    K Number
    K251041

    Validate with FDA (Live)

    Date Cleared
    2025-10-31

    (211 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GripTract-GI Endoscopic Tissue Manipulator (GripTract) Upper GI Model is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    GripTract Upper GI Model is indicated for use in the upper gastrointestinal tract with any standard endoscope as follows:

    Endoscope Distal Tip Outer Diameter (mm)Endoscope Working Length (cm)GripTract Model#
    9.8 - 10.5103 - 110GT-GS100
    Device Description

    The GripTract-GI™ Endoscopic Tissue Manipulator (GripTract) Upper GI Model GT-GS100 is a disposable, non-sterile accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    The GripTract Handpiece attaches to the endoscope control body just below the working channel. The soft End Cap with two integrated Fingers for tissue manipulation is placed on the distal end of the endoscope. Two Control Knobs on the Handpiece each operate a corresponding Finger, permitting the User to rotate and extend/retract the Fingers for tissue manipulation and visualization independent of the endoscope's movement or the presence of other tools in the endoscope's working channel.