Found 2 results

510(k) Data Aggregation

    K Number
    K101250

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-03-07

    (307 days)

    Product Code
    Regulation Number
    876.1300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PillCam Platform with a PillCam SB capsules is intended for visualization of the small bowel mucosa.

    • The PillCam Platform with PillCam SB capsules may be used in the visualization and . monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
    • The PillCam Platform with PillCam SB capsules may be used in the visualization of . lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
    • The PillCam Platform with PillCam SB capsules may be used in the visualization of ● lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.

    The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.

    The PillCam Platform with PillCam SB capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.

    Device Description

    The Given PillCam® Platform is comprised of three main subsystems; (1) the ingestible PillCam capsule, (2) the RAPID® software, and (3) the Given® Workstation and Hardware.

    Ingestible PillCam Capsule: The disposable, ingestible PillCam Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via a RF communication channel to the DataRecorder located outside the body.

    RAPID Software: The RAPID Software is a software application that is utilized to process, analyze, store, and view the acquired images collected from the DataRecorder to create a RAPID video of the images. The software also includes a reporting function to create detailed clinical reports, in-service training videos, and patient instruction forms.

    Given Workstation and Hardware: The Workstation is a modified standard personal computer that is the operational platform for the RAPID software. The DataRecorder is an external receiving/recording unit that receives acquired images from the capsule. The SensorArray receives data from the PillCam capsule and transfers the data to the DataRecorder. The RAPID Real Time is a handheld device that allows for real-time viewing of acquired images through the GI tract. Other accessories include a flat panel LCD monitor, a high-capacity mass storage device, and a high-capacity USB portable storage device.

    AI/ML Overview

    The provided document K101250 is a 510(k) summary for the Given PillCam Platform with PillCam SB Capsules. It primarily focuses on demonstrating substantial equivalence to a predicate device (Given Diagnostic System with PillCam SB2 Capsule, K090557) and does not contain detailed information about a de novo study with specific acceptance criteria and performance metrics for the device itself beyond its "Suspected Blood Indicator (SBI) feature".

    However, the document states: "The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas." And later: "The devices meet the guidance entitled "Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA dated November 28, 2001. Clinical data has been summarized to show safety and effectiveness for the proposed indication for use."

    Based on the limited information, here's what can be extracted and inferred, focusing on the SBI feature as the most "AI-like" component described in terms of its function:


    Description of Acceptance Criteria and Study for Given PillCam® Platform with PillCam® SB Capsules (with emphasis on SBI feature)

    The provided submission primarily focuses on the substantial equivalence of the Given PillCam® Platform with PillCam® SB Capsules to a predicate device (Given® Diagnostic System with PillCam® SB2 Capsule, K090557). It does not outline specific, quantitative acceptance criteria for the overall device's performance (e.g., diagnostic accuracy for specific conditions) in the way one might expect for a de novo device with novel algorithms.

    However, the document mentions a "Suspected Blood Indicator (SBI) feature" and states that "The devices meet the guidance entitled 'Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA dated November 28, 2001. Clinical data has been summarized to show safety and effectiveness for the proposed indication for use.'"

    Given the lack of explicit quantitative acceptance criteria or detailed study results for the SBI feature within this 510(k) summary, specific numerical performance metrics cannot be provided directly from this document. The "Performance Data" section broadly states that clinical data was summarized to show safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since the document does not provide specific acceptance criteria or reported performance metrics for the SBI feature (or the overall device's diagnostic ability) in a quantitative format, this table cannot be fully populated as requested. The "acceptance criteria" here implicitly refers to meeting the "Special Controls Guidance Document" and demonstrating safety and effectiveness for the intended use through clinical data summarization.

    Feature/MetricAcceptance Criteria (Not explicitly stated numerically in this document)Reported Device Performance (Not explicitly stated numerically in this document)
    Suspected Blood Indicator (SBI) FunctionalityThe SBI feature is intended to accurately mark frames of video suspected of containing blood or red areas, aiding in the visualization of lesions that may be a source of obscure bleeding or IDA. (Implicitly, the performance should be clinically acceptable and comparable to the predicate device, although specific thresholds are not provided.)The document states: "The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas." It further mentions that "Clinical data has been summarized to show safety and effectiveness for the proposed indication for use" which includes visualizing lesions that may be a source of obscure bleeding or IDA. However, no specific metrics like sensitivity, specificity, or predictive values for the SBI feature are reported. The basis for approval relies on substantial equivalence and the general claim of safety and effectiveness as supported by the clinical data summary (which is not detailed here).
    General Device Safety and EffectivenessDevice meets the guidance document "Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA dated November 28, 2001." Demonstrates safety and effectiveness for its intended uses.The conclusion states: "Based on the technological characteristics and clinical performance of the devices, Given Imaging Ltd. believes that the Given PillCam® Platform with PillCam® SB Capsules and the predicate devices selected are substantially equivalent and do not raise new issues of safety or effectiveness." No specific performance numbers (e.g., for detection rates of Crohn's, obscure bleeding, or IDA) are provided in this summary.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified in this 510(k) summary. The document mentions "Clinical data has been summarized to show safety and effectiveness," implying a study was conducted, but details regarding the size of the test set (number of patients or cases) are absent.
    • Data Provenance: Not specified. Standard clinical studies for such devices often involve multiple clinical sites, potentially across different countries. The document does not indicate if the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified in this 510(k) summary.

    4. Adjudication method for the test set:

    • Not specified in this 510(k) summary.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size:

    • Not explicitly stated that an MRMC comparative effectiveness study was done specifically to quantify how much human readers improve with AI vs. without AI assistance for the SBI feature. The primary focus of this 510(k) is substantial equivalence based on overall device performance and safety, not on the comparative effectiveness of a specific AI assist feature.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • It is not explicitly stated whether a standalone performance study of the SBI algorithm was conducted, divorced from human review. The SBI is described as "intended to mark frames... suspected of containing blood or red areas," suggesting it's an assist feature for human review.

    7. The type of ground truth used:

    • Not explicitly stated. For a device visualizing the small bowel mucosa and detecting lesions, the ground truth would typically be established through consensus of expert gastroenterologists interpreting the capsule endoscopy video, possibly supplemented by findings from conventional endoscopy, biopsy (pathology), or subsequent clinical outcomes if those were part of the study design.

    8. The sample size for the training set:

    • Not applicable/Not specified. The 510(k) submission primarily relies on substantial equivalence to a predicate device and summarizes clinical data rather than describing the development and training of a novel algorithm from scratch. If the SBI feature involved machine learning, details of its training set would typically be provided in a more detailed technical report, which is not part of this summary.

    9. How the ground truth for the training set was established:

    • Not applicable/Not specified, as information on a specific training set for a novel algorithm within this context is not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103088

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-14

    (87 days)

    Product Code
    Regulation Number
    876.1300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SB Indications for Use: The Given PillCam® Platform with the PillCam® SB Capsules is intended for visualization of the small bowel mucosa. It may be used as a tool in the detection of abnormalities of the small bowel in adults and children from two years of age. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.

    Device Description

    The Given PillCam® Platform is comprised of three main subsystems; { } } the ingestible PillCam capsule, (2) the RAPID® software, and (3) the Given® Workstation and Hardware. 1. Ingestible PillCam Capsule The disposable, ingestible PillCam Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via a RF communication channel to the DataRecorder located outside the body. RAPID Software The RAPID Software is a software application that is utilized to process, analyze, store, and view the acquired images collected from the DataRecorder to create a RAPID video of the images. The software also includes a reporting function to create detailed clinical reports, in-service training videos, and patient instruction forms. Given Workstation and Hardware 3 The Workstation is a modified standard personal computer that is the operational platform for the RAPID software. The DataRecorder is an external receiving/recording unit that receives acquired images from the capsule. The PillCam® SensorBelt receives data from the PillCam capsule and transfers the data to the DataRecorder. The RAPID Real Time is a handheld device that allows for real-time viewing of acquired images through the GI tract. Other accessories include a flat panel LCD monitor, a high-capacity mass storage device, and a high-capacity USB portable storage device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state numerical acceptance criteria in a table format. Instead, it describes a clinical validation study aimed at demonstrating the device's performance in a specific population (overweight individuals) to support the removal of a warning statement. The "performance" described is the device's ability to operate without new risks or issues in this expanded population.

    Acceptance Criteria (Implied)Reported Device Performance
    No new risks or issues when using PillCam® SensorBelt in subjects with BMI ≥ 30 kg/m^2.The clinical validation test conducted with 25 healthy volunteers (BMI ≥ 30 kg/m^2) who underwent the SB procedure using PillCam® SB2 capsule, DR2 DataRecorder, and PillCam® SensorBelt showed no new risks. The performance of the SensorBelt has not been studied in patients with BMI greater than 43.3 kg/m^2.
    Substantial equivalence to the predicate device for the aforementioned use case.The clinical validation data demonstrated substantial equivalence to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): 25 healthy volunteers.
    • Data Provenance: The text does not explicitly state the country of origin but is associated with Given Imaging Ltd. in Israel. The study appears to be prospective as it involves healthy volunteers undergoing the procedure.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts or their qualifications for establishing ground truth in this particular study. The study focuses on the performance of the SensorBelt in overweight individuals, implying an assessment of technical function and safety rather than diagnostic accuracy against a "ground truth" of disease.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method. The study's focus was on the performance of the SensorBelt in overweight individuals, assessing for new risks or failures rather than diagnostic discrepancies requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, an MRMC comparative effectiveness study, as typically understood in the context of diagnostic AI, was not performed. The study aimed to validate the device's performance in an expanded population, not to compare human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? The device itself (PillCam® Platform with PillCam® SB Capsules with PillCam® SensorBelt) is a medical imaging system, not purely an algorithm. The "Suspected Blood Indicator (SBI)" feature is mentioned as an intended function to mark frames, but no specific standalone performance study for this algorithmic component (e.g., sensitivity/specificity of SBI flagging against a ground truth of bleeding) is detailed in this document. The clinical validation was for the overall system in an overweight population.

    7. Type of Ground Truth Used

    The "ground truth" for this specific clinical validation study was the observation of the device's performance and the absence of new risks or adverse events in the healthy overweight volunteers. It was not a ground truth for a diagnostic condition (e.g., pathology-confirmed lesions or outcomes data).

    8. Sample Size for the Training Set

    The document does not mention a training set. This study is a clinical validation of the device's operation in a specific population, not a study involving the training of a new AI model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set or ground truth for a training set is mentioned.