    K Number
    K171656

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-13

    (38 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Genesys Spine Apache® Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

    Device Description

    The Genesys Spine Apache Interbody Fusion System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of a Genesys Spine interbody fusion device, which may be implanted as a single device via:

    • Bi-laterally via a posterior (PLIF) approach
    • As a single device via a transforaminal (TLIF) approach or
    • As a single device via an anterior/anterolateral (ALIF) approach or
    • As a single device via a lateral (LLIF) approach.
      The Genesys Spine Apache Interbody Fusion System implants are manufactured from implantable, medical grade Ti-6Al-4V ELI titanium alloy per ASTM F136. The additional implant offering being proposed has similar technological characteristics and identical indications as the currently cleared product line.
    AI/ML Overview

    I'm sorry, but this document does not contain the information requested regarding acceptance criteria and a study proving device performance as it pertains to AI/ML devices.

    The provided text is a 510(k) premarket notification for the "Genesys Spine Apache Interbody Fusion System," which is a traditional medical device (an intervertebral body fusion device), not an AI/ML software device.

    Therefore, the following information is not present in the document:

    1. Table of acceptance criteria and reported device performance: This would typically involve metrics like sensitivity, specificity, AUC, etc., which are not relevant to a physical spinal implant.
    2. Sample size used for the test set and data provenance: No test set of clinical data for AI/ML performance is mentioned.
    3. Number of experts used to establish ground truth and qualifications: Not applicable for a physical implant.
    4. Adjudication method: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not conducted as this is not an AI-assisted diagnostic device.
    6. Standalone (algorithm only) performance: Not applicable.
    7. Type of ground truth used: Not applicable. Device safety and efficacy for a physical implant are typically assessed through material properties, mechanical testing, and clinical outcomes.
    8. Sample size for the training set: Not applicable as there is no AI/ML model to train.
    9. How ground truth for the training set was established: Not applicable.

    Instead, the document primarily focuses on regulatory approval based on substantial equivalence to previously cleared predicate devices. It mentions:

    • Mechanical performance data: Stating "Not Required" as cross-sectional area calculations, FEA analysis, and projected area calculations showed that a new "worst-case device" was not created. The Finite Element analysis tool used was validated by the predicate system's mechanical test data. This indicates that the performance assessment relied on engineering analysis and comparison to established mechanical properties of similar devices, rather than clinical study of AI/ML diagnostic accuracy.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153123

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-04-26

    (180 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Genesys Spine Apache@ Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine. from L2 to S1, in skeletally mature patients who had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

    Device Description

    The Genesys Spine Apache® Interbody Fusion System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of a Genesys Spine interbody fusion device, which may be implanted as a single device via:

    • Bi-laterally via a posterior (PLIF) approach ■
    • As a single device via a transforaminal (TLIF) approach or ■
    • I As a single device via an anterior/anterolateral (ALIF) approach.
      The system implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.
    AI/ML Overview

    This document is a 510(k) summary for the Genesys Spine Apache® Interbody Fusion System, which is a medical device. The document describes the device, its intended use, and the studies conducted to demonstrate its substantial equivalence to other legally marketed predicate devices.

    Here's the breakdown of the acceptance criteria and study information, based solely on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Static Subsidence (per ASTM F2267-04)The Genesys Spine Apache® Interbody Fusion System was tested in static subsidence per ASTM F2267-04. The results, combined with Cross-Sectional Area calculations and validated FEA analysis, showed that a new worst-case device was not created when compared to the predicate Genesys Spine Interbody Fusion System (K103034). The performance data were sufficient to allow for a substantial equivalence designation.
    Static Expulsion (per ASTM draft standard F-04.25.02.02)The Genesys Spine Apache® Interbody Fusion System was tested in static expulsion per ASTM draft standard F-04.25.02.02. The results, combined with Cross-Sectional Area calculations and validated FEA analysis, showed that a new worst-case device was not created when compared to the predicate Genesys Spine Interbody Fusion System (K103034). The performance data were sufficient to allow for a substantial equivalence designation.
    (Implied) Equivalent technological characteristics to predicate devicesThe additional implant offering being proposed has similar technological characteristics and identical indications as the currently cleared product line (predicate devices). The overall technology characteristics lead to the conclusion of substantial equivalence.

    Limitations: The document does not explicitly state numerical acceptance criteria for static subsidence or expulsion. Instead, it relies on a comparative assessment against a predicate device, aiming to demonstrate that a "new worst-case device was not created" and that the performance data were "sufficient to allow for a substantial equivalence designation."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document states that the system "was tested" but does not provide details on the number of individual devices or tests conducted for static subsidence and static expulsion.
    • Data Provenance: The studies are laboratory-based mechanical performance tests. The country of origin for the data is not explicitly stated but can be inferred as the "Submitter's Address" (Austin, Texas, USA) or the location of the testing facility. The data is prospective in the sense that the tests were conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to the provided document. This submission is for a physical medical device (intervertebral fusion system) and the "ground truth" is established through engineering and mechanical testing standards (ASTM F2267-04, ASTM draft standard F-04.25.02.02, FEA analysis), not through expert consensus on diagnostic images or clinical outcomes.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human readers/interpreters to resolve discrepancies in expert opinions. The presented studies are mechanical tests, governed by standardized protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. An MRMC study is not relevant to the type of device and studies described in this document. This typically applies to diagnostic devices where the performance of human readers (e.g., radiologists) is evaluated with and without AI assistance. This document describes a surgical implant and its mechanical properties.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    No. This question is not applicable. The device is a physical implant, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluation of this device is based on established engineering and mechanical standards and comparative analysis to predicate devices. Specifically:

    • ASTM F2267-04 for static subsidence.
    • ASTM draft standard F-04.25.02.02 for static expulsion.
    • Cross-Sectional Area calculations.
    • Validated FEA (Finite Element Analysis).
    • The performance of the legally marketed predicate devices.

    There is no "expert consensus," "pathology," or "outcomes data" specifically used to establish the "ground truth" for these mechanical tests in the context of this 510(k) summary.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is not an AI/ML algorithm that requires a "training set." The studies performed are mechanical tests for a physical implant.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device submission.