Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222080

    Validate with FDA (Live)

    Date Cleared
    2022-10-06

    (83 days)

    Product Code
    Regulation Number
    892.1650
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Garion is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.

    Device Description

    GARION is mobile fluoroscopy system is designed to provdie fluoroscopic and spot-film images of thepatient during diagnostic, surgical procedures. This device is a digital X-ray radiographic equipment that consists of high voltage generator, X-ray control unit, X-ray tube unit, Collimator, image processing unit. The image processing unit consists of medical image detector, power supply, medical image collecting unit and relevant software.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Garion", an image-intensified fluoroscopic x-ray system. The document focuses on demonstrating substantial equivalence to a legally marketed predicate device, "Zenition 70".

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document focuses on demonstrating substantial equivalence to a predicate device rather than defining specific performance acceptance criteria for the Garion device against a set of quantitative metrics. The "performance" is implicitly assessed through a comparison of technical characteristics to the predicate.

    Acceptance Criteria (Implicit from Substantial Equivalence Claim)Reported Device Performance (Comparison to Predicate: Zenition 70 K183040)
    Indications for UseSame as predicate
    X-ray Tube Anode TypeRotating Anode (Same as predicate)
    X-ray Tube Anode Target Angle10° (Same as predicate)
    X-ray Tube Focal Size0.3/0.6 (Same as predicate)
    Fluoroscopic Mode kV range40-125 kV (Similar, slightly higher peak kVp than predicate's 40-120 kV)
    Fluoroscopic Mode mA range0.1-100 mA (Similar, slightly higher mA than predicate's 0.5-60 mA)
    Pulse FluoroYES (Same as predicate)
    ABS functionYES (Same as predicate)
    FPD ScintillatorCesium Iodide (Same as predicate)
    FPD Detector typeAmorphous silicon detector (Same as predicate)
    FPD Active detector size228.6 mm x 228.6 mm (Similar, slightly larger image than predicate's 207x207 mm)
    FPD Total pixel matrix1024x1024 (Similar to predicate's 1344x1344)
    FPD Pixel pitch205 μm (Similar to predicate's 154 μm)
    FPD A/D Conversion16 bit (Same as predicate)
    FPD MTF (1.0 lp/mm)0.64 (Similar, slightly better than predicate's 0.59)
    FPD DQE (0 lp/mm)0.77 (Same as predicate)
    SID980mm (Similar to predicate's 993mm)
    C-arm Rail Rotation Range±180° (Similar to predicate's ±200°)
    Linear FR-arm Movement Range200mm (Same as predicate)
    Linear T-arm Movement Range400mm (Similar to predicate's 490 mm)
    Swing-arm Rotation Range±15° (Similar, slightly better than predicate's ±10°)
    Collimator ControlMotor control / rotation (Same as predicate)

    The acceptance criterion, in this context, is that the Garion device's technological characteristics are substantially equivalent to the predicate device, implying similar safety and effectiveness. The reported performance is the direct comparison of these characteristics.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states: "9. Summary of clinical testing: Not required for a finding of substantial equivalene." This indicates that no clinical test set was used to evaluate the device's performance against specific acceptance criteria. The evaluation relies on comparing the technical specifications of the new device to the predicate device and adherence to relevant IEC standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Since no clinical test set was used, there were no experts establishing ground truth for a test set.

    4. Adjudication Method for the Test Set:

    Not applicable, as no clinical test set was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The document states that clinical testing was not required for substantial equivalence.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The "Garion" is an X-ray system, not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance is not applicable in the same way as it would be for a pure software device. The evaluation is of the hardware and integrated software for image acquisition and display.

    7. The Type of Ground Truth Used:

    The "ground truth" for the substantial equivalence claim is effectively the established safety and effectiveness of the legally marketed predicate device (Zenition 70), supported by the device's compliance with applicable performance standards (IEC and FDA Radiation Safety Performance Standard). The comparison of technical specifications aims to show that the Garion device is sufficiently similar to the predicate to assume similar safety and effectiveness.

    8. The Sample Size for the Training Set:

    The concept of a "training set" is typically associated with machine learning or AI algorithms. Since this submission is for an X-ray system and does not mention any AI components requiring supervised learning, there is no training set in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no mention of a training set or AI algorithms requiring ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1