Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122612

    Validate with FDA (Live)

    Date Cleared
    2012-09-10

    (14 days)

    Product Code
    Regulation Number
    870.3450
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HEMASHIELD Vascular Grafts are indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. The prosthesis is also recommended for use in patients requiring systemic heparinization prior to, or during, surgery.

    Device Description

    The HEMASHIELD Vascular Grafts are Woven Double Velour polvester grafts impregnated with highly purified collagen. The HEMASHIELD Vascular Grafts are designed to reduce . bleeding at implant and thereby eliminate the operative preclotting step. The collagen is designed to be gradually resorbed by the patient. In addition to collagen, the grafts also contain glycerol as a softening agent.

    AI/ML Overview

    HEMASHIELD Woven Vascular Grafts - Acceptance Criteria and Study Details

    The provided document describes the acceptance criteria and study performed to demonstrate substantial equivalence for the HEMASHIELD Woven Vascular Grafts. As this is a 510(k) submission for a predicate device modification, the focus is on non-clinical bench testing rather than clinical trials with human readers or AI algorithms.

    1. Acceptance Criteria and Reported Device Performance

    The document states that the results of non-clinical tests meet the specified acceptance criteria and are substantially equivalent to the predicate device. However, the exact quantitative acceptance criteria for each test and the reported numerical performance values for the proposed device are not explicitly provided in the given text. The tests performed are listed as:

    Acceptance Criteria (Explicit values not provided)Reported Device Performance
    Water PermeabilityMet acceptance criteria
    Longitudinal Tensile StrengthMet acceptance criteria
    Factory Anastomotic StrengthMet acceptance criteria
    Burst StrengthMet acceptance criteria
    Usable LengthMet acceptance criteria
    Relaxed Internal DiameterMet acceptance criteria
    Pressurized Internal DiameterMet acceptance criteria
    Wall ThicknessMet acceptance criteria
    Suture Retention Strength (Suture Pull Out)Met acceptance criteria
    GuidelineMet acceptance criteria

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: The document does not specify the exact sample size (number of grafts) used for each of the bench tests.
    • Data Provenance: The data is from bench testing conducted by MAQUET Cardiovascular LLC. The context of a 510(k) submission implies this testing was done internally by the manufacturer to support regulatory submission. No country of origin is explicitly stated, but the company address is in Wayne, New Jersey, USA. The testing is prospective in the sense that it was conducted specifically to evaluate the modified device against predetermined criteria.

    3. Number of Experts and Qualifications for Ground Truth of Test Set

    • Not Applicable. This study is a non-clinical bench test evaluating physical properties of a vascular graft, not a diagnostic or AI device that requires expert review for ground truth establishment.

    4. Adjudication Method for Test Set

    • Not Applicable. As this is a non-clinical bench test, there is no need for expert adjudication of results. Each test would have objective measurement outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. An MRMC study was not performed. This type of study is relevant for diagnostic devices involving human readers and potentially AI assistance, which is not the nature of this vascular graft evaluation.

    6. Standalone (Algorithm Only) Performance Study

    • No. This is not an AI algorithm or software-only device. The evaluation focuses on the physical properties of a medical implant.

    7. Type of Ground Truth Used

    • The "ground truth" for the bench tests would be the objective measurements obtained from established testing standards and methodologies (e.g., ISO or ASTM standards, or equivalent internal validated methods). The results of these measurements were then compared against predetermined acceptance criteria, likely derived from the predicate device's performance or relevant standards.

    8. Sample Size for Training Set

    • Not Applicable. This is a non-clinical bench test, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1