Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K063753

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-06-20

    (183 days)

    Product Code
    Regulation Number
    882.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lesion Probe and Generator system is intended to create radiofrequency (RF) heat lesions in nerve tissue. It is intended for use only by trained clinicians in a hospital or clinical setting.

    Device Description

    The GFX Nerve Ablation System provides a minimally invasive technique for creating a neural lesion inhibiting the function of the target nerve. The Generator and probe are used as a system to both stimulate the nerve for the purpose of locating the probe correctly and to create a neural lesion to inhibit nerve function through the application of RF energy. The GFX Nerve Ablation probes are single use devices supplied sterile to the customer.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantitative acceptance criteria or detailed performance metrics from a clinical study. Instead, it relies on a comparison to predicate devices and bench testing to demonstrate substantial equivalence.

    Acceptance Criteria (Not explicitly stated as quantitative targets)Reported Device Performance
    Technological Substantial Equivalence to Predicate Devices:
    - Indications For UseSubstantially equivalent
    - RF Energy Output Type (Frequency, Power, Wave Type)Substantially equivalent
    - Treatment TimeSubstantially equivalent
    - Functions (stimulation and ablation)Substantially equivalent
    - Safety Monitoring Systems (Temperature, Impedance, Power Output)Substantially equivalent
    - Probe Type (single use)Substantially equivalent
    Biocompatibility:Meets ISO 10993-1, AAMI/ISO 10993-7
    Sterilization:Meets ISO 11135
    Packaging:Validated and 100% visual verification for integrity
    Device Integrity and Functionality:Qualified/validated using routine manufacturing samples; meets ACI Inc.'s in-house requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical test set in the traditional sense, such as a cohort of patients for an effectiveness or safety study. The "test set" from a regulatory perspective here refers to:

    • Bench Testing Samples: Devices produced under routine manufacturing conditions. The exact quantity of samples is not specified.
    • Biocompatibility Testing Samples: Not specified, but implied to be sufficient for meeting ISO standards.
    • Sterilization Validation Batches: Not specified, but implied to be sufficient for meeting ISO standards.

    The data provenance for these tests is internal to ACI Inc. and would be considered retrospective in the sense that they are internal validation tests, not a forward-looking clinical trial. There is no mention of country of origin for any human data, as no human data is presented for performance evaluation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable or provided in the document. No clinical test set with human data requiring expert ground truth establishment for a diagnostic or AI-assisted system is described. The device is a direct treatment tool, and its "performance" is assessed through engineering and bench testing, not expert interpretation of outputs.

    4. Adjudication Method for the Test Set

    This is not applicable or provided. There is no clinical test set requiring human adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done and is not described in this document. The device is a therapeutic radiofrequency ablation system, not an imaging or diagnostic aid that would typically involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. The GFX Nerve Ablation System is a physical device (generator and probe) that creates heat lesions. It is not an algorithm, software, or AI system that operates in a "standalone" or "human-in-the-loop" capacity in the way an AI diagnostic tool would. Its performance is evaluated based on its physical and electrical characteristics and ability to produce lesions, as demonstrated through bench testing and comparison to predicate devices.

    7. The Type of Ground Truth Used

    For the bench testing and validation, the ground truth would be based on:

    • Engineering specifications and standards: For RF energy output, temperature control, impedance monitoring, etc. (e.g., meeting specific voltage, current, and temperature targets).
    • Biocompatibility standards: (ISO 10993 series for material safety).
    • Sterilization efficacy standards: (ISO 11135 for sterility assurance).
    • Physical integrity standards: For packaging and device robustness.

    There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for this type of device submission's performance evaluation.

    8. The Sample Size for the Training Set

    This is not applicable. The GFX Nerve Ablation System is not an AI or machine learning device. It does not have a "training set" in the context of algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1