Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160752

    Validate with FDA (Live)

    Date Cleared
    2016-06-29

    (103 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Harmonic HD 1000i Shears instrument is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gynecologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.

    Device Description

    The HARMONIC HD 1000i Shears are a sterile, single patient use device used for dissection, grasping, coagulation, and cutting between the blade and clamp arm. It consists of an ergonomic grip housing assembly with two hand control buttons: 1) Energy button for power levels 1-5 and, 2) Energy with Advanced Hemostasis button for large vessel sealing. The device is available in two shaft lengths, 20 cm and 36 cm, and both device shafts can be rotated continuously to facilitate visualization and access to targeted tissue. An integrated audible and tactile mechanism in the grip housing indicates full trigger closure. The device has a clamp arm and coated curved blade that is designed to work through a 5 mm trocar, through a 5 mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The blade used in the HARMONIC HD 1000i Shears has a slightly different blade design than that of the predicate device, the HARMONIC ACE+Shears with Advanced Hemostasis (cleared under K132612 on October 17, 2013), and is also slightly longer and more tapered than the predicate device. Additionally, two dashes have been added to the device which are intended to represent relative vessel size. The distal/front Energy button is indicated for vessels up to 5 mm in diameter. The Energy with Advanced Hemostasis button (the Green button), which is found on the side of the device handle, is designed for larger vessels and is indicated for vessels up to 7 mm in diameter. The device utilizes Adaptive Tissue Technology (the technological characteristics of which were first cleared under K120729 on May 17, 2012). This technology provides the generator with the ability to identify and monitor the instrument during use. The HARMONIC HD 1000i Shears are designed for use exclusively with the Ethicon Generator 11 (GEN11).

    AI/ML Overview

    The provided text describes the HARMONIC HD 1000i Shears. Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance & Adjudication method

    The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, it describes various tests performed and states that the "Data generated from the bench testing met the predetermined acceptance criteria" and that "the results of the acute/survival studies demonstrated that the tissue effects were not different than the predicate device."

    Here's an attempt to structure the information into a table format based on the descriptions:

    Test / Performance MetricAcceptance Criteria (Implicit)Reported Device PerformanceAdjudication Method
    SterilizationSterilized to the same sterility assurance level as the predicate device.Both the subject and predicate devices are sterilized via ethylene oxide and to the same sterility assurance level.Not explicitly stated, but implicitly a pass/fail against established sterility assurance level.
    BiocompatibilityNo new materials introduced, therefore no new biocompatibility concerns.Biocompatibility testing was not required as no new materials were introduced.Not applicable (no test performed).
    Electrical SafetyCompliance with IEC 60601-1:2005.The system was shown to comply with IEC 60601-1:2005 for electrical safety.Not explicitly stated, but typically involves objective measurement against standard thresholds.
    EMCCompliance with IEC 60601-1-2:2007.The system was shown to comply with IEC 60601-1-2:2007 for electromagnetic compatibility.Not explicitly stated, but typically involves objective measurement against standard thresholds.
    Bench TestingPredetermined acceptance criteria (specific values not provided in text).Data generated from the bench testing (tissue pad life, tissue pad removal force, instrument grasping force, sealed vessel burst test) met the predetermined acceptance criteria.Not explicitly stated, but implies a comparison against pre-defined numerical limits or ranges for each metric.
    Acute Animal TestingTissue effects not different from the predicate device (for hemostatic transection, lateral thermal damage, tissue dissection, back cutting).The results of the acute studies demonstrated that the tissue effects were not different than the predicate device for hemostatic transection of vessel pedicles (< 5mm using Energy button, ≤ 7mm using Advanced Hemostasis button), lateral thermal damage, tissue dissection, and back cutting on a porcine model.Comparison with predicate device, likely involving expert observation and measurement of tissue effects (e.g., thermal spread, completeness of transection, hemostasis). The standard is "not different," implying qualitative and quantitative comparison.
    Survival Animal TestingTissue effects not different from the predicate device (for hemostatic transection, healed vessel seals under blood pressure challenge).The results of the survival studies demonstrated that the tissue effects were not different than the predicate device for hemostatic transection of vessel pedicles (≤ 5mm using Energy button, ≤ 7mm using Advanced Hemostasis button) in splenectomy and carotid artery procedures, and in blood pressure challenge for healed vessel seals.Comparison with predicate device, likely involving expert observation during necropsy and analysis of tissue samples, including evaluation of healed vessel seals under stress. The standard is "not different."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated for bench tests. For animal studies, the document mentions "an acute porcine study" and "a chronic survival study" without specifying the number of animals or specific data points.
    • Data Provenance: The animal studies were conducted using a porcine (pig) model. The country of origin of the data is not specified, but the submission is to the FDA in the USA. The studies are prospective in nature, as they involve performing new tests to evaluate the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their qualifications. However, for animal studies, evaluation of tissue effects and necropsy would typically involve veterinary pathologists or surgeons with expertise in animal physiology and surgical pathology.

    4. Adjudication Method for the Test Set

    The adjudication method is not explicitly described as a formal process like "2+1" or "3+1." The determination of "not different than the predicate device" in animal studies implies expert judgment and comparison, likely by the researchers conducting the study. For bench testing, it would involve confirming measurements against predefined criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done, as this is a medical device (surgical shears) and not an imaging or diagnostic AI device that would typically involve multiple readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. The HARMONIC HD 1000i Shears are a surgical instrument requiring human operation, not a standalone algorithm.

    7. Type of Ground Truth Used

    The ground truth for the performance evaluations (especially animal studies) appears to be:

    • Expert Observation/Assessment: Direct observation and assessment of tissue effects (hemostasis, thermal damage, completeness of transection) by qualified personnel during and after surgical procedures.
    • Pathology/Histology: Implied for detailed assessment of tissue damage and healing in animal studies (though not explicitly called out as "pathology," it's the standard for evaluating tissue effects).
    • Objective Measurements: For bench tests (e.g., burst pressure, grasping force).

    8. Sample Size for the Training Set

    Not applicable. This device is a mechanical surgical instrument, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1