Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201608
    Date Cleared
    2021-01-28

    (227 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GenesisZr® UHT700 Zirconia blanks are indicated for Monolitic-ceramic for single-unit anterior or posterior prostleses and for three-unit prostheses not involving molar restoration, adhesively cemented. Also indicated for fully covered substructure for single-unit anterior or posterior prostheses and for three-unit prostheses not involving molar restoration_adhesively, or non-adhesively cemented.

    Device Description

    GenesisZr® UHT700 Zirconia are disc and block shaped dental porcelain zirconia oxide blanks that come in various sizes that are used in custom restorations by the dental laboratory. The dental laboratory will further process the blank by milling the blank based upon the anatomically rendering of the patient's teeth (done at the dental office) through "Computer Aided Drafting/ Computer Aided Machining (CAD/CAM). Once the custom rendered blank is milled the product is fully sintered and colored (if required) and fitted to the patient's teeth as a crown.

    AI/ML Overview

    The provided text describes the 510(k) Premarket Submission for GenesisZr® UHT700 Zirconia. This submission focuses on demonstrating substantial equivalence to a predicate device (CCRI, Inc., Pavati® Z40.2 (K160867)) primarily through non-clinical testing of physical, chemical, and material properties, and comparison of manufacturing processes and indications for use.

    Crucially, no information about a study proving the device meets acceptance criteria for an AI or software algorithm is present in this document. The document pertains to a dental material (zirconia blanks) and its physical and chemical properties, not a software device requiring performance metrics like accuracy, sensitivity, or specificity.

    Therefore, I cannot provide the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, training set details, or ground truth establishment for a test set or training set. This document does not describe such a study for this medical device.

    The only "performance" discussed relates to the material's physical and chemical properties as measured against established dental material standards (ISO 6872:2015 and ISO 13356:2015).

    Here's a breakdown of what is available in the document regarding the device's characteristics, which are the basis for its clearance, but not an AI/ML study:


    The available document details the non-clinical testing performed to demonstrate substantial equivalence of the GenesisZr® UHT700 Zirconia blanks to the predicate device, Pavati® Z40.2 (K160867). This is a material science characterization, not a performance study of a diagnostic or therapeutic device with AI/ML components.

    Here's the relevant information about the acceptance criteria and the "study" (non-clinical testing) that proves the device meets these criteria in the context of a dental material:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by international standards for dental ceramic materials (ISO 6872:2015 and ISO 13356:2015) and comparison to the predicate device's properties. The "performance" reported is the measured physical and chemical properties of the GenesisZr® UHT700 Zirconia.

    PropertyAcceptance Criteria (from ISO standards / Predicate)Reported GenesisZr® UHT700 Zirconia Performance
    Sintered Density (g/cm³)≥ 6.0 (ISO 13356:2015, Sec. 4.1 Req't.)6.04
    Thermal Expansion Coefficient (20-500°C)Comparable to predicate (10.3 µm/m °C)10.2 µm/m °C
    Bending Strength (MPa)> 600 MPa (and comparable to predicate)600 MPa
    Chemical Solubility (µg/cm²)< 100 (ISO 6872:2015)Meets ISO Standard (< 100)
    Grain sizeNot specified as an explicit acceptance criteria comparison, but reported.0.66 µm
    Fracture toughness (MPa m0.5)> 2.0 (and comparable to predicate)2.4
    ZrO2 + HfO2 + Y2O3+ Al2O3 (%)> 99.9> 99.9
    Y2O3 (%)9.85% ±0.659.85% ±0.65
    Al2O3 (%)≤ 0.1≤ 0.01
    SiO2 (%)≤ 0.02≤ 0.02
    Fe2O3 (%)≤ 0.01≤ 0.01
    HfO2 (%)≤ 5.0Not explicitly listed, but implied by total ZrO2+HfO2+Y2O3+Al2O3 > 99.9
    Freedom from extraneous materials (Uranium238 Bq g-1)< 1.0 (ISO 6872:2015 Section 5.2)< 0.03

    2. Sample size used for the test set and the data provenance:

    This document describes non-clinical bench testing of material properties, not testing on human subjects or clinical data. Therefore, the concept of "test set sample size" in the context of an AI/ML study is not applicable. The sample sizes for the material property tests (e.g., number of specimens for bending strength) are not specified in this summary, but would be standard for material characterization.

    • Data Provenance: The data comes from "sintered tests coupled with mechanical bench testing" performed by United Dental Resources Corporation. The location/country of origin of this testing is not specified, nor is whether it was retrospective or prospective in conventional clinical study terms, as it's lab testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. Ground truth for material properties is established by physical and chemical measurements using validated laboratory methods and instruments, not by expert human interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. This is not a study involving human readers or interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a dental material, not an AI-assisted diagnostic or therapeutic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. No algorithm is involved.

    7. The type of ground truth used:

    The ground truth used for this device's non-clinical testing is measured physical and chemical properties validated against established international standards (ISO 6872:2015, ISO 13356:2015).

    8. The sample size for the training set:

    Not applicable. There is no AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable. There is no AI/ML algorithm requiring a training set.


    In summary, the provided document details a 510(k) submission for a dental material based on substantial equivalence through non-clinical laboratory testing and comparison to an existing predicate device, not through clinical studies of an AI-powered device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1