Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K190335

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-06

    (265 days)

    Product Code
    Regulation Number
    862.1310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GSP Neonatal Total Galactose kit is intended for the quantitative determination of total galactose and galactose-1-phosphate) concentrations in blood specimens dried on filter paper as an aid in screening newborns for galactosemia using the GSP® instrument.

    Device Description

    The GSP Neonatal Total Galactose test system measures total galactose, i.e. both galactose and galactose-1-phosphate, using a fluorescent galactose oxidase method. The fluorescence is measured using an excitation wavelength of 505 nm and an emission wavelength of 580 nm. The GSP Neonatal Total Galactose kit contains sufficient reagents to perform 1152 assays. The kit contains Calibrators, Controls, Neonatal Total Galactose Assay Reagent 1, Neonatal Total Galactose Assay Reagent 2, Neonatal Total Galactose Assay Buffer, Neonatal Total Galactose Assay Reconstitution Solution, and Neonatal Extraction Solution.

    AI/ML Overview

    The provided document describes the K190335 submission for the GSP Neonatal Total Galactose kit, a device used for screening newborns for galactosemia. It primarily focuses on demonstrating the substantial equivalence of the new device (3309-002U) to a previously cleared predicate device (3309-001U).

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text, with the understanding that this is a medical device clearance document, not an AI/ML model acceptance study. Therefore, some of the requested information (like number of experts for AI ground truth, MRMC study, training set details) are not directly applicable to this type of device and submission.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the comparison to the predicate device and standard analytical performance metrics for in vitro diagnostic tests. The goal is to show that the new device performs equivalently to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance Criteria (typically similar to predicate performance or within acceptable ranges)Reported Device Performance (GSP Neonatal Total Galactose kit - 3309-002U)
    Intended UseQuantitative determination of total galactose and galactose-1-phosphate in dried blood specimens as an aid in screening newborns for galactosemia.Same as predicate.
    Test MethodologyEnzymatic assayEnzymatic assay
    Detection MethodFluorescence – measured at 505 nm and 580 nm wavelengthsFluorescence – measured at 505 nm and 580 nm wavelengths
    Instrument PlatformGSP instrument, automated (K090846)GSP instrument, automated (K090846)
    Sample TypeDried blood spotDried blood spot
    Reportable Range1.15 - 50 mg/dL (Predicate)1.2 - 50 mg/dL
    Limit of Blank (LoB)0.34 mg/dL (Predicate)0.3 mg/dL (17 umol/L)
    Limit of Detection (LoD)0.97 mg/dL (Predicate)0.7 mg/dL (39 umol/L)
    Limit of Quantitation (LoQ)1.15 mg/dL (Predicate)1.2 mg/dL (67 umol/L), defined as lowest concentration with total CV ≤ 20%
    CalibratorsSpecific values (0.5, 2.5, 5.0, 10.0, 20, 50 mg/dL)Specific values (0.5, 2.5, 5.0, 10, 20, 50 mg/dL)
    Total Variation (%CV)Not explicitly stated as acceptance criteria, but demonstrates precision.Ranged from 10.0 to 13.9 %CV
    LinearityDemonstrated to be linear throughout the measuring range.Linear from 1.2 mg/dL to 50 mg/dL
    RecoveryNot explicitly stated as acceptance criteria, but demonstrates accuracy.Average recovery: Galactose 98%, Galactose-1-phosphate 115%, Combined 102%
    InterferenceBias exceeding ±15% is considered significant interference.Acetaminophen, Conjugated Bilirubin, Intralipid, Hemoglobin/Bilirubin combinations tested (see detailed tables in source for significant interference levels)
    Hook EffectNo hook effect expected within relevant range.No hook effect found up to 500 mg/dL
    Method Comparison (vs. 3309-001U)Close correlation with predicate device.mg/dL: y = 1.00x + 0.33; 95% CI: slope (0.96; 1.04), intercept (0.25; 0.42) (n=545)
    Overall Percent Agreement (95th percentile)High agreement with predicate.98.7 % (95%CI 98.1 % - 99.1 %)
    Positive Percent Agreement (95th percentile)High agreement with predicate.87.7 % (95%CI 80.3 % - 93.1 %)
    Negative Percent Agreement (95th percentile)High agreement with predicate.99.3 % (95%CI 98.9 % - 99.6 %)
    Overall Percent Agreement (99th percentile)High agreement with predicate.99.4 % (95%CI 98.9 % - 99.6 %)
    Positive Percent Agreement (99th percentile)High agreement with predicate.74.1 % (95%CI 53.7% - 88.9 %)
    Negative Percent Agreement (99th percentile)High agreement with predicate.99.7 % (95%CI 99.3 % - 99.9 %)

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Analytical Validation (Precision, LoB, LoD, LoQ, Linearity, Recovery, Interference, Hook Effect): Various sample sizes specific to each experiment (e.g., 150 for LoB, 60 for LoD, 40 plates/80 results for repeatability, 15 plates/75 results for between-instrument, 15 plates/75 results for between-lot variation). Samples were human red blood cell enriched with galactose, human blood enriched with galactose and galactose-1-phosphate, whole blood with added substances, or contrived dried blood spot samples.
    • Method Comparison:
      • Comparison with 3029-0010 Neonatal Total Galactose kit (different method): n=139 samples (routine newborn screening dried blood spot samples and dried adult human whole blood samples spiked with galactose and galactose-1-phosphate).
      • Comparison with 3309-001U GSP Neonatal Total Galactose kit (predicate): n=545 routine newborn screening dried blood spot samples.
    • Screening Performance Study:
      • Test Set Size: 2161 samples.
      • Data Provenance: Conducted at one newborn screening laboratory in the United States.
      • Retrospective/Prospective: The samples included "routine newborn screening dried blood spot samples" (implying prospective collection in a screening program context) and "retrospective galactosemia diagnosed screening samples" (implying retrospective identification of confirmed positive cases). Specifically, the 95th percentile analysis included 5 retrospective galactosemia diagnosed screening samples and 1 retrospective galactosemia screening sample collected 22 hours after birth. The 99th percentile analysis included 4 retrospective galactosemia diagnosed screening samples, 1 retrospective galactosemia diagnosed screening sample collected 22 hours after birth, and 1 retrospective galactosemia diagnosed screening sample collected 16 hours after birth.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This is an in vitro diagnostic (IVD) device, not an AI/ML model for image interpretation. The "ground truth" for the screening performance study is clinical diagnosis of galactosemia, or the results from the predicate device using routine newborn screening samples.
    • The document does not specify the number or qualifications of experts (e.g., radiologists) in the context of establishing ground truth, as this is not a study involving subjective interpretation like medical imaging by human experts. The 'truth' is derived from the biochemical measurements and clinical outcomes associated with galactosemia screening performed by a validated screening program.

    4. Adjudication Method for the Test Set:

    • Not applicable in the context of an IVD device. The 'comparison' and 'screening performance' results are based on quantitative measurements by both the new device and the predicate device compared to each other, and against known clinical results (for retrospective samples). There is no "adjudication" between human readers or AI outputs in the way it's understood for image interpretation or diagnosis.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. An MRMC study is relevant for perception tasks like image interpretation where human readers' performance is evaluated. This is an in vitro diagnostic assay that produces quantitative results. The comparison is between the new device's quantitative output and the predicate device's quantitative output, as well as their agreement on screening classification (positive/negative) against the established screening program's outcomes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, in a sense. The GSP Neonatal Total Galactose kit is a standalone in vitro diagnostic system. Its performance (accuracy, precision, linearity, etc.) is measured intrinsically and compared to the predicate, independent of human operators' subjective interpretation. The "screening performance study" evaluates the device's ability to classify samples as screen positive or negative based on its quantitative output, without direct human "interpretation" of the assay result itself. Human decision-making uses this quantitative result but isn't part of the direct device performance.

    7. The Type of Ground Truth Used:

    • For analytical performance (Precision, LoB, LoD, LoQ, Linearity, Recovery, Interference, Hook Effect): The ground truth is established by known concentrations of analytes (galactose, galactose-1-phosphate) in spiked samples, or by established measurement principles for blank and low-level samples.
    • For Method Comparison: The ground truth is the measurement obtained from the predicate device and the 3029-0010 Neonatal Total Galactose kit. This establishes agreement.
    • For Screening Performance Study: The ground truth is a combination of:
      • Predicate Device Results: For routine samples, the predicate device's classification (screen positive/negative) serves as the comparator.
      • Clinical Diagnosis/Outcomes: For "retrospective galactosemia diagnosed screening samples," the confirmed clinical diagnosis of galactosemia (presumably through follow-up testing and clinical presentation) serves as the ultimate truth for these specific cases.

    8. The Sample Size for the Training Set:

    • This device is an IVD kit, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The device's calibration curve is established using known calibrators provided in the kit. The "training" of an IVD like this involves chemical formulation, assay optimization, and manufacturing process control, not data-driven learning from a "training set" like an AI model.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set for this type of device. The calibrators have been prepared from human red blood cells enriched with galactose and calibrated against primary calibrators gravimetrically prepared using a U.S. Pharmacopeia Reference Standard Preparation for galactose. This establishes the "truth" for calibration.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133652

    Validate with FDA (Live)

    Date Cleared
    2014-04-28

    (152 days)

    Product Code
    Regulation Number
    862.1310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GSP Neonatal Total Galactose kit is intended for the quantitative determination of total galactose (galactose and galactose-1-phosphate) concentrations in blood specimens dried on filter paper as an aid in screening newborns for galactosemia using the GSP® instrument.

    Device Description

    The GSP Neonatal Total Galactose kit contains sufficient reagents to perform 1152 assays. The GSP Neonatal Total Galactose test system measures total galactose, i.e. both galactose and galactose-1-phosphate, using a fluorescent galactose oxidase method. The fluorescence is measured using an excitation wavelength of 505 nm and an emission wavelength of 580 nm. The kit contains Neonatal Total Galactose Assay Reagent 1, Neonatal Total Galactose Assay Reagent 2, Neonatal Total Galactose Assay Buffer, Neonatal Total Galactose Assay Reconstitution Solution, and Neonatal Extraction Solution. Calibrators and Controls are also included.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Precision (Total Variation)Ranged from 9.3% to 14.1% CV.
    Limit of Blank (LoB)0.34 mg/dL
    Limit of Detection (LoD)0.97 mg/dL
    Limit of Quantitation (LoQ)1.15 mg/dL (defined as the lowest concentration with a total CV equal to or less than 20%).
    LinearityDemonstrated linear performance throughout the measuring range (from 1.15 mg/dL to 50 mg/dL).
    RecoveryAverage recovery of 109% for galactose, 117% for galactose-1-phosphate, and 103% for both combined from three contrived dried blood spot samples.
    Interference- Acetaminophen: Concentrations above 2.75 mg/dL caused a significant decrease (>15%) in measured total galactose. Maximum tested (5.5 mg/dL) caused a decrease of ~20-22%.- Conjugated Bilirubin: Concentrations above 16.6 mg/dL caused a significant decrease (>15%) in measured total galactose. At 24.9 mg/dL and above, the decrease was 100% at some total galactose concentrations.- Intralipid: Concentrations above 250 mg/dL (at 5 and 10 mg/dL total galactose) or 375 mg/dL (at 15 mg/dL total galactose) caused a significant increase (>15%) in measured total galactose. Maximum tested (1500 mg/dL) caused an increase of ~52-77%.- Hemoglobin (with Bilirubin): Hemoglobin levels at 198 g/L and above in combination with an elevated bilirubin level of 15 mg/dL caused a significant increase (>15%) in measured total galactose at certain total galactose concentrations. For example, at 5 mg/dL total galactose, 198 g/L Hb led to a 26.3% increase.- Non-Interfering Substances: Unconjugated bilirubin (20 mg/dL), ß-Nicotinamide adenine dinucleotide (100 µmol/L), Glutathione (3 mmol/L), Human Serum Albumin (30 mg/mL), Ascorbate (6 mg/dL), D-glucose (1000 mg/dL), D-mannose (100 mg/dL), D-fructose (18 mg/dL), Ampicillin (152 µmol/L), and Lithium heparin (0.375 mg/ml), and Hematocrit levels from 30% to 66% (102-230 g/L Hemoglobin) were found not to interfere.
    Screening Performance vs. Predicate (95th percentile)Overall percent agreement = 96.0%Positive percent agreement = 63.6%Negative percent agreement = 97.9%
    Screening Performance vs. Predicate (99th percentile)Overall percent agreement = 98.8%Positive percent agreement = 53.3%Negative percent agreement = 99.4%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Precision Study: 7 samples, with 216 total measurements per sample (4 replicates per sample, in 27 runs over 21 days using 3 kit lots and 3 GSP instruments).
    • Sample Size for LoD: 216 determinations of 4 low-level samples.
    • Sample Size for LoB: 150 blank samples.
    • Sample Size for Recovery: 3 contrived dried blood spot samples.
    • Sample Size for Interference Studies: Not explicitly stated for each concentration, but involved various concentrations of interfering substances at three total galactose concentrations (5, 10, and 15 mg/dL).
    • Sample Size for Internal Method Comparison: 141 routine screening and spiked blood spot specimens.
    • Sample Size for Screening Performance Study: 2320 samples (6 confirmed positive samples and 2314 routine samples).
    • Data Provenance: The screening performance study was conducted at "one newborn screening laboratory in the United States." Other non-clinical studies (precision, linearity, LoB/LoD/LoQ, recovery, interference) appear to be internal laboratory studies without specific geographic provenance mentioned, but presumably also conducted in the US or Finland (Wallac Oy headquarters). The studies were retrospective, using banked samples and contrived samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish ground truth for the test set. For the screening performance study, "6 confirmed positive samples" are mentioned, implying prior clinical diagnosis as the ground truth. The qualifications of who confirmed these positive cases or how the "routine samples" were classified as normal are not specified.

    4. Adjudication Method for the Test Set

    Not applicable. The document does not describe any expert adjudication process for the test set. Ground truth for confirmed positive samples likely came from clinical diagnosis.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an in-vitro diagnostic test kit (laboratory assay) for quantitative determination, not an imaging device or AI-assisted diagnostic tool that would involve human readers interpreting results in a MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the assay itself. The entire submission details the standalone performance of the GSP Neonatal Total Galactose kit (assay only) without human-in-the-loop interpretation beyond standard laboratory procedures and reporting. The "GSP® instrument" is automated, as stated in the comparison chart ("GSP instrument, automated").

    7. The Type of Ground Truth Used

    • For Analytical Performance (Precision, LoB, LoD, LoQ, Linearity, Recovery, Interference): Ground truth was established by preparing samples with known concentrations of total galactose or specific interfering substances. For example, for recovery, the "recovery of galactose, galactose-1-phosphate, and both combined was determined from three contrived dried blood spot samples," meaning these samples were prepared with known amounts.
    • For Screening Performance Study: The ground truth for the 6 positive samples was "confirmed positive." This implies a clinical diagnosis of galactosemia, likely through follow-up diagnostic testing. The other 2314 samples are referred to as "routine samples" and classified as "normal" in the context of screening performance, likely based on either their negative predicate device result or their actual clinical status. The document also compares the new device's results against the predicate device's results as a reference for "Manual result."

    8. The Sample Size for the Training Set

    Not applicable in the conventional sense of machine learning training sets. This is a chemical assay, not an AI/ML device that requires a distinct training set for model development. The development and optimization of the assay would involve various experiments, but these are not referred to as "training sets" in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/ML model for this chemical assay. The development of the calibrators and controls (which are prepared with known concentrations of galactose and galactose-1-phosphate) serves an analogous function in ensuring accuracy and consistency of the assay.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1