510(k) Data Aggregation

    K Number
    K160654

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-05-04

    (57 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GastriSail™ Gastric Positioning System is indicated for use in gastric and bariatric surgical procedures for the application of suction, decompression and to radially expand the stomach, to drain gastric fluids, to test for leaks, to provide visible and tactile delineation of the lesser curvature of the stomach and to serve as a sizing guide for gastric and bariatric procedures, such as sleeve gastrectomy.

    Device Description

    The GastriSail™ Gastric Positioning System is a 40FR flexible, non-sterile, single-use, dual lumen gastric tube designed to be inserted into the esophagus by an anesthesiologist for use in gastric and bariatric surgical procedures, such as sleeve gastrectomy.

    The GastriSail™ Gastric Positioning System consists of a dual lumen flexible tube. One lumen houses the Sail feature and integrated LED guidelights, and the other lumen has rows of distal holes which, when connected to suction provides vacuum to the stomach for evacuation of stomach contents and suction fixation along the lesser curvature of the stomach. This lumen is also used to perform leak testing at the end of the procedure.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, the GastriSail™ Gastric Positioning System. It primarily focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the device meets specific acceptance criteria through a clinical study with defined performance metrics for a novel AI algorithm.

    Therefore, many of the requested categories relating to AI algorithm performance and clinical study design are not applicable or cannot be extracted from this document.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a table of specific numerical acceptance criteria (e.g., sensitivity, specificity, accuracy) for a device's performance in the way an AI algorithm study would. Instead, it details non-clinical and in-vivo testing to demonstrate that the modified device's characteristics are comparable to the predicate device.

    The "acceptance criteria" are implied by the types of tests conducted and their successful completion, indicating that the device performs as expected and is substantially equivalent to the predicate.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility (for new tip material)Testing conducted in accordance with ISO 10993-1:2009. (Implies successful completion, as the conclusion states substantial equivalence).
    Stability (for new tip material)Testing conducted. (Implies successful completion, as the conclusion states substantial equivalence).
    Performance (Nonclinical - Visual Inspection)Testing conducted. (Implies satisfactory results).
    Performance (Nonclinical - Dimensional Measurement)Testing conducted. (Implies satisfactory results, ensuring dimensions are within acceptable tolerances, especially the 40 Fr/Ch size).
    Performance (Nonclinical - Vacuum Pressure Evaluation)Testing conducted. (Implies satisfactory results, demonstrating proper suction function).
    Performance (Nonclinical - Tensile Test Evaluation)Testing conducted. (Implies satisfactory results, ensuring material strength and integrity).
    Performance (Nonclinical - Battery Test)Testing conducted. (Implies satisfactory results for the LED guide lights).
    Performance (In Vivo - Esophageal Insertion Test in Porcine)Evaluation conducted. (Implies successful insertion and functionality in an in-vivo model, confirming safety and performance related to insertion and positioning).
    Substantial Equivalence to predicate device (K143088)"The results of testing demonstrate that the modified GastriSail™ Gastric Positioning System 40 Fr/Ch (13.3mm) is substantially equivalent to the legally marketed GastriSail™ Gastric Positioning System (K143088)."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes for the "test set" in terms of number of patients or cases. The non-clinical and in-vivo tests likely used a predetermined number of devices/specimens for each test, but these are not explicitly stated.

    • Test Set Sample Size: Not specified.
    • Data Provenance: Not specified, but the in-vivo testing was conducted in "porcine," indicating animal model data, not human patient data. These studies are typically prospective laboratory/animal model studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable. This device is a physical surgical tool, not an AI algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for the performance tests would be objective measurements and observations determined by engineers and researchers during the testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This concept applies to human expert review and consensus, typically for establishing ground truth in diagnostic studies, which is not the nature of this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic device, and no MRMC study was conducted. The device is a physical surgical instrument.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the non-clinical and in-vivo testing would be based on:

    • Objective Measurements: For dimensional measurements, vacuum pressure, tensile strength, battery performance.
    • Engineering Specifications: Whether the device meets predefined criteria for its physical and functional attributes.
    • Observation/Evaluation by Trained Personnel: For visual inspection and the esophageal insertion test in porcine models, assessing proper function and absence of adverse events.

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143088

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-02-25

    (120 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GastriSail™ Gastric Positioning System is indicated for use in gastric and bariatric surgical procedures for the application of suction, decompression and to radially expand the stomach, to drain gastric fluids, to test for leaks, to provide visible and tactile delineation of the lesser curvature of the stomach and to serve as a sizing guide for gastric and bariatric procedures, such as sleeve gastrectomy.

    Device Description

    The GastriSail™ Gastric Positioning System is a 36FR flexible, non-sterile, single-use, dual lumen gastric tube designed to be inserted into the esophagus by an anesthesiologist for use in gastric and bariatric surgical procedures, such as sleeve gastrectomy.

    The GastriSail™ Gastric Positioning System consists of a dual lumen flexible tube. One lumen houses the Sail feature and integrated LED guidelights, and the other lumen has rows of distal holes which, when connected to suction provides vacuum to the stomach for evacuation of stomach contents and suction fixation along the lesser curvature of the stomach. This lumen is also used to perform leak testing at the end of the procedure.

    AI/ML Overview

    The provided text describes a medical device, the GastriSail™ Gastric Positioning System, and its clearance through the FDA's 510(k) pathway, which relies on demonstrating substantial equivalence to a legally marketed predicate device. The document focuses on regulatory approval rather than a detailed study proving the device meets specific quantitative acceptance criteria in a clinical setting.

    Therefore, much of the requested information (like effect size of MRMC studies, details of expert adjudication, ground truth for training sets, etc.) is not present in this regulatory document, as it is generally not required for a 510(k) submission unless clinical studies are specifically called for to demonstrate substantial equivalence.

    However, I can extract the available information related to performance and testing:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" for a study in the sense of predefined thresholds for clinical performance metrics. Instead, it lists various non-clinical and comparative tests conducted to demonstrate the device's functionality and substantial equivalence to predicate devices. The "reported device performance" is essentially that the device passed these tests.

    Acceptance Criteria (Implied by Test Type)Reported Device Performance (Summary)
    Dimensional analysisPassed (Dimensions are acceptable)
    Tensile testing of adhered bondsPassed (Bonds are sufficiently strong)
    Battery functional testingPassed (Battery operates as intended)
    Vacuum relief testingPassed (Vacuum relief functions correctly)
    Suction tubing adapter testPassed (Adapter performs as expected)
    Decompression of the stomach (Comparative)Comparable to Realize™ Gastric Calibration Tube
    Suction fixation and evacuation of stomach contents (Comparative)Comparable to Realize™ Gastric Calibration Tube
    Sizing of a Gastric Sleeve (Comparative)Comparable to Realize™ Gastric Calibration Tube
    Esophageal insertion test in porcine (In-vivo)Passed (Safe and effective insertion)
    Biocompatibility (ISO 10993-1:2009)Materials tested and passed
    Electrical Safety Testing (IEC 60601-1: Ed3, IEC 60601-1-2: Ed 4)Passed (Meets electrical safety standards)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not specified for any of the non-clinical or comparative tests. For the in-vivo evaluation, it was performed "in porcine," meaning animal subjects, but the number of animals or procedures is not given.
    • Data Provenance: The studies are non-clinical (bench testing) and in-vivo animal testing. It's not human clinical data, so country of origin isn't relevant in the typical sense. The regulatory submission is from the USA (Covidien LLC, North Haven, CT).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • This information is not provided. For non-clinical and comparative performance tests, "ground truth" and "experts" as in a clinical study are not typically applicable in the same way. Performance is measured against engineering specifications or direct comparison with a predicate device. For the in-vivo test, veterinary or surgical expertise would be involved, but specific numbers and qualifications are not detailed.

    4. Adjudication Method for the Test Set

    • Not applicable/Not specified for the non-clinical and animal performance testing.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC or human-in-the-loop study was conducted. This device is a physical medical instrument, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable, as this is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    • For non-clinical safety and performance tests: Engineering specifications, direct measurements, and comparison to the predicate device's established performance.
    • For biocompatibility: Adherence to ISO 10993-1:2009 standards.
    • For electrical safety: Adherence to IEC 60601-1 and 60601-1-2 standards.
    • For the in-vivo test: Observed physiological response and device function in a porcine model.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not use machine learning or AI, and therefore does not have a "training set" in that context.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable as there is no training set.