510(k) Data Aggregation

    K Number
    K012952

    Validate with FDA (Live)

    Date Cleared
    2001-10-02

    (28 days)

    Product Code
    Regulation Number
    870.3450
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hemashield Gold ™ Woven Double Velour Vascular Graft is indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. The prosthesis is also recommended for use in patients requiring systemic heparinization prior to, or during, surgery.

    Device Description

    The Branch Graft Configurations is different from the rest of the Hemashield Gold Woven Double Velour Vascular Graft product line only in the 1-4 branch configuration in which the grafts are sewn. The modified vascular grafts have the following similarities to those which previously received 510(k) concurrence: Identical indications for use, Identical labeling, Identical manufacturing processes, Identical operating principle, Incorporate identical materials, Have the identical shelf-life (5 years), Are packaged and sterilized using identical packaging materials and processes.

    AI/ML Overview

    The provided submission for the Hemashield Gold Woven Double Velour Branch Graft is a Special 510(k) for a device modification, focusing on the branch configuration. The document explicitly states that there is "no change to these tests or to their acceptance criteria for the Branch Graft line extension" and that "all testing was done with standard test methods for these parameters". This indicates that the device's performance for these criteria was demonstrated to meet existing standards, rather than new studies being conducted specifically for this modification.

    Due to the nature of this submission being a device modification to an already approved device and the stated lack of change in acceptance criteria for the branch graft line extension, detailed new studies with specific sample sizes, ground truth establishment, or multi-reader multi-case analyses are not part of this document. The performance is assumed to be in compliance with the established methods for the predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedAcceptance CriteriaReported Device Performance
    Visual Inspection - Stains$\le$ 0.05mm²Met (Implied by "Pass" for related categories)
    Visual Inspection - Foreign Matter$\le$ 0.03mm²Met (Implied by "Pass" for related categories)
    Water PermeabilityAccept: if $\le$ 4.0 mL $\cdot$ cm $^{-2}\cdot$ min $^{-1}$Reject: if > 4.5 mL $\cdot$ cm $^{-2}\cdot$ min $^{-1}$Repeat test if > 4.0 and $\le$ 4.5 mL $\cdot$ cm $^{-2}\cdot$ min $^{-1}$If re-test values are all < 4.5 mL $\cdot$ cm $^{-2}\cdot$ min $^{-1}$ then batch is acceptable.Met
    Burst> 399.9 lb/inch²Met
    Crimp12.0 turns/inch - circularMet
    Usable LengthMain Graft - 45 ± 5cmHead Branches - 17.5 ± 2.5cmPerfusion Branch - 22.5 ± 2.5cmMet
    Inner Diameter$\le$ 10.0 mm sizes; nominal diameter ± 0.5 mm> 10.0 mm sizes; nominal diameter ± 1.0 mmMet
    Visual Inspection - SewingNo sewing defectsMet (Implied by "Pass" for related categories)
    Air Permeability$\le$ 24.5 - 43.5 l/m (depending upon size)Met
    Pyrogenicity: Rabbit PyrogenPassPass
    Acute Systemic Toxicity: Acute Systemic InjectionPassPass
    Irritation: Acute Intracutaneous ReactivityPassPass
    Hemocompatibility: HemolysisPassPass
    Cytotoxicity: ISO L929 MEM ElutionPassPass
    Mutagenicity: Ames Test (Saline extract)PassPass
    Subchronic toxicity: 14 day IV injection in ratsPassPass
    Implantation: ISO Muscle Implantation with histopathology (2 week)PassPass
    Implantation: ISO Muscle Implantation with histopathology (12 week)PassPass
    PhysiochemicalPassPass
    Shelf Life TestingMeets Product Specification after agingMet Product Specification after aging (Implied by "Pass")

    Information Not Applicable to this Submission:

    The following information is not provided in the original text as the submission is for a device modification where existing acceptance criteria and methods were used, rather than a new de novo study with specific performance endpoints outlined in these categories. The emphasis is on substantial equivalence to the predicate device, confirming that the modified device continues to meet the established performance and safety characteristics of the existing product line.

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
      • Not specified. The document indicates that "All testing was done with standard test methods for these parameters" and that "There is no change to these tests or to their acceptance criteria for the Branch Graft line extension." This suggests that the established test methods for the predicate device were applied, without specifying new sample sizes or data provenance for this specific modification.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
      • Not applicable. This is a medical device for vascular repair, not an image analysis or diagnostic device requiring expert interpretation for ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set
      • Not applicable. This is a medical device for vascular repair.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
      • Not applicable. This is a medical device for vascular repair, not an AI or diagnostic device involving human reader performance.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
      • Not applicable. This is a medical device for vascular repair, not an AI or diagnostic device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
      • For the physical and biological tests, "ground truth" is established by the specified acceptance criteria for each test (e.g., specific dimensions, permeability rates, "Pass" for biocompatibility).
    7. The sample size for the training set
      • Not applicable. This is a medical device for vascular repair, not a machine learning model requiring a training set.
    8. How the ground truth for the training set was established
      • Not applicable. This is a medical device for vascular repair.