Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141829

    Validate with FDA (Live)

    Date Cleared
    2014-09-24

    (79 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GlucoVitaal H1A Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, forearm, upper arm, thigh, or calf. The GlucoVitaal H1A Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The GlucoVitaal H A Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes controls. The Gluco Vitaal H1 A Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GlucoVitaal H1A Blood Glucose Test Strips are for use with the Glood Glucose Test Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, forearm, upper arm, palm, thigh, or calf.

    The GlucoVitaal H1A Glucose Control Solution is for use with the GlucoVitaal H1A Blood Glucose Test Meter and Test strips as a quality control check to verify that the meter and test strips are working together properly and that the test is performing correctly.

    Device Description

    The GlucoVitaal H1A Blood Glucose Monitoring System consists of GlucoVitaal H1A Blood Glucose Test Meter, GlucoVitaal H1A Blood Glucose Test Strips, GlucoVitaal H1A Glucose Control Solution 1 and Control Solution 2, a Lancing device. Control Solution 1 and Control Solution 2 are required but not included with the meter. Control Solution 1 and Control Solution 2 are always provided as a set.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GlucoVitaal H1A Blood Glucose Monitoring System, seeking substantial equivalence to a predicate device (CERA-CHEK 1070 Blood Glucose Monitoring System).

    Here's an analysis of the acceptance criteria and study data based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a regulated performance standard. Instead, it aims to demonstrate substantial equivalence to a predicate device by comparing various specifications. The "reported device performance" is not given as numerical results against predefined criteria, but rather as statements of identical or similar specifications to the predicate device.

    Acceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance (GlucoVitaal H1A)
    Detection MethodAmperometry (Same as Predicate)
    EnzymeGlucose dehydrogenase (FAD-GDH) (Same as Predicate)
    Test Time5 seconds (Same as Predicate)
    Memory1000 blood glucose test results with date and time (Same as Predicate)
    Sample Volume0.5 uL (Same as Predicate)
    Humidity range10-85% (Same as Predicate)
    Temperature range50-104°F (10-40°C) (Same as Predicate)
    Power (Battery)One 3-volt lithium battery (CR2032) (Same as Predicate)
    Test range20-600 mg/dL (Same as Predicate)
    Level of QC2 Levels (Control 1, Control 2) (Same as Predicate)
    Dimensions94mm(H) x 53.6(W) x 14.9mm(T) (Same as Predicate)
    Weight35g (Same as Predicate)
    Hematocrit range10-70% (Same as Predicate)
    Code SettingNo code (Predicate device required a code key; candidate device has one built-in)
    LanguageEnglish, Spanish and Portuguese (Predicate device only had English)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size used for any test set or the data provenance. It focuses on comparing specifications and features for substantial equivalence, rather than presenting a performance study with detailed results from a specific test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The submission is for a blood glucose monitoring system, which typically relies on laboratory reference methods (e.g., YSI analyzer) for ground truth, not expert human readers.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided in the document given the nature of the device and the substantial equivalence claim. Adjudication methods are typically relevant for image-based diagnostic systems where human interpretation can vary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study is not applicable and not mentioned in the document. This type of study is relevant for AI-assisted diagnostic tools involving human interpretation, which is not the case for a blood glucose monitoring system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    While the device itself operates "standalone" in measuring glucose, the document does not describe a specific performance study in terms of an algorithm-only evaluation. Instead, it asserts that the modification of code setting (from code key to no code) would have "no effect on the performance test," implying that prior performance data (likely from the predicate device) is considered relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not explicitly state the type of ground truth used for any performance evaluation. However, for blood glucose monitoring systems, the ground truth is typically established using a highly accurate laboratory reference method, such as a YSI glucose analyzer.

    8. The sample size for the training set

    The document does not mention a training set or its sample size. This is a 510(k) submission focused on substantial equivalence to an existing device, not the development and training of a novel algorithm from scratch.

    9. How the ground truth for the training set was established

    Since no training set is mentioned, the method for establishing its ground truth is not provided.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1