510(k) Data Aggregation

    K Number
    K150299

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-19

    (286 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gmate® SMART Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, palm, thigh or calf. The Gmate® SMART Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The Gmate® SMART Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Gmate® SMART should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternate site testing should be done only during steady state times (when glucose is not changing rapidly).

    The Gmate® Blood Glucose Test Strips are with the GMATE® SMART meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf.

    The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management.

    Device Description

    The Gmate® SMART Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood. The Gmate® SMART System is intended for self-testing outside the body (in vitro diagnostic use only) by people with diabetes at home as an aid to monitor the effectiveness of their diabetes management. The Gmate® SMART Blood Glucose Monitoring System consists of a glucose meter, test strips, and one control material (additional levels of control material are available upon request). The small Gmate® SMART meter does not require coding or calibration, no need of batteries, and no settings are required. The Gmate® SMART meter is powered on by plugging it into the headphone jack of the Apple/Android device. Insert the test strip into the meter, apply the blood or control solution to the strip and the meter will begin the 5 seconds count down of your test result. The Gmate™ SMART App converts the signal generated from the meter and test strip and displays the test result on the Apple/Android device. The Gmate® SMART Blood Glucose Monitoring System uses the smartphone technology, currently with Apple's iOS (with use of the Apple iPhone 3GS, iPhone 4, iPhone 4S, iPhone 5, iPod Touch 4th generation, iPad, and iPad2) and now with the following Android devices: Samsung Galaxy S3, S4, and S5; to view glucose test results. A simple download of the Gmate® SMART App, enables use of many functions. The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider or healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management. The test principle is: This device is an in vitro diagnostic only product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the glucose oxidase that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. The meter measures the current and calculates your blood glucose level. Combined with the Gmate™ SMART App, it displays the test result and stores them on your Apple/Android device.

    AI/ML Overview

    Here’s an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state quantitative acceptance criteria in a table format for accuracy. However, it does state an overall conclusion: "The results demonstrated that the Gmate® SMART met all the reliability requirements and performance claims." and "Based on the comparisons completed of the clinical and non-clinical tests performed, the devices passed all of the tests based on pre-determined Pass/Fail Criteria."

    To infer the acceptance criteria for accuracy, we typically refer to ISO 15197 for blood glucose monitoring systems. While not directly stated, the study implies adherence to these standards by mentioning "FDA Guidance Documents and CLSI reference standards." The YSI 2300 Glucose Analyzer is the widely accepted reference method for accuracy in these types of studies. The performance is reported as meeting "all the reliability requirements and performance claims" and "passed all of the tests based on pre-determined Pass/Fail Criteria."

    Performance MetricAcceptance Criteria (Inferred from regulatory standards)Reported Device Performance
    System AccuracyTypically based on ISO 15197 (e.g., >95% of results within ±15 mg/dL or ±15% of reference for specific glucose ranges)"Met all the reliability requirements and performance claims" and "passed all of the tests based on pre-determined Pass/Fail Criteria."
    PrecisionNot explicitly defined, but implied by "passed all of the tests""Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    LinearityNot explicitly defined, but implied by "passed all of the tests""Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    InterferenceNot explicitly defined, but implied by "passed all of the tests""Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    HematocritNot explicitly defined, but implied by "passed all of the tests""Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    AltitudeNot explicitly defined, but implied by "passed all of the tests""Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    Temperature/HumidityNot explicitly defined, but implied by "passed all of the tests""Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    User PerformanceSuccessful completion of tests by users"Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"
    Alternate Site TestingSuccessful completion of tests"Evaluations... conducted to establish the performance, the functionality and the reliability characteristics"

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: "Glucose levels were measured on 100 persons with diabetes and healthcare professionals at a clinic center."
    • Data Provenance: The document does not explicitly state the country of origin where the clinical study took place. It mentions Philosys, Inc. has a New York Office and a Seoul Office (South Korea). It also states "clinical tests... were performed in accordance with FDA Guidance Documents and CLSI reference standards," which suggests the study was designed to meet US regulatory requirements. The study appears to be prospective as it involves performance evaluations on individuals, rather than analyzing existing datasets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • The ground truth was established using the YSI 2300 Glucose Analyzer, a laboratory instrument.
    • The document mentions "healthcare professionals at a clinic center" were involved in measuring glucose levels, but it doesn't specify the number or specific qualifications of these professionals for establishing the ground truth. Their role was likely to operate the YSI 2300 and collect samples, not to interpret or adjudicate results in the way an "expert" would for image-based diagnostics.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • No adjudication method (like 2+1 or 3+1 consensus) is described. This is expected for a blood glucose monitoring system where the YSI 2300 Glucose Analyzer itself serves as the objective reference standard, not human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This type of study is typically relevant for interpretative devices (e.g., radiology AI) where human readers make diagnoses. The Gmate® SMART is a quantitative measurement device, not an interpretative one in the sense of an MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance evaluation was done. The device's accuracy was assessed by comparing its glucose readings directly against a laboratory reference instrument (YSI 2300 Glucose Analyzer). While users operate the device, the core performance being evaluated is the algorithm/system's ability to accurately measure glucose, independent of human interpretive bias. The study also included "user performance" which might incorporate human interaction, but the primary accuracy assessment is standalone.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth used was reference laboratory measurement from a YSI 2300 Glucose Analyzer. This is considered a highly accurate and objective standard for glucose concentration.

    8. The sample size for the training set:

    • The document does not provide information about a training set. This is common for this type of device, where analytical and clinical performance studies are conducted, rather than machine learning model training in the typical sense. The device's underlying measurement principle is electrochemical, not a learned algorithm requiring a separate training set.

    9. How the ground truth for the training set was established:

    • As no training set is mentioned or implied, this question is not applicable based on the provided text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131230

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-04

    (461 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gmate® SMART Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf. The Gmate® SMART Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The Gmate® SMART Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Gmate® SMART should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternate site testing should be done only during steady state times (when glucose is not changing rapidly).

    The Gmate® Blood Glucose Test Strips are for use with the GMATE® SMART Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf.

    The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management.

    Device Description

    The Gmate® SMART Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood. The Gmate® SMART System is intended for self-testing outside the body (in vitro diagnostic use only) by people with diabetes at home as an aid to monitor the effectiveness of their diabetes management. The Gmate® SMART Blood Glucose Monitoring System consists of a glucose meter, test strips, and one control material (additional levels of control material are available upon request). The small Gmate® SMART meter does not require coding or calibration, no need of batteries, and no settings are required. The Gmate® SMART meter is powered on by plugging it into the headphone iack of the Apple device. Insert the test strip into the meter, apply the blood or control solution to the strip and the meter will begin the 5 seconds count down of your test result. The Gmate™ SMART App converts the signal generated from the meter and test strip and displays the test result on the Apple device. The Gmate® SMART Blood Glucose Monitoring System uses the smartphone technology, currently Apple's iOS (with use of the Apple iPhone 3GS, iPhone 4, iPhone 4S, iPhone 5, iPod Touch 4tt generation, iPad, and iPad2), to view glucose test results. A simple download of the Gmate® SMART App, enables use of many functions. The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider or healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management. The test principle is: This device is an in vitro diagnostic only product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the glucose oxidase that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. The meter measures the current and calculates your blood glucose level. Combined with the Gmate™ SMART App, it displays the test result and stores them on your Apple device.

    AI/ML Overview

    Acceptance Criteria and Study for Gmate® SMART Blood Glucose Monitoring System

    This document describes the acceptance criteria and the study performed for the Gmate® SMART Blood Glucose Monitoring System, based on the provided 510(k) summary (K131230).

    1. Acceptance Criteria and Reported Device Performance

    The provided document states that the performance evaluations were conducted to establish the performance, functionality, and reliability characteristics of the Gmate® SMART Blood Glucose System, and "The devices passed all of the tests based on pre-determined Pass/Fail Criteria." However, the specific quantitative acceptance criteria (e.g., specific accuracy percentages for different glucose ranges) are not explicitly detailed in the provided text. It refers to compliance with ISO15197 and other reference standards, which would define these criteria.

    Based on the general requirements of ISO 15197, for a blood glucose monitoring system, the acceptance criteria typically specify the percentage of results that must fall within certain accuracy ranges when compared to a laboratory reference method. A common set of criteria from ISO 15197:2003 (which was likely relevant in 2013/2014) is:

    • For glucose concentrations < 75 mg/dL (4.2 mmol/L): at least 95% of the measured glucose values shall fall within ± 15 mg/dL (± 0.83 mmol/L) of the reference method values.
    • For glucose concentrations ≥ 75 mg/dL (4.2 mmol/L): at least 95% of the measured glucose values shall fall within ± 20% of the reference method values.

    Since the document states the device "passed all of the tests based on pre-determined Pass/Fail Criteria" and was performed in accordance with ISO15197, we infer that the device performance met the criteria of ISO15197. Without the exact quantitative data presented in the summary, we cannot create a detailed table of reported performance against specific numerical acceptance criteria.

    Therefore, a table of acceptance criteria and reported device performance cannot be generated with specific numerical values from the provided text. The document only broadly states that the device met the pre-determined Pass/Fail Criteria based on ISO15197.

    2. Sample Size and Data Provenance for the Test Set

    The 510(k) summary does not specify the exact sample size used for the clinical test set, nor does it explicitly state the country of origin or whether the study was retrospective or prospective. It generally refers to "Non-clinical and clinical tests."

    3. Number and Qualifications of Experts for Ground Truth

    The document does not provide information regarding the number of experts used to establish the ground truth for the test set or their specific qualifications. For blood glucose monitoring systems, the "ground truth" is typically established by laboratory reference methods (e.g., using a YSI glucose analyzer) rather than expert consensus on visual interpretation.

    4. Adjudication Method

    The document does not mention an adjudication method, which is generally not applicable for performance studies of blood glucose meters where comparisons are made against a laboratory reference method rather than subjective expert interpretations.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    An MRMC study is not relevant for this type of device (blood glucose monitoring system). Blood glucose meters are assessed for their accuracy in quantifying glucose levels, not for improved human interpretation or diagnostic capabilities with AI assistance. Therefore, no information on an MRMC study or effect size of AI assistance for human readers is provided or expected.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The "Non-clinical and clinical tests for the Gmate® SMART Blood Glucose System" evaluated the device's performance, functionality, and reliability. This would inherently be a standalone evaluation of the algorithm and device's measurement capabilities without human-in-the-loop performance being a primary metric. The system's ability to accurately measure glucose and display results on the connected Apple device constitutes its standalone performance.

    7. Type of Ground Truth Used

    The ground truth for the clinical performance evaluation (specifically system accuracy) would be laboratory reference method measurements of blood glucose. While not explicitly stated, this is the standard for evaluating blood glucose monitoring systems against ISO 15197. These reference measurements are typically obtained using highly accurate laboratory instruments (e.g., a YSI analyzer).

    8. Sample Size for the Training Set

    The document does not mention a "training set" or "training data" in the context of device development. Blood glucose monitors primarily use a fixed algorithm and calibration rather than machine learning models that require a distinct training set. The calibration and factory settings would be established during manufacturing based on extensive testing and quality control processes.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a "training set" for a machine learning model is not explicitly detailed or indicated for this device. The device's underlying "algorithm" (the conversion of the electrochemical signal to a glucose reading) is based on established electrochemical principles and calibration. The ground truth for establishing and verifying this algorithm/calibration would be through laboratory reference method measurements using standard glucose solutions and blood samples across the device's measurement range, similar to how the clinical test set's ground truth is established.