Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K041924

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-09-09

    (55 days)

    Product Code
    Regulation Number
    890.3690
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hausted® POWERTRAN™ Series Stretcher is a powered patient transport. This device is a motorized device intended for medical purposes to assist in the transfer of patients to and from the bath, beds, chairs, treatment modalities, and transport vehicles. The FDA has classified powered patient transports as a Class II device.

    Device Description

    The Hausted® POWERTRAN™ Series Stretcher is a powered patient transport. This device is a motorized device intended for medical purposes to assist in the transfer of patients to and from the bath, beds, chairs, treatment modalities, and transport vehicles. The FDA has classified the device as a Class II device. The FDA has classified powered patient transports as a Class II device.

    The Hausted® POWERTRAN™ Series Stretcher is designed to assure compliance with IEC 60601-1 (Electrical Safety) and IEC 60601-1-2 (Electromagnetic Compatibility). The device will also carry the ETL (to UL 2601-1) and cETL (to CAN/CSA C22.2 No. 601.1-M90) markings. Electrical testing will be certified by the accredited testing facility shown at the end of this certification. Software testing is being certified by STERIS as the manufacturer.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "The Hausted® POWERTRAN™ Series Stretcher." This document describes a submission for regulatory clearance, not a study that proves the device meets acceptance criteria in the way clinical studies for AI/diagnostic devices do.

    Therefore, many of the requested categories for AI/diagnostic device studies (like sample size for test/training sets, ground truth establishment, expert qualifications, MRMC studies, and standalone performance) are not applicable to this submission.

    However, I can extract the relevant information regarding the acceptance criteria and how "proof" is presented in this context, which is primarily through safety and performance testing against recognized standards and comparison to predicate devices.

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Standard/Requirement)Reported Device Performance/Compliance
    Electrical SafetyIEC 60601-1 (Electrical Safety), Amendment 1, 1991-11, Amendment 2, 1995-03The device is designed to assure compliance with IEC 60601-1. Electrical testing will be certified by an accredited testing facility, and the device will carry ETL (to UL 2601-1) and cETL (to CAN/CSA C22.2 No. 601.1-M90) markings. Safety Testing and performance characteristics have been conducted and successfully completed.
    EMCIEC 60601-1-2 (Electromagnetic Compatibility - Requirements and Tests), 1993-04The device is designed to assure compliance with IEC 60601-1-2. Safety Testing and performance characteristics have been conducted and successfully completed.
    General Safety (US)UL 2601-1 (2nd Ed.) Standard for Medical Electrical Equipment, Part 1: General Requirements for SafetyThe device will carry ETL (to UL 2601-1) marking. Safety Testing and performance characteristics have been conducted and successfully completed.
    General Safety (Canada)CAN/CSA C22.2 No. 601.1-M90 Standard for Medical Electrical EquipmentThe device will carry cETL (to CAN/CSA C22.2 No. 601.1-M90) marking. Safety Testing and performance characteristics have been conducted and successfully completed.
    Risk ManagementISO 9001, Design Control 4.4 and (implied) assessment of known and reasonable hazardsAn assessment of known and reasonable hazards has been conducted to ensure that any risk associated with the device as of the date of product release is as low as reasonably possible. Design review has been conducted by a cross-functional team.
    Software Quality(Not explicitly-stated standard, but mentioned as distinct testing)Software testing is being certified by STERIS as the manufacturer.
    Substantial EquivalenceComparison to Predicate Devices: Stryker ZOOM® Motorized Stretchers and Stryker Powered Wheeled StretcherThe Hausted® POWERTRAN™ Series Stretcher is substantially equivalent to the predicate devices in function and intended use. Minor differences do not raise new issues of safety or effectiveness. The intended use, basic technology, and performance characteristics are the same.

    Non-Applicable / Not Provided Information for AI/Diagnostic Device Studies:

    1. Sample size used for the test set and the data provenance: Not applicable. This is a physical medical device, not an AI/diagnostic software. Testing involves physical device evaluations against engineering standards, not data sets for AI.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's safety and performance is established by compliance with engineering and safety standards, and by comparison to predicate devices, not by expert consensus on data.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For a physical device like a stretcher, "ground truth" relates to engineering specifications and compliance with recognized safety standards (like IEC, UL, CSA). It means the device physically meets the requirements set forth in those standards and performs its intended function safely, as demonstrated through testing and design reviews.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of "Study" (in this context):

    The "study" or rather, the evidence presented for this 510(k) submission, is a comprehensive set of safety and performance characteristic tests conducted by the manufacturer (STERIS) and certified by accredited testing facilities. These tests ensure compliance with various national and international electrical, electromagnetic compatibility (EMC), and general safety standards (IEC 60601-1, IEC 60601-1-2, UL 2601-1, CAN/CSA C22.2 No. 601.1-M90). Additionally, a risk assessment and design review process (following ISO 9001 principles) were conducted by a cross-functional team. The core of the 510(k) argument is the demonstration of substantial equivalence to already legally marketed predicate devices (Stryker ZOOM® Motorized Stretchers and Stryker Powered Wheeled Stretcher) in terms of intended use, basic technology, and performance characteristics. The minor differences were deemed not to raise any new issues of safety or effectiveness.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1