Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202885

    Validate with FDA (Live)

    Date Cleared
    2021-09-15

    (352 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GlucoSure HT Plus Blood Glucose Monitoring System is intended to quantitatively measure blood glucose in fresh capillary whole blood drawn from fingertips, palm, or forearm. Alternative site testing for glucose test should be performed only during steady-state (when glucose is not changing rapidly). It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control and should only be used by a single patient and it should not be shared. It is not intended for diagnosis or screening of diabetes or for neonatal use.
    The GlucoSure HT Blood Glucose Test Strips are to be used with the GlucoSure HT Plus Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm.

    Device Description

    The GlucoSure HT Plus blood glucose monitoring system consists of the GlucoSure HT Plus meter and GlucoSure HT Test Strips. It is used for testing of blood glucose by self-testers at home.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the GlucoSure HT Plus Blood Glucose Monitoring System, which is a medical device for measuring blood glucose. The information necessary to fully answer your request regarding acceptance criteria and study proving device meets criteria (especially as it pertains to AI/algorithm performance) is not fully detailed in this document, as it focuses on the substantial equivalence to a predicate device for a glucose monitoring system, and not an AI-assisted diagnostic tool in the typical sense this question implies.

    However, based on the provided text, I can infer and extract some relevant information as best as possible, particularly regarding the clinical testing and the overall performance evaluation for this type of device.

    Key takeaway for AI-related questions: This document describes a blood glucose monitoring system, not an AI-based diagnostic device where radiologists or extensive adjudication methods for image analysis would typically apply. The "algorithm" mentioned refers to the measurement algorithm within the glucose meter, not necessarily a complex AI/ML algorithm for image interpretation.

    Here's a breakdown based on the available text:

    Acceptance Criteria and Device Performance (Inferred from device type and general FDA requirements for glucose meters):

    For a blood glucose monitoring system, the primary acceptance criteria revolve around the accuracy of glucose measurements compared to a reference method. While specific numerical acceptance criteria (e.g., % within 15% range) are not explicitly stated in the provided summary, these are standard for blood glucose meters. The summary states:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred for Blood Glucose Meters)Reported Device Performance (as stated in summary)
    Accuracy of Glucose Measurement: (Typically assessed against a laboratory reference method, with specific percentages of measurements falling within predefined error margins, e.g., >95% within ±15 mg/dL for <100 mg/dL and ±15% for ≥100 mg/dL)"Clinical and analytical testing demonstrated that the GlucoSure HT Plus Blood Glucose Monitoring System perform in a substantially equivalent manner to that of the predicate."
    Ease of Use for Home Users: (User comprehension, ability to perform test correctly, no significant errors due to device complexity)"User studies for glucose was conducted with home users, including evaluation of ease of use and ease of understanding of the user manual. Results demonstrate substantial equivalence to the predicate system."
    Stability (e.g., aluminum foil, battery life): (Maintain performance over time and under expected conditions)"Battery life test, aluminum foil stability test were also done, and results demonstrate substantial equivalence to the predicate system."
    Software Verification and Validation: (Software functions as intended, no critical errors)"Software verification and validation were done."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not explicitly state the specific number of participants or samples used in the "user studies for glucose" (clinical testing) or "analytical testing."
    • Data Provenance: The studies were conducted by Apex BioTechnology Corp. No specific country of origin for the data (beyond the applicant's location in Taiwan) is mentioned. The studies appear to be for regulatory submission, implying they are prospective in nature, designed specifically to validate the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This question is more applicable to diagnostic devices interpreted by human experts (like radiologists for imaging). For a blood glucose monitoring system, the "ground truth" for glucose levels is typically established by laboratory reference methods (e.g., hexokinase method on a clinical chemistry analyzer), not by human experts interpreting data. The text does not mention human experts establishing ground truth for glucose values.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable in the context of a blood glucose monitoring system where ground truth is established by objective laboratory methods. Adjudication methods like 2+1 or 3+1 are used for expert consensus on subjective interpretations (e.g., reviewing medical images).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a standalone blood glucose meter, not an AI-assisted diagnostic tool that humans (readers) would use to interpret medical cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a "standalone" measurement system. The "algorithm" mentioned (likely the embedded algorithm within the meter that processes the chemical reaction on the test strip to produce a glucose reading) performs without human interpretation in the loop beyond the user performing the test. The "analytical testing" would assess this standalone performance against a reference method.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Based on common practices for blood glucose meters, the ground truth would typically be laboratory reference standard glucose measurements obtained from clinical chemistry analyzers (e.g., YSI analyzer, hexokinase method). The document implicitly refers to "analytical testing" which would involve this comparison.

    8. The sample size for the training set:

    The document does not provide details about a "training set" for an AI model. For traditional medical devices like blood glucose meters, there isn't a "training set" in the machine learning sense. The device's internal algorithm is developed and calibrated, and then its performance is validated through analytical and clinical studies.

    9. How the ground truth for the training set was established:

    Not applicable, as there's no mention of an AI/ML training set in the document. The device's measurement algorithm is likely based on electrochemical principles and calibration data rather than a machine learning training process from labeled data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1