510(k) Data Aggregation

    K Number
    K183279

    Validate with FDA (Live)

    Date Cleared
    2019-02-21

    (90 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HEALIX TiTM Anchors with DYNACORD™ is intended for:

    Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromioclavicular; Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction:

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. Patellar Tendon repair and secondary fixation in ACL/PCL reconstruction repair:

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction;

    Hip: Capsular repair, Acetabular Labral Repair.

    Device Description

    The HEALIX TiTM Anchor with DYNACORDTM Suture is a threaded suture anchor preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. HEALIX TiTM Anchors with DYNACORDTM Suture is available in Titanium material, which is a common material used on existing Depuy Mitek devices. Devices with needles will be offered to facilitate suture passage through tissue. The HEALIX TiTM Anchor with DYNACORDTM Suture is provided sterile and is for single use only.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the HEALIX Ti™ ANCHOR with DYNACORD™. It establishes substantial equivalence to a predicate device, meaning it does not require a new study to prove its safety and effectiveness in the same way a novel device would. Therefore, the information typically found in an AI/Software as a Medical Device (SaMD) submission regarding acceptance criteria, training/test sets, expert adjudication, and MRMC studies is not applicable here.

    The provided text focuses on the non-clinical testing performed to demonstrate the device's substantial equivalence to a predicate, rather than a clinical study evaluating the performance of an AI/SaMD.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance:

    The document describes non-clinical testing, not performance against specific clinical acceptance criteria of an AI/SaMD. The "acceptance criteria" here relate to engineering and material performance for a physical implant.

    Acceptance Criteria (Type of Test)Reported Device Performance (Summary)
    Anchor fixation testingPerformed; results demonstrated suitability for intended use
    Torque testingPerformed; results demonstrated suitability for intended use
    Suture testing (per USP)Performed; results demonstrated suitability for intended use
    Approximation force testingPerformed; results demonstrated suitability for intended use
    Biological evaluationConducted
    SterilityConducted
    PackagingConducted
    Shelf-life testingConducted
    Bacterial endotoxin testingMet endotoxin limits

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in terms of number of anchors tested, but general non-clinical testing was conducted.
    • Data Provenance: Not applicable in the context of clinical data for AI/SaMD. This refers to lab-based testing of physical properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is not an AI/SaMD and does not involve human expert labeling of data. The "ground truth" for these tests would be the established engineering and materials science standards.

    4. Adjudication method for the test set:

    • Not applicable for non-clinical engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical medical device, not an AI/SaMD.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an AI/SaMD.

    7. The type of ground truth used:

    • For the non-clinical testing, the "ground truth" is defined by established engineering standards, USP specifications for suture materials, and biological safety standards (e.g., ISO standards for biocompatibility, endotoxin limits).

    8. The sample size for the training set:

    • Not applicable. This is not an AI/SaMD.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/SaMD.

    In summary, the provided document is a regulatory submission for a physical medical device (suture anchor) based on demonstrating substantial equivalence to an existing predicate device through non-clinical performance and safety testing, not through clinical trials or evaluations of an AI/SaMD.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082282

    Validate with FDA (Live)

    Device Name
    HEALIX TI ANCHOR
    Date Cleared
    2008-11-07

    (88 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shoulder: Rotor Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction; Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair; Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Patella tendon repair and secondary fixation in ACL / PCL reconstruction repair. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Hip: Capsular repair, acetabular labral repair.

    Device Description

    The Healix Ti Anchor is a non-absorbable threaded suture anchor manufactured of Titanium material. The threaded anchor comes preloaded on a disposable inserter assembly and is intended for fixation of #2 suture to bone. The anchor is provided in three sizes: one with an outer diameter of 4.5mm, another with an outer diameter of 5.5mm and the third with an outer diameter of 6.5mm. The suture options may or may not include tapered needles to facilitate suture passage through The Healix Anchors are currently offered with absorbable tissue. Panacryl, non-absorbable Ethibond or partially absorbable Orthocord suture options.

    AI/ML Overview

    The provided text is a 510(k) summary for the Healix Ti Anchor, a medical device. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a clinical study or performance testing that would generate the kind of data requested in your prompt. This type of submission relies on comparisons to already approved devices.

    Therefore, many of the requested items (e.g., acceptance criteria table, sample size, ground truth, MRMC study, training set information) are not applicable or available within this document.

    Here's an assessment based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Not applicable / Not provided. The 510(k) summary states, "Results of performance and safety testing have demonstrated that the modified device is substantially equivalent to the predicate devices." It does not list specific acceptance criteria (e.g., pull-out strength, fatigue life) or provide quantitative performance data for such criteria directly within this document. The "tests" performed are implicit in demonstrating substantial equivalence, likely through mechanical bench testing, but the details are not given here.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not provided. No specific 'test set' in the context of clinical or performance data is described in this summary. The substantial equivalence argument relies on comparison to predicate devices and likely bench testing, not a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not provided. Ground truth, in the context of expert consensus or clinical outcomes, is not relevant to this 510(k) submission, which is for a mechanical bone anchor and focused on substantial equivalence to existing devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. Adjudication methods are typically for clinical studies or image interpretation, which are not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a mechanical bone anchor, not an AI-powered diagnostic tool, so an MRMC study is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a mechanical bone anchor, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable / Not provided. Ground truth in the sense of clinical outcomes or expert labels is not part of this 510(k) summary for a mechanical device. The "ground truth" for substantial equivalence for such a device is typically established through a combination of engineering analysis, material properties testing, and mechanical performance tests (e.g., pull-out strength, fatigue) compared against the predicate devices. These specific tests and their "ground truth" values (e.g., meeting a certain strength threshold) are not detailed in this high-level summary.

    8. The sample size for the training set

    • Not applicable. This document describes a mechanical medical device, not a machine learning model, so there is no 'training set'.

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no training set for this type of device.

    Summary based on the document:

    The provided document is a 510(k) summary for the Healix Ti Anchor. Its purpose is to demonstrate substantial equivalence to legally marketed predicate devices, not to present performance data against specific acceptance criteria from a primary clinical study. The "study that proves the device meets the acceptance criteria" is implicitly the "performance and safety testing" mentioned, but the details of these tests, specific criteria, and quantitative results are not included in this summary. Such details would typically be found in the comprehensive 510(k) submission itself, but not in this summary section.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1