience, all specializing in fetal cardiology.
      • One obstetrician with more than 25 years of experience (supervisor).

    4. Adjudication Method

    • The document implies a consensus-based approach among the three experts for classifying images and manually drawing anatomy areas, supervised by a fourth obstetrician. No specific "2+1" or "3+1" adjudication method is explicitly stated, but the process involves multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not explicitly mentioned or described for comparing human readers with and without AI assistance. The study focuses on the standalone performance of the AI algorithm.

    6. Standalone Performance

    • Yes, the study describes the standalone performance of the algorithm for segmentation and size measurement without human intervention after the initial image acquisition and ground truthing.

    7. Type of Ground Truth Used

    • Expert consensus. All acquired images for training, tuning, and validation were classified into correct views by experts, and corresponding anatomy areas were manually drawn by them.

    8. Training Set Sample Size

    • Not explicitly stated, but the document mentions that a dataset was acquired for training, tuning, and validation.

    9. How Ground Truth for Training Set Was Established

    • All acquired images for training, tuning, and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each image. The participating experts were composed of an obstetrician with more than 20 years of experience and two sonographers with more than 10 years of experience, all in fetal cardiology, and the entire process was supervised by another obstetrician with more than 25 years of experience.

    ViewAssist

    1. Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Performance
    View Recognition TestAverage recognition accuracy ≥ 89%94.70% (threshold 89%)
    Anatomy Annotation (Segmentation) TestAverage Dice-score ≥ 0.80.875 (threshold 0.8)

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 1,320 fetal heart and fetal biometry images.
    • Data Provenance: Images collected retrospectively and prospectively at two hospitals in South Korea and the United States.

    3. Number of Experts and Qualifications

    • Experts for Test Set Ground Truth: Three participating experts and one supervising obstetrician.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (supervisor).
      • Two sonographers with more than 10 years of experience, all specializing in fetal cardiology.
      • One obstetrician with more than 25 years of experience (supervisor).

    4. Adjudication Method

    • Similar to BiometryAssist, the document implies a consensus-based approach among the three experts for classifying images and manually drawing anatomy areas, supervised by a fourth obstetrician. No specific "2+1" or "3+1" adjudication method is explicitly stated, but the process involves multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not explicitly mentioned or described for comparing human readers with and without AI assistance. The study focuses on the standalone performance of the AI algorithm.

    6. Standalone Performance

    • Yes, the study describes the standalone performance of the algorithm for view recognition and anatomy annotation (segmentation) without human intervention after the initial image acquisition and ground truthing.

    7. Type of Ground Truth Used

    • Expert consensus. All acquired images for training, tuning, and validation were classified into correct views by experts, and corresponding anatomy areas were manually drawn by them.

    8. Training Set Sample Size

    • Not explicitly stated, but the document mentions that a dataset was acquired for training, tuning, and validation.

    9. How Ground Truth for Training Set Was Established

    • All acquired images for training, tuning, and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each image. The participating experts were composed of an obstetrician with more than 20 years of experience and two sonographers with more than 10 years of experience, all in fetal cardiology, and the entire process was supervised by another obstetrician with more than 25 years of experience.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220043

    Validate with FDA (Live)

    Date Cleared
    2022-04-05

    (90 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan+™ Mode, Combined modes, Multi-Image modes (Dual, Quad), 3D/4D modes.

    Device Description

    The HERA W9/ HERA W10 are general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDD), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S-Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes.

    The HERA W9/HERA W10 also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W10 have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    This document is a 510(k) summary for the HERA W9/W10 Diagnostic Ultrasound System, which explicitly states that clinical studies were not required to support substantial equivalence. Therefore, the document does not contain information about acceptance criteria, device performance, sample sizes, ground truth establishment, or clinical study designs (MRMC, standalone) in the context of proving device performance against acceptance criteria.

    The submission focuses on demonstrating substantial equivalence to a predicate device (HERA W9/W10, K211824) primarily through technological characteristics, intended use, and compliance with non-clinical safety and performance standards.

    The "Summary of Clinical Tests" section on page 7 explicitly states: "The subject of this premarket submission, HERA W9/ HERA W10, did not require clinical studies to support substantial equivalence."

    Therefore, I cannot provide the requested information from the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182595

    Validate with FDA (Live)

    Date Cleared
    2018-12-18

    (89 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    Device Description

    The HERA W10 is general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound dasta and to display the data as B-mode. M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S-Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes. The HERA W10 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W10 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    I am sorry, but the provided text does not contain sufficient information to answer your request. The document is a 510(k) summary for the HERA W10 Diagnostic Ultrasound System, which primarily focuses on establishing substantial equivalence to predicate devices and detailing the device's intended uses and compliance with safety standards.

    Specifically, the text does not include:

    • A table of acceptance criteria and reported device performance.
    • Details about sample sizes used for test sets, data provenance, number or qualifications of experts for ground truth, or adjudication methods for test sets.
    • Information about a multi-reader multi-case (MRMC) comparative effectiveness study or its effect size.
    • Information about a standalone (algorithm only) performance study.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    • The sample size for the training set or how its ground truth was established.

    In fact, section 11 explicitly states: "The subject of this premarket submission, HERA W10, did not require clinical studies to support substantial equivalence." This indicates that detailed performance studies as you've requested were not part of this specific FDA submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1