510(k) Data Aggregation

    K Number
    K062901

    Validate with FDA (Live)

    Device Name
    HEMOCATH II
    Manufacturer
    Date Cleared
    2007-02-08

    (134 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoCath Hemodialysis/Apheresis Catheter is indicated for use in attaining long term vascular access for hemodialysis or apheresis therapy via the jugular or subclavian vein. The catheter is intended for implantation dwell time of greater than 30 days.

    Device Description

    The HemoCath II is radiopaque silicone, comprised of 2 single lumens used to remove and return blood through two segregated circular lumen passages. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed Dacron cuff allows for tissue ingrowth for long-term placement.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text, focused on the HemoCath II device:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Flow Rate: Substantially equivalent to legally marketed predicate devices intended for hemodialysis and apheresis treatments.Data submitted in premarket notification includes in vitro performance data for flow rate, demonstrating substantial equivalence to legally marketed predicate devices.
    Tensile Strength: Substantially equivalent to legally marketed predicate devices intended for hemodialysis and apheresis treatments.Data submitted in premarket notification includes in vitro performance data for tensile strength, demonstrating substantial equivalence to legally marketed predicate devices.

    Study Information

    1. Sample size used for the test set and the data provenance: Not applicable. The submission states, "Clinical data was not deemed necessary since in vitro testing was sufficient to demonstrate safety and efficacy by way of comparison to legally marketed predicate device intended for hemodialysis and apheresis treatments." The testing involved in vitro performance data.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical data or expert-based ground truth was established for a test set.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable, as no clinical data or expert-based ground truth was established for a test set.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a medical catheter and not an AI-driven diagnostic or assistive tool.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a medical catheter and does not involve an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance was established by comparison to the performance of legally marketed predicate devices through in vitro testing. There was no clinical ground truth established for this submission.

    7. The sample size for the training set: Not applicable. There was no "training set" in the context of an AI/ML algorithm. The comparison was made against existing predicate device performance.

    8. How the ground truth for the training set was established: Not applicable. The device is not an AI/ML algorithm requiring a training set with established ground truth. Its safety and efficacy were demonstrated by comparing its in vitro performance characteristics (flow rate, tensile strength) to those of already legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043292

    Validate with FDA (Live)

    Device Name
    HEMOCATH
    Manufacturer
    Date Cleared
    2005-03-16

    (107 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoCath Hemodialysis/Apheresis Catheter is indicated for use in attaining long term vascular access for hemodialysis or apheresis therapy via the jugular or subclavian vein. The catheter is intended for implantation dwell time of greater than 30 days.

    Device Description

    The HemoCath is a radiopaque silicone, double lumen catheter used to remove and return blood through two segregated circular lumen passages. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed Dacron cuff allows for tissue ingrowth for long-term placement.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    This document is a 510(k) summary for a medical device (HemoCath Hemodialysis/Apheresis Catheter) seeking clearance from the FDA. It does not describe an "AI device" or a study in the typical sense of evaluating software performance with AI. Instead, it describes a medical device undergoing a regulatory review process for substantial equivalence to existing legally marketed devices. Therefore, many of the requested fields related to AI, software performance, expert reviews, ground truth, and training sets are explicitly not applicable.

    Here's the information extracted and contextualized:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria/Performance Goal: To demonstrate substantial equivalence to legally marketed predicate devices in terms of performance. The specific performance criteria for central venous catheters (like flow rate and tensile strength) are implied to be met.

    Performance MetricAcceptance Criteria (Implied)Reported Device PerformanceComments
    Flow RateSubstantially equivalent to predicate devicesSubstantially equivalent to predicate devicesThe document states, "Information submitted... includes in vitro performance data for the HemoCath including flow rate... that is substantially equivalent to the legally marketed devices." Detailed quantitative values are not provided in this summary. The expectation is that the HemoCath's flow rate performance is comparable to or within acceptable ranges of the predicate devices.
    Tensile StrengthSubstantially equivalent to predicate devicesSubstantially equivalent to predicate devicesSimilar to flow rate, the document confirms that "tensile strength that is substantially equivalent to the legally marketed devices" was demonstrated. This refers to the catheter's ability to resist breaking under tension, a critical safety and durability aspect.
    Other Performance CharacteristicsSubstantially equivalent to predicate devicesNot explicitly detailed, but implied as "substantially equivalent"The summary indicates that substantial equivalence was shown for "design, material type, performance, and method of sterilization." This implies other relevant performance characteristics for a long-term intravascular catheter (e.g., biocompatibility, resistance to kinking, radiopacity) were also addressed and found to be equivalent, even if not itemized with specific data.

    Study that Proves the Device Meets the Acceptance Criteria:

    The document explicitly states that "Clinical data was not deemed necessary since in vitro testing was sufficient to demonstrate safety and efficacy by way of comparison to legally marketed predicate device intended for hemodialysis and apheresis treatments."

    Therefore, the "study" was an in vitro performance study.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in the 510(k) summary. For in vitro testing of a physical device like a catheter, the sample size would typically refer to the number of catheters tested for each metric (e.g., X number of catheters for flow rate, Y for tensile strength). This level of detail is usually found in the full 510(k) submission, not the summary.
    • Data Provenance: The data is from in vitro testing. It does not involve human subjects or retrospective/prospective collection of patient data. Therefore, country of origin is not applicable in the patient data sense; the testing would have been conducted by the manufacturer or a contract lab.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not applicable. This was an in vitro engineering/performance test, not a study requiring expert clinical review or interpretation of patient data to establish ground truth. The "ground truth" here would be objective physical measurements of the catheter's performance against established engineering standards or predicate device performance.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no human interpretation or subjective assessment of the test set data that would require an adjudication method. The results were objective measurements (e.g., flow rate in ml/min, tensile strength in N).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This is a physical medical device (catheter), not an AI algorithm, and therefore this type of study is not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Objective physical measurements and engineering standards.
      • For flow rate: Measurements of fluid volume over time.
      • For tensile strength: Measurements of force at which the catheter breaks.
      • Comparison was made against the performance of legally marketed predicate devices (DermaPort™ Access Device and RetrO™ Catheter) and likely against relevant international or industry standards for such devices.

    8. The Sample Size for the Training Set

    • Not applicable. This pertains to a physical medical device developed through engineering and manufacturing processes, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, no training set was used for an AI algorithm. The "ground truth" in the context of developing this catheter would involve engineering specifications, materials science data, and performance characteristics derived from predicate devices and industry standards.