Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K153046

    Validate with FDA (Live)

    Device Name
    HAMILTON-MR1
    Date Cleared
    2016-09-06

    (323 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HAMILTON-MR1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.
    Intended areas of use:

    • In the MRI department
    • In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
    • During transfer of ventilated patients within the hospital
      The HAMILTON-MR1 ventilator is classified as MR Conditional with the use of 1.5 Tesla and 3.0 Tesla static magnetic field scanners.
      The HAMILTON-MR1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
    Device Description

    The HAMILTON-MR1 is an MR-Conditional ventilator which increases the availability of appropriate modes of therapy for ventilated hospital patients requiring MR imaging. It is designed for adult, pediatric, infant, and neonatal patients requiring invasive or noninvasive ventilation support. It covers a full range of clinical requirements, including invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and noninvasive ventilation. It can be used at the gauss line, in the presence of either 1.5T or 3T MRI machines.
    With the shielded, MRI-compatible HAMILTON-MR1 ventilation performance and MR image quality is as safe and as effective as the predicate device throughout the MRI procedure. The integrated gaussmeter is programmed to alarm when the clinician is placing the HAMILTON-MR1 too close to the MRI magnet, which helps the clinician to properly position the HAMILTON-MR1 at the 50mT (500 gauss) line or less.
    The 510(k) submission intends to add the following new features to the previously cleared ventilator HAMILTON-MR1:

    • Neonatal patients with a minimum weight of 0.2 kg and a minimal tidal volume of 2 mL
    • . The following two new modes for the neonatal patient group: nCPAP and nCPAP-PC
    • An increase in the battery duration from 5 hours and 20 minutes to 9 hours and 20 minutes.
    AI/ML Overview

    The provided document describes a 510(k) submission for the HAMILTON-MR1 ventilator, focusing on its substantial equivalence to previously cleared predicate devices. The submission concerns adding new features, specifically the capability to ventilate neonatal patients and two new ventilation modes (nCPAP and nCPAP-PC).

    This document does not include information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML-driven medical device, as it is a ventilator and not an AI/ML diagnostic tool. Therefore, many of the requested items (e.g., sample size for test/training sets, number of experts for ground truth, MRMC studies, standalone AI performance) are not applicable to this type of device submission.

    However, I can extract information regarding performance testing and regulatory compliance relevant to this specific device.


    Acceptance Criteria and Device Performance (Based on the provided document)

    Since this is a ventilation device rather than an AI/ML diagnostic, "acceptance criteria" here refer to compliance with relevant medical device standards and demonstrating functionality for the newly added features.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied by Regulatory Compliance & Testing)Reported Device Performance (Summary from Document)
    Safety and Essential Performance Standards: - ANSI/AAMI ES60601-1 (2005/(R) 2012): Medical electrical equipment General — Requirements for SafetyCompliant: Testing demonstrated compliance with this standard.
    - IEC 60601-1-2 (2007): Electromagnetic CompatibilityCompliant: Testing demonstrated compliance with this standard.
    - ISO 80601-2-12 (2011): Critical Care Ventilators Specific RequirementsCompliant: Testing demonstrated compliance with this standard.
    - IEC 60601-1-8 (2006 + Am.1: 2012): Alarm SystemsCompliant: Testing demonstrated compliance with this standard.
    - IEC 60601-1-6 (2010 + A1 :2013): UsabilityCompliant: Testing demonstrated compliance with this standard.
    - IEC 62366 (2008)+A1(2014): Usability EngineeringCompliant: Testing demonstrated compliance with this standard.
    - IEC 62304 (2006): Medical Device Software Life-Cycle ProcessesCompliant: Software design and validation process, together with bench testing, demonstrated the device operates as intended. Additional software verification and validation conducted as per FDA guidance. Software classified as "major" level of concern.
    - ISO 80601-2-55 (2011): Respiratory Gas Monitors Specific RequirementsCompliant: Testing demonstrated compliance with this standard.
    Added Features Functionality: - Neonatal patient compatibility (min. 0.2 kg, min. 2 mL tidal volume)Demonstrated Functionality: Covered by the overall testing of modified HAMILTON-MR1.
    - New modes nCPAP and nCPAP-PC performance for neonatal useFunctionality Demonstrated & Substantially Equivalent: New ventilation modes subjected to waveform performance testing. Data provided showed substantial equivalence to legally marketed devices (specifically, the HAMILTON-C1/T1 which already has these modes and neonatal capability).
    Extended Battery Duration: - Increase from 5 hours 20 minutes to 9 hours 20 minutesDemonstrated: Battery duration increased to 9 hours 20 minutes (with typical run time stated as 8 hours).
    MR-Conditional Performance: - Does not generate artifacts in MRI - Not negatively influenced by magnetic fields (1.5T and 3.0T static magnetic fields)Verified: MRI environment test for RF-Noise redone to verify no artifacts. System test for magnetic immunity redone to verify no negative influence by magnetic field. The device is classified as MR Conditional with the use of 1.5 Tesla and 3.0 Tesla static magnetic field scanners, and the integrated gaussmeter alarms when too close to the MRI magnet (50mT line or less).
    Material Biocompatibility: - Biocompatibility of new materialsConfirmed: Only biocompatible materials or materials already used in earlier 510(k)s were introduced; no additional testing conducted (implying prior acceptance).

    Study Details (Contextualized for a Ventilator Submission):

    2. Sample size used for the test set and the data provenance:

    • This document does not specify a "test set" in terms of patient data or clinical trials, as it's a 510(k) for a ventilator modification based on bench testing, engineering verification, and comparison to predicate devices, rather than a clinical performance study involving human subjects.
    • The data provenance is from bench testing and software verification/validation activities conducted in support of the submission. The company is Hamilton Medical AG, located in Bonaduz, Grisons, Switzerland. The testing is implied to be prospective as it was conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k) submission for a ventilator. Ground truth as typically defined for AI/ML diagnostic devices (e.g., expert consensus on medical images) is not relevant here. The "ground truth" for this device's performance is its adherence to engineering specifications and international standards, demonstrated via bench testing and engineering verification.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This relates to clinical study design for image interpretation or diagnosis, which is not part of this ventilator submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a ventilator, not an imaging device or an AI assistant for human readers. MRMC studies are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is a ventilator, a physical machine that provides mechanical ventilation. Its "performance" is its ability to deliver specified ventilation parameters safely and effectively, not an algorithm's diagnostic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this ventilator device is its compliance with internationally recognized medical device standards (e.g., IEC, ISO, ANSI/AAMI), its engineering specifications, and its demonstrated performance in bench testing (e.g., waveform performance testing for new ventilation modes, magnetic immunity, RF-noise generation). The comparison to predicate devices also serves as a baseline for "truth" in terms of substantial equivalence.

    8. The sample size for the training set:

    • Not applicable. This device does not use an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set is used for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122438

    Validate with FDA (Live)

    Device Name
    HAMILTON-MR1
    Date Cleared
    2013-11-20

    (467 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HAMILTON-MR1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics.

    Intended areas of use:

    • · In the MRI department
    • · In the intensive care ward or in the recovery room
    • · During transfer of ventilated patients within the hospital

    The HAMILTON-MR1 ventilator is classified as MR Conditional with the use of 1.5 T and 3.0 Tesla static magnetic field scanners.

    The HAMILTON-MR1 ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    Device Description

    The HAMILTON-MR1 is an MR-Conditional ventilator which increases the availability of appropriate modes of therapy for ventilated hospital patients requiring MR imaging. It covers a full range of clinical requirements such as: invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and non-invasive Ventilation (NIV). It can be used at the 500 gauss line, in the presence of either 1.5T or 3T magnets.

    In the MRI environment, where strong magnetic fields pose a danger for both the patient and the operator, safety is the highest priority. With the effectively shielded, MRI-compatible HAMILTON-MR1 ventilator, ventilation performance and MR image quality is guaranteed throughout the procedure. The integrated gaussmeter is programmed to alarm when the clinician is placing the HAMILTON-MR1 too close to the MRI magnet, which helps the clinician to properly position the HAMILTON-MR1 at the 50mT (500 gauss) line or less.

    The HAMILTON-MR1 can be used in ICU special care areas, cardiac surgery recovery rooms. step-down or sub-acute care units, and when transporting patients to the MRIdepartment. In these cases, the HAMILTON-MR1 guarantees uncompromised continuous ventilation care from the ICU to the MRI and back with its 6-hour long batteries. Alternatively, the HAMILTON-MR1 can be used as an MRI-proprietary ventilator, waiting in the MRI department for the patient. In the MR environment -- when the clinician is unable to stay close to the patient for routine adjustments -- the ASV mode adapts to the patient's lung condition.

    Positioning a medical device too close to the MRI can have fatal consequences and cause serious injury to the patient or clinician. In addition, significant financial losses can occur if an MRI shutdown is required. The HAMILTON-MR1's integrated gaussmeter continuously monitors the magnetic field and gives the clinician both an audible and a visual signal if the HAMILTON-MR1 is getting too close to the MRI magnet. For increased MRI safety and ease of use in the MR environment, the integrated gaussmeter continues monitoring -- even when the ventilator is not in use. Close proximity of the ventilator to the MRI machine is crucial. The HAMILTON-MR1 is a ventilator able to be used at a magnetic field strength of 500 gauss, without creating any MR image artifacts.

    Both the HAMILTON-MR1's software and ventilation modes are identical to the HAMILTON-C1. One can operate the HAMILTON-MR1 with the touch-screen or with a single-turn wheel. Hard keys give direct access to the most important functions. The two devices are identical with the exception of modifications related to the MR environment (e.g. integrated gaussmeter, special alarm indicators, MR-specific labeling, the reduction in ferromagnetic materials, the removal of the options board, and increased safety measures for battery removal or replacement). Due to these modifications, the HAMILTON-MR1 is able to withstand the challenging conditions found in the MR environment. With the large alarm lamp, a clinician can immediately identify an alarming HAMILTON-MR1 ventilator because the alarm lamp is located at the top of the device, even if the clinician is a long distance away or when several different devices are operating simultaneously in the same room. The high-performance turbine can deliver up to 210 L/min flow, which is potentially helpful when using NIV modes of ventilation. The HAMILTON-MR1 includes a trolley made out of nonferrous materials, which will not be attracted to the powerful electromagnetic forces emanating from the MRI's magnet. The trolley also has a "fail-safe" braking system (i.e., Dead man's brake).

    AI/ML Overview

    The provided text describes the Hamilton-MR1 ventilator and its compliance with various standards and its suitability for use in an MRI environment. However, it does not explicitly detail a study in the format of a clinical trial or a comparative effectiveness study with human readers that proves the device meets specific acceptance criteria.

    The information primarily focuses on non-clinical performance testing and substantial equivalence to a predicate device (HAMILTON-C1) and other MR-compatible ventilators.

    Here's an attempt to extract and organize the information based on your request, with a note on where the requested information is not available in the provided text:


    Acceptance Criteria and Device Performance for HAMILTON-MR1 Ventilator

    The acceptance criteria are generally implied through compliance with various international standards and specific performance thresholds related to MRI compatibility. The device performance is reported through non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Target/Threshold)Reported Device Performance (HAMILTON-MR1)Reference/Standard
    MRI CompatibilityMagnetic Field Threshold (Operation)Green Light - Acceptable: $\leq$ 50mT (500 gauss) Yellow Light - Too close: 50 - 80mT Red Light - Too close: > 80mTInternal testing, Philips Achieva 3.0T and 1.5T MRI scanners
    Gaussmeter AccuracyIntegrated gaussmeter accuracy: $\pm$ 10%Comparison with commercially available, calibrated third-party gaussmeter
    Functionality at Higher Gauss LevelsCan function "in accordance to its 'essential performance criteria'" at > 100mT (1,000 gauss) (though not recommended for use > 50mT)Non-clinical performance bench-testing (Philips Achieva 3.0T and 1.5T MRI scanners)
    Magnetically Induced TorquePulled towards MRI magnet at $\geq$ 445mT (4,450 gauss)ASTM F2213-06 (Magnetically Induced Torque on Medical Devices in the MR Environment)
    Magnetically Induced Displacement ForceForce of 64.2 Newtons at $\geq$ 445mT (4,450 gauss) when brake activatedASTM F2052-06 (Magnetically Induced Displacement Force on Medical Devices in the MR Environment)
    MR Image ArtifactsNo MR image artifacts created at 500 gaussASTM F2119-07 (Evaluation of MR Image Artifacts from Passive Implants)
    Distance to MRI Bore Opening$\sim$1 m (3.25 ft) at 50mT (500 gauss)Comparison to other MR-compatible ventilators
    General PerformanceTidal Volume (minimal)$\geq$ 20 mLComparison to other MR-compatible ventilators
    Maximum Inspiratory Flow210 L/minComparison to predicate device (HAMILTON-C1)
    Maximum Working Pressure Limit60 cmH20Comparison to predicate device (HAMILTON-C1)
    EnvironmentalOperation Temperature Range5 to 40° C (41 to 104 °F)Comparison to predicate device (HAMILTON-C1) and IEC 60601-1
    Operating Humidity10 to 95%, non-condensingComparison to predicate device (HAMILTON-C1)
    Operating Atmospheric Pressure1100 to 700 hPaComparison to predicate device (HAMILTON-C1)
    Software/SafetySoftware Verification/ValidationSoftware passed verification/validation tests. Complete revision level history, hazard analysis, and traceability analysis. Risk management report, system-level validation, verification testing according to applicable standards.Internal software development processes, IEC 62304, IEC 60601-1-4, IEC 60601-1-8, ISO 14971, IEC 60601-1-6, FDA Draft Reviewer Guidance for Ventilators (1995)
    Battery SafetyNon-spillableIEC 62133

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test set sample size: Not specified. The testing described is primarily non-clinical bench testing with specific MRI scanners (Philips Achieva 3.0T and 1.5T). The "test set" in this context refers to the device prototypes/units physically tested.
    • Data provenance: The document indicates testing was conducted using Philips Achieva 3.0T and 1.5T MRI scanners, and comparison with commercially available third-party gaussmeters. The location of these tests is not explicitly stated, but the manufacturer is Hamilton Medical AG based in Switzerland. Therefore, the data provenance is likely European-based non-clinical lab testing. The testing is non-clinical/bench testing, implying it is prospective testing of the device's physical properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of experts: Not applicable for the described non-clinical performance and safety testing. The "ground truth" for these tests relates to physical measurements and compliance with engineering standards, not expert medical opinion on images or patient outcomes.
    • Qualifications of experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication method: Not applicable. The tests described are objective measurements (e.g., magnetic field strength, displacement force, artifact presence/absence based on established protocols), not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: No, a multi-reader multi-case comparative effectiveness study was not done. The document focuses on the ventilator's physical and functional compatibility with the MRI environment, not on AI assistance for human readers or interpretation of medical images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone performance: No, this is not applicable. The HAMILTON-MR1 is a physical medical device (ventilator) with integrated software and hardware components, not an AI algorithm for diagnostic purposes. Its performance is evaluated as a standalone device in terms of its operational capabilities and safety in an MRI environment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of ground truth: The ground truth for the non-clinical tests is based on defined physical parameters and measurements (e.g., magnetic field strength, force, temperature, flow rates), and compliance with standardized testing protocols (e.g., ASTM standards for MRI compatibility, IEC standards for medical electrical equipment). For the integrated gaussmeter, the ground truth was established by comparison with a "commercially available, calibrated third-party gaussmeter."

    8. The sample size for the training set

    • Training set sample size: Not applicable. The HAMILTON-MR1 is a medical device, not an AI model trained on a dataset. The software undergoes verification and validation, but this is distinct from machine learning training.

    9. How the ground truth for the training set was established

    • Ground truth for training set: Not applicable, as there is no "training set" in the context of an AI model for this device. Software verification and validation are typically based on predefined requirements, specifications, and established testing methodologies to confirm the software performs as intended.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1