Found 3 results

510(k) Data Aggregation

    K Number
    K181216

    Validate with FDA (Live)

    Date Cleared
    2019-08-02

    (452 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates.

    Intended areas of use:

    · In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room

    · During transfer of ventilated patients within the hospital

    The HAMILTON-C1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates.

    Intended areas of use:

    · In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room

    • · For emergency medical care
    • · During transport within and outside the hospital
    • · During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship

    The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    Device Description

    The HAMILTON-C1 and HAMILTON-T1 ventilators are designed for adults, pediatrics, infants and neonatal patients requiring invasive or non-invasive ventilation support. All ventilators cover a range of clinical requirements, including invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and Non-Invasive Ventilation.

    The previously cleared ventilators, HAMILTON-C1 and HAMILTON-T1 (both K140939), have been bundled together in this 510(k) submission in order to add the following new features:

    A modification to the software allows compatibility with the Nihon Kohden -SpO2 sensors to be used with HAMIILTON-C1 and HAMILTON-T1 ventilators. These ventilators are already compatible and cleared for use with Masimo SpO2 sensors (K140939).

    • cFlow was added, which continuously delivers an air/gas mixture.
    • -A modification to the software, which allows HAMILTON-C1/T1 ventilators to be compatible for use with speaking valves. A speaking valve allows tracheostomized adult and pediatric patients to communicate verbally.
    AI/ML Overview

    The provided text describes the Hamilton-C1 and Hamilton-T1 ventilators and their clearance by the FDA. The submission focuses on adding new features to already cleared devices.

    Here's an analysis of the provided information regarding acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria for the new features (cFlow, Speaking Valve, and Nihon Kohden SpO2 sensor compatibility) and their reported performance. Instead, it states that "the data provided from these tests was shown to be equivalent to the legally marketed devices." This implies that the acceptance criterion was equivalence to predicate/reference devices, and the reported performance met this criterion.

    However, the "SUMMARY OF THE TECHNOLOGY AND PERFORMANCE SPECIFICATIONS COMPARISON WITH THE PREDICATED DEVICES" (Table 1 on pages 5-6) compares various parameters of the proposed HAMILTON-C1/T1 with the predicate HAMILTON-C1/T1 (K140939). For all listed parameters (Indications of Use, Control settings, Modes of ventilation, Alarms), the comparison states "Substantially Equivalent" or "Equivalent". While this table doesn't detail performance metrics for the new features, it confirms that the device, with its existing features, is considered equivalent to its predicate.

    For the new features, the text mentions:

    • cFlow: "cFlow was added, which continuously delivers an air/gas mixture." It later states, "The Nihon Kohden NKV-550 Series Ventilator System is used as a reference device for the proposed HAMILTON-C1/T1 as both the reference device and the proposed device are intensive care ventilators which have the O2 therapy/cFlow feature." This implies the cFlow feature in HAMILTON-C1/T1 is equivalent to that in the Nihon Kohden NKV-550.
    • Nihon Kohden SpO2 sensors: "A modification to the software allows compatibility with the Nihon Kohden -SpO2 sensors to be used with HAMIILTON-C1 and HAMILTON-T1 ventilators." The document states, "The HAMILTON-C3 is used as a reference device for the proposed HAMILTON-C1/T1 as both the reference device and the proposed device are intensive care ventilators which can be used with Nihon Kohden SpO2 sensors and accessories." This suggests compatibility and performance with these sensors are equivalent to the HAMILTON-C3.
    • Speaking Valve: "A modification to the software, which allows HAMILTON-C1/T1 ventilators to be compatible for use with speaking valves." "The Esprit Ventilator V200 with Speaking mode Option is used as a reference device for the proposed Speaking valve compatibility on the modified HAMILTON-C1/T1..." This indicates equivalence to the Esprit Ventilator V200's speaking mode.

    Implicit Acceptance Criteria and Reported Performance for New Features:

    Acceptance Criteria (Implicit)Reported Device Performance
    cFlow Feature: Performance and safety of continuous delivery of air/gas mixture should be equivalent to the reference device (Nihon Kohden NKV-550 Series Ventilator System)."The data provided from these tests was shown to be equivalent to the legally marketed devices." (Referring to comparison testing with legally marketed devices for the new features).
    Nihon Kohden SpO2 Sensor Compatibility: Successful compatibility and performance with Nihon Kohden SpO2 sensors should be equivalent to the reference device (HAMILTON-C3), which also supports these sensors. Compatibility with existing Masimo SpO2 sensors should be maintained."The data provided from these tests was shown to be equivalent to the legally marketed devices." (Referring to comparison testing with legally marketed devices for the new features).
    Speaking Valve Compatibility: Safe and effective operation with speaking valves for tracheostomized adult and pediatric patients should be equivalent to the reference device (Esprit Ventilator V200 with Speaking mode Option) and maintain functionality in invasive modes. Waveform characteristics should be comparable to legally marketed devices."The data provided from these tests was shown to be equivalent to the legally marketed devices." Additionally, "waveform comparison testing was completed for the Speaking Valve Feature." and "The data provided from these tests was shown to be equivalent to the legally marketed devices."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify exact sample sizes (e.g., number of patients, number of test cases) used for the comparison testing of the new features. It only states that "Testing of the modified HAMILTON-C1/T1, with the new features, was conducted."
    The provenance of the data (country of origin, retrospective/prospective) is also not mentioned. The non-clinical performance tests are generally bench testing and software validation, not human studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the document describes non-clinical performance tests for a ventilator, not diagnostic imaging or AI performance where human expert consensus would establish ground truth. The "ground truth" for these tests would be established through engineering specifications, validated test equipment, and established medical device standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The testing described is non-clinical performance and software validation, not a multi-reader clinical study requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a ventilation device, not a diagnostic AI tool, so no MRMC study or AI assistance improvement for human readers would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not entirely applicable in the context of a ventilator. However, the software validation and bench testing can be considered "standalone" in the sense that the device's technical performance and software functions were tested independently of a human operator making clinical decisions or interpreting results in a real patient scenario. The tests confirm the device operates as intended according to its design specifications and relevant standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical performance tests and software validation would be based on:

    • Engineering Specifications: The defined functional requirements and performance limits of the device.
    • International and National Standards: Compliance with standards like ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 80601-2-12, etc. (listed in Section IX) which define acceptable ranges and behaviors for critical care ventilators, electromagnetic compatibility, alarm systems, and usability.
    • Predicate/Reference Device Performance: The established, legally marketed performance of the predicate HAMILTON-C1/T1 and the reference devices (Nihon Kohden NKV-550, HAMILTON-C3, Esprit Ventilator V200) for the new features. Equivalence to these cleared devices serves as a form of "ground truth" for substantial equivalence.

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning product that requires a training set. The changes described are software modifications for compatibility and feature additions to a pre-existing ventilator design.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140939

    Validate with FDA (Live)

    Date Cleared
    2015-09-02

    (509 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.
    Intended areas of use:

    • In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
    • For emergency medical care
    • During transport within and outside the hospital
    • During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship
      The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.
    Intended areas of use:

    • In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
    • During transfer of ventilated patients within the hospital
      The HAMILTON-C1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
    Device Description

    The HAMILTON-C1 and HAMILTON-T1 are designed for adults, pediatrics, infants and neonatal patients requiring invasive or non-invasive ventilation support. Both ventilators cover a full range of clinical requirements: including invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and non-invasive ventilation.
    The two previously cleared ventilators, the HAMILTON-T1 and the HAMILTON-C1, have been bundled together in this 510(k) submission, in order to add the following new features to both ventilators:
    Neonatal patients with a minimum weight of 0.2 kg and a minimal tidal volume of 2 mL. The following two new modes for the neonatal patient group: nCPAP and nCPAP-PC; SpO2 monitoring with MASIMO PULSE OXIMETERS;
    An increase in the battery duration from 5.5 hours on the HAMILTON-T1, to 9.25 hours (maximum) instead and an increased battery duration from 2 hours to 4.30 hours on the HAMILTON-C1.
    Increased temperature range for the HAMILTON-T1 to 50°C [122°F] for adult and pediatric patients. Increased altitude operating condition for the HAMILTON-T1 from 15.091 ft to 25.000 ft for adult and pediatric patients.

    AI/ML Overview

    This document is a 510(k) summary for the Hamilton-T1 and Hamilton-C1 ventilators, detailing their substantial equivalence to predicate devices and the performance data supporting this claim.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a quantitative table format with corresponding reported device performance for specific features. However, it implies that the devices meet a set of performance and safety standards, and that their performance on new features (neonatal ventilation, improved battery life, increased temperature/altitude range, SpO2 monitoring, and new modes) has been demonstrated to be "substantially equivalent" to legally marketed predicate devices.

    The "Difference Status" column in the comparison tables (Table 1 for HAMILTON-T1 and Table 2 for HAMILTON-C1) implicitly acts as a form of "acceptance criteria met" by referencing substantial equivalence to specific predicate devices for each new or changed feature.

    FeatureAcceptance Criteria (Implied)Reported Device Performance (Implied)
    Neonatal Ventilation (HAMILTON-T1 & C1)Must be substantially equivalent to the HAMILTON-C2 (K121225) for neonatal patient group. Minimum weight 0.2 kg, minimum tidal volume 2 mL.Demonstrated substantial equivalence to HAMILTON-C2 (K121225).
    nCPAP and nCPAP-PC modes (HAMILTON-T1 & C1)Must be substantially equivalent to MAQUET Servo-I (K073179).Demonstrated substantial equivalence to MAQUET Servo-I (K073179). New ventilation modes were subjected to waveform performance testing, as described in ASTM F1100-90, and data was shown to be substantially equivalent to legally marketed devices.
    SpO2 monitoring (HAMILTON-T1 & C1)Must be substantially equivalent to the HAMILTON-G5 (K131774) and ensure SpO2 and pulse rate values are not corrupted during communication between OEM system and host device.Demonstrated substantial equivalence to HAMILTON-G5 (K131774). Testing demonstrated that SpO2 and pulse rate values calculated by the OEM system are not corrupted during communication to the HAMILTON-T1 or HAMILTON-C1 host device.
    Battery Duration (HAMILTON-T1)Maximal run time of 9.25 hrs (increased from 5.5 hrs).Achieved a maximal run time of 9.25 hrs.
    Battery Duration (HAMILTON-C1)Maximal run time of 4.30 hrs (increased from 2 hrs).Achieved a maximal run time of 4.30 hrs.
    Temperature Range (HAMILTON-T1)5 to 50°C (41 to 122°F) for adults/pediatrics (increased from 5 to 40°C).Components can withstand 50°C (122°F) in the adult/pediatric mode.
    Altitude Operating Condition (HAMILTON-T1)1013 to 376 hPa (25.000 ft) for adults/pediatrics (increased from 1013 to 600 hPa).Performance in high altitude has been increased for adult/pediatric patients to 376 hPa (25.000 ft).
    O2 consumption monitoring (HAMILTON-T1)Must be substantially equivalent to the Dräger Oxylog 3000 (K062267).Demonstrated substantial equivalence to the Dräger Oxylog 3000 (K062267).
    VOC and Particulate MatterOutput gas from the device must meet requirements for allowable levels of particulate matter for the most vulnerable patient population.Gas sample analysis comprising VOC and particular matter testing has demonstrated that the output gas from the device meets the requirements.
    General Safety and PerformanceCompliance with relevant FDA guidelines and international standards (e.g., ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 80601-2-12, IEC 60601-1-8, IEC 60601-1-6, IEC 62366, IEC 62304, ISO 80601-2-55). Software is a "major" level of concern.Testing demonstrated compliance with all listed standards and guidelines. Software verification and validation testing were conducted as per FDA guidance for "major" level of concern software.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document extensively references "bench testing" and "waveform performance testing" for the new ventilation modes as described in ASTM F1100-90. It also mentions "gas sample analysis" for VOC and particulate matter.

    • Sample Size: The document does not specify the sample size for the test sets used in these bench tests.
    • Data Provenance: The data appears to be from bench testing conducted by the manufacturer, Hamilton Medical AG, located in Bonaduz, Grisons, Switzerland. The nature of these tests (laboratory-based performance evaluation) suggests they are prospective in relation to the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The testing described (bench testing, waveform performance testing, gas sample analysis) is focused on technical performance metrics against specified standards and predicate devices, rather than clinical ground truth established by medical experts for a diagnostic or treatment outcome.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and therefore not provided in the document. The tests described are objective engineering and performance evaluations, not subjective assessments requiring adjudication of ground truth by multiple human reviewers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done. This document is for a ventilator, not an AI-assisted diagnostic or therapeutic device requiring human interpretation of output. Therefore, information on differences in human reader improvement with or without AI assistance is not applicable and not provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a ventilator, which is a therapeutic medical device, not an algorithm for diagnosis or interpretation. The document describes the performance of the integrated device and its new features. Therefore, the concept of "standalone (algorithm only)" performance as typically applied to AI/ML devices is not directly applicable here. The performance data presented refers to the functionality of the device itself.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the technical performance evaluations, the "ground truth" implicitly refers to the established standards (e.g., ASTM F1100-90), technical specifications, and the performance characteristics of the identified predicate devices. For example, for waveform performance, the waveforms generated by the new modes would be compared against expected waveforms defined by the standard and/or predicate device performance. For VOC and particulate matter, the ground truth is the "requirements for allowable levels" of these substances. This is not clinical ground truth derived from human subjects or pathology.

    8. The sample size for the training set

    The document describes a mechanical ventilator with hardware and embedded software, not a machine learning or AI algorithm that typically has a "training set." Therefore, the concept of a "training set sample size" is not applicable and not provided. The software mentioned (considered "major" level of concern) undergoes verification and validation, but this is distinct from machine learning model training.

    9. How the ground truth for the training set was established

    As the concept of a "training set" for an AI/ML algorithm is not applicable to this device, the method for establishing its "ground truth" is also not relevant and not provided. The software verification and validation would follow established software engineering principles and regulatory guidance, ensuring the software performs its intended functions correctly and safely.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120670

    Validate with FDA (Live)

    Device Name
    HAMILTON-T1
    Date Cleared
    2012-04-03

    (29 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics.

    Intended areas of use:

    • · In the intensive care ward or in the recovery room.
    • · For emergency medical care or primary care.
    • · During transport within and outside the hospital.
    • · During transfer by rescue vehicles, jet or helicopter.

    The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    Device Description

    The HAMILTON-T1 is designed for adults and pediatrics requiring invasive or noninvasive ventilation support. Due to its compact design, a fully-loaded weight of only 6.5 kg (14.3 lbs), a twin hot-swappable battery supply, and a built-in turbine; the HAMILTON-T1 can accompany a ventilated patient everywhere within the hospital or outside the hospital when transport is needed. The HAMILTON-T1 can run using AC or DC power. It does not need compressed air or O2 to drive the pneumatics, which reduces the weight load in the aircraft needed to operate the ventilator.

    Since the HAMILTON-T1 has been tested and evaluated for flight and high-altitude environments, it can be also used during patient transfer by emergency rescue vehicles, fixed-wing aircraft, or helicopter. This makes the HAMILTON-T1 especially relevant for Aeromedical Evacuations and Medevac operations.

    The HAMILTON-T1 ventilator uses the same graphical user interface (GUI) used by the HAMILTON-C2 and HAMILTON-G5, featuring a touchscreen "Ventilation Cockpit", this provides the exact information that the user needs and helps focus on what is important. In addition, the HAMILTON-T1 includes the ASV ventilation-mode which automatically applies lung-protective strategies, reduces the risk of operator error, and promotes early weaning.

    The HAMILTON-T1's microprocessor system controls gas delivery and monitors the patient. The gas delivery and monitoring functions are cross-checked by an alarm controller. This cross-checking helps prevent simultaneous failure of these two main functions and minimizes the possible hazards of software failure.

    The HAMILTON-T1 is intended as a transport ventilator, based on the existing HAMILTON-C2, with minor adaptations to make the HAMILTON-T1 capable of being used in highaltitude flight environments. The HAMILTON-T1's changes include the following:

      1. The HAMILTON-T1 software is identical to the HAMILTON-C2's software, except that some of the options are not available with the HAMILTON-T1, (e.g. Neonatal Ventilation & nCPAP-PS). Other features like Trends & Loops, NIV, NIV-ST, APRV, and DuoPAP are standard with the C2, but are only optional with the HAMILTON-T1.
      1. The HAMILTON-T1 has increased immunity from EMI, including >30 V/m. It also has extra safety features for the EMD, ESD, and RFI environments found on aircraft.
      1. The unit is contained within an impact resistant case which protects the controls from damage and inadvertent manipulation. The enclosure for the HAMILTON-T1 has been ruggedized to withstand shock, vibrations, water ingress, and drops from >1 meter heights.
      1. The HAMILTON-T1 was tested for use in fixed and rotary-wing aircraft.. Because mechanical ventilation can be challenging during air-medical transport, particularly due to the impact of changing barometric pressure with different altitude levels, the HAMILTON-T1 automatically compensates for altitude changes. Adjusting provided- and measured-patient volumes accordingly, thereby eliminating the need for manual calculation and reducing the risk of error.
      1. The HAMILTON-T1 has a "lock-button" which prevents an inadvertent change of settings. If screen lock is active, the following items are inactive: Touchscreen. Power/Standby switch, Print-screen key, Press-and-turn knob. Active are Alarm Silence, Manual Breath, O2 enrichment, Nebulizer. To switch off power, the user must press the On/Off button for > 3 s.
    AI/ML Overview

    Acceptance Criteria and Study for HAMILTON-T1 Ventilator

    This document describes the acceptance criteria and the non-clinical study that demonstrates the HAMILTON-T1 ventilator meets these criteria, as presented in the provided 510(k) summary (K120670).

    1. Table of Acceptance Criteria and Reported Device Performance

    The HAMILTON-T1 is essentially a modified HAMILTON-TC1 (K112006) for use in high-altitude flight environments. Therefore, its performance is largely measured against established standards and its predicate device. The acceptance criteria are derived from these recognized standards and confirmed through a battery of non-clinical tests.

    Acceptance Criteria CategorySpecific Criteria (Standard/Guidance Document)HAMILTON-T1 Performance (Reported in 510(k) Summary)
    Safety - GeneralIEC 60601-1: General Requirements for SafetyNon-clinical test results show device is safe.
    Safety - Electromagnetic Compatibility (EMI)IEC 60601-1-2: Electromagnetic CompatibilityIncreased immunity from EMI, including >30 V/m. Meets RTCA/DO-160F, Section 21, Category M (maximum level of radiated radio frequency interference).
    Safety - SoftwareIEC 60601-1-4: Programmable electrical medical systems; IEC 62304: Software life-cycle processesSoftware verification and validation testing demonstrates all specified requirements implemented correctly and completely. Hazard analysis and traceability analysis performed.
    Safety - AlarmsIEC 60601-1-8: Alarm SystemsNot explicitly detailed, but implied through general safety and predicate equivalence.
    Safety - Critical Care VentilatorsIEC 60601-2-12: Critical Care VentilatorsNot explicitly detailed, but implied through general safety and predicate equivalence, and its basis on HAMILTON-C2.
    UsabilityIEC 62366: Application of usability engineering to medical devices; AAMI/ANSI HE75: Human factors engineering. Design of medical devicesTouchscreen "Ventilation Cockpit" provides exact information, helps focus. "Lock-button" prevents inadvertent changes.
    Risk ManagementEN ISO 14971: Application of risk management to medical devicesHazard analysis performed.
    Airborne Environmental ConditionsRTCA/DO-160F (Environmental Conditions and Test Procedures for Airborne Equipment), specific sections: 7 (Normal surge voltage), 8 (Low/Abnormal surge voltage, Voltage spikes), 16.6 (DC input power leads), 17 (Vibration), 18.3.1 (Operational Shocks and Crash Safety), 19.3.1 (Abnormal operating conditions), 20 (Magnetic field induced into equipment), 21 (Radio Frequency Susceptibility - Radiated & Conducted), 25 (Electrostatic Discharge & Radiated Electromagnetic Field)Tested and evaluated for flight and high-altitude environments. Meets or exceeds all specified RTCA/DO-160F sections. Automatically compensates for altitude changes.
    Road AmbulancesEN 1789: Medical vehicles and their equipment - Road ambulancesUnit ruggedized to withstand shock, vibrations, water ingress, and drops.
    Air AmbulancesEN 13718-1: Medical vehicles and their equipment - Air ambulances - Part 1: Requirements for medical devices used in air ambulancesTested for use in fixed and rotary-wing aircraft. Automatically compensates for altitude changes.
    Mechanical Performance (Waveform)ASTM F1100-90: Standard Specification for Ventilators Intended for Use in Critical Care (for waveform standard analysis)Ventilator subjected to waveform performance testing.
    Ruggedness/DurabilityEN 794-3: Particular requirements for emergency and transport ventilators (Equal to EN 60068-2-6; -29; -64)Unit contained in impact resistant case, ruggedized to withstand shock, vibrations, water ingress, and drops from >1 meter heights.
    Input PowerIEC 62133: Battery Safety. Non-Spillable.Twin hot-swappable battery supply. Runs using AC or DC power.
    Software EquivalenceN/ASoftware is identical to HAMILTON-C2's software, with some options unavailable.
    Predicate EquivalenceN/AIntended use, technological characteristics, and performance specifications are equivalent to HAMILTON-TC1 (K112006).

    2. Sample Sizes Used for Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a number of devices. The "test set" refers to the HAMILTON-T1 device itself being subjected to various non-clinical tests.
    • Data Provenance: The testing was conducted by HAMILTON MEDICAL AG, a Swiss company. The tests are non-clinical, involving the device's physical and software performance under simulated conditions as per internationally recognized standards (e.g., IEC, ISO, RTCA/DO-160, ASTM, EN, MIL-STD). This is a prospective evaluation of the manufactured device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable for this submission. The "ground truth" for non-clinical performance refers to the established requirements and specifications outlined in the referenced standards (e.g., a certain level of EMI immunity, specific waveform characteristics). Compliance is measured against these objective, quantifiable standards, not against expert human interpretations of clinical data. Testing personnel performing these evaluations would be qualified engineers and technicians.

    4. Adjudication Method for the Test Set

    This section is not applicable for a non-clinical performance study. The adjudication method as described typically refers to how disagreements among human readers or expert reviewers are resolved in a clinical study where subjective interpretation is involved. Non-clinical tests against objective standards do not require such adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses on non-clinical performance and equivalence to a predicate device, not on diagnostic or treatment efficacy evaluated through human-in-the-loop performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation of the device's engineering and software attributes was done through the extensive non-clinical testing described. This includes:

    • Software verification and validation testing: To ensure specified requirements are implemented correctly and completely.
    • Waveform performance testing: As described in ASTM F1100-90.
    • EMI/EMC testing: Against IEC 60601-1-2 and RTCA/DO-160F.
    • Environmental stress testing: For shock, vibration, water ingress, drops, and high-altitude/low-pressure conditions.
    • Safety testing: Against a range of IEC 60601-1 series standards.

    These tests evaluate the device's intrinsic functional and safety characteristics without direct human intervention in its operation for a clinical outcome.

    7. The Type of Ground Truth Used

    The ground truth used for this submission is engineering specifications and recognized international consensus standards. These include:

    • Performance Specifications: Quantifiable parameters like FiO2 range, Tidal Volume, Peak Flow, which are compared against known requirements for ventilators.
    • Safety Standards: Such as IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-8, IEC 60601-2-12, IEC 62304, IEC 62366, EN ISO 14971.
    • Environmental and Transportation Standards: Such as RTCA/DO-160F, EN 794-3, EN 1789, EN 13718-1.
    • Waveform Standards: ASTM F1100-90.

    The device's performance in designated tests is measured against the limits and criteria defined by these standards.

    8. The Sample Size for the Training Set

    This refers to a machine learning context. Since the HAMILTON-T1 device is not described as utilizing a machine learning algorithm that requires a "training set" of data, this section is not applicable. The device's software is based on the HAMILTON-C2's software and verified through traditional software validation methods, not machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, this section is not applicable as there is no "training set" in the context of this device's development or evaluation.