Found 2 results

510(k) Data Aggregation

    K Number
    K222270

    Validate with FDA (Live)

    Date Cleared
    2022-10-21

    (84 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HEDRON C® Spacers and HEDRON IC® Spacers are interbody fusion devices indicated at one or more levels of the cervical spine (C2-T1) in patients with cervical disease, instability, trauma including fractures, deformity defined as kyphosis, lordosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. Cervical disc disease is defined as intractable radiculopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment.

    HEDRON C® Spacers and HEDRON IC® Spacers are intended to be used with supplemental fixation, such as an anterior cervical plate or posterior cervical fixation. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

    When the HEDRON IC® Spacer is used with the COALITION AGX® Plate, the plate-spacer assembly (HEDRON IC® Plate-Spacer) is a stand-alone device intended for use at one or two levels of the cervical spine (C2-TI) in patients with cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. These devices are to be used with two titanium alloy screws which accompany the implants (≥20°) must be used with supplemental fixation in addition to the two screws.

    HEDRON® Lumbar Spacers (HEDRON A™, HEDRON P®, HEDRON T™, and HEDRON RT™) are lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T -T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be sketally mature and have had at least six (6) months of non-operative treatment. HEDRON® Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.

    HEDRON IA® Integrated Lumbar Spacers are integrated lumbar interbody fusion devices indicated for use at one or more levels of the lumbosacral spine (L1-S1), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DD), disc herniation (with myelopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. These devices are intended to be used with or without three screws and or anchors which accompany the implants. These devices are with supplemental fixation (e.g. facet screws or posterior fixation). In addition, these devices are intended for stand-alone use in patients with DDD at one or two levels only when <259 lordotic implants are used with three screws per implant. HEDRON IA® Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.

    SABLE® Expandable Spacer is an interbody fusion device intended for use at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The spacer is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone and is to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, and anterior screw and rod systems).

    The ExcelsiusGPS® is intended for use as an aid for precisely locating anatomical structures and for spationing and orientation of an instrument holder or guide tube to be used by surgeons for navigating compatible surgical instruments in open or percutaneous provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody spacers.

    Device Description

    HEDRON® Cervical Spacers (HEDRON C® and HEDRON IC®) are anterior cervical interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. HEDRONIC® Spacer may be assembled to the COALITION AGX® Plate to create the HEDRON IC® Plate-Spacer, a standalone cervical interbody fusion device. HEDRON® Spacers are additively manufactured from titanium allov. as specified in ASTM F3001.

    HEDRON® Lumbar Spacers are lumbar interbody fusion devices used to provide structural stability following discectomy. Each HEDRON® spacer has a different shape to accommodate various surgical approaches to the spine, including anterior, anterolateral, lateral, posterior or transforaminal approaches. All approaches are used in the lumbar spine; only anterolateral, or lateral approaches are used in the thoracic spine.

    HEDRON® Integrated Lumbar Spacers are integrated anterior lumbar interbody fusion devices used to provide structural stability following discectomy. They are used with or without screws and/or anchors.

    All HEDRON® Lumbar Spacers are additively manufactured from titanium powder. The mating screws and anchors are manufactured from titanium alloy and/or cobalt chrome alloy. Titanium screws and anchors are available with or without hydroxyapatite (HA) coating.

    The SABLE® Expandable Spacer is an expandable lumbar interbody fusion device used to provide structural stability in skeletally mature individuals following discectomy. The device is available in various heights and geometric options to fit the anatomical needs of a wide variety of patients.

    SABLE® Spacers are manufactured from titanium alloy. The endplates are additively manufactured from titanium alloy powder and an internal component is manufactured from radiolucent PEEK polymer. The drive screw is manufactured from cobalt chromium alloy.

    ExcelsiusGPS® Instruments are nonsterile, reusable instruments that can be operated with ExcelsiusGPS® and may be used for a freehand navigated surgical procedure.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for several spinal implant devices (HEDRON® Cervical Spacers, HEDRON® Lumbar Spacers, SABLE® Expandable Spacer, and ExcelsiusGPS® Instruments). However, it does not contain information about an AI/ML-based device.

    Therefore, I cannot provide the detailed information requested regarding machine learning acceptance criteria, study design, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.

    The document focuses on the substantial equivalence of the new devices to existing predicate devices based on:

    • Mechanical testing: Dynamic compression-shear testing was conducted in accordance with "Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Intervertebral Fusion Device," June 12, 2007, and ASTM F2077.
    • Technological Characteristics: The subject implants are stated to have the same technological characteristics as the predicate devices, including design, intended use, material composition, and range of sizes.
    • Basis of Substantial Equivalence: The information provided within the premarket notification supports substantial equivalence of the subject spacers to the predicate devices with respect to technical characteristics, performance, and intended use.

    In summary, the document addresses the regulatory requirements for hardware medical devices and does not involve AI/ML components or related performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191243

    Validate with FDA (Live)

    Date Cleared
    2019-09-17

    (131 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HEDRON C™ Spacers and HEDRON IC™ Spacers are interbody fusion devices indicated at one or more levels of the cervical spine (C2-T1) in patients with cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, lordosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. Cervical disc disease is defined as intractable radiculopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment.

    HEDRON C™ Spacers and HEDRON IC™ Spacers are intended to be used with supplemental fixation, such an anterior cervical plate or posterior cervical fixation. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

    When the HEDRON IC™ Spacer is used with the COALITION AGX® Plate, the plate-spacer assembly (HEDRON IC™ Plate-Spacer) is a stand-alone device intended for use at one or two levels of the cervical spine (C2-T1) in patients with cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. These devices are to be used with two titanium alloy screws which accompany the implants (≥20°) must be used with supplemental fixation in addition to the two screws.

    Device Description

    HEDRON™ Cervical Spacers (HEDRON CTM and HEDRON ICT™) are anterior cervical interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. HEDRON ICT™ Spacer may be assembled to the COALITION AGX® Plate to create the HEDRON ICT™ Plate-Spacer, a stand-alone cervical interbody fusion device. HEDRON™ Spacers are additively manufactured from titanium alloy, as specified in ASTM F3001.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (HEDRON™ Cervical Spacers), which focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and performance data. It does not describe an AI medical device or a clinical study that establishes acceptance criteria and proves the device meets those criteria from an AI performance perspective.

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving device performance for an AI model, as this document is not about an AI medical device.

    The document discusses:

    • Device Name: HEDRON™ Cervical Spacers
    • Intended Use: Interbody fusion devices for cervical spine
    • Mechanical Testing: Static and dynamic compression, torsion, compression-shear, subsidence, and expulsion in accordance with specific ASTM standards and FDA guidance.
    • Basis of Substantial Equivalence: Similar technological characteristics (design, intended use, material composition, sizes) to predicate devices.

    However, none of this relates to AI performance metrics, ground truth establishment, expert adjudication, or MRMC studies for AI devices.