510(k) Data Aggregation

    K Number
    K122493

    Validate with FDA (Live)

    Device Name
    GENESISPLUS
    Manufacturer
    Date Cleared
    2013-05-15

    (272 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cutera GenesisPlus Nd.YAG laser is intended for use in the medical specialties of general and plastic surgery, dermatology, endoscopic.laproscopic.laproscopic.label specialites of general otorhinolaryngology (ENT), neurosurgery, oculoplastics, pulmonary.hnoracis surgery, podiaty and urology for surgical and aesthetic applications.

    Dermatology:
    The Cutera GenesisPlus faser is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, rosaceal diffuse redness, poikiloderma of civatte, scar reduction (including hypertropic and keloid scars), and warts.

    The Cutera GenesisPlus laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.

    The GenesisPlus laser is indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.

    Podiatry:
    Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:

    • Matrixectomy .
    • Periungual and subungual warts .
    • . Plantar warts
    • . Radical nail excision
    • Neuromas

    The Cutera GenesisPlus laser is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophyton rubrum and T. mentagrophytes, and/or yeast Condida Albicans, etc.).

    Device Description

    The Cutera GenesisPlus Laser unit and controls are contained in a single console. Electrical power is supplied to the console by the facility's power source. Laser energy produced within the device is delivered to the tissue by means of a handpiece using a fiber optic delivery system with an optical lens at the aperture. The user activates laser emission by means of a footswitch.

    The Cutera GenesisPlus Laser is designed to provide laser energy for use in a variety of dermatology and podiatry procedures.

    AI/ML Overview

    The Cutera GenesisPlus Laser System, under K122493, is not an AI/ML device. Therefore, the questions related to AI/ML device performance, such as sample sizes for test and training sets, ground truth establishment methods, expert qualifications, adjudication methods, and MRMC studies, are not applicable.

    This submission describes a medical laser system and focuses on demonstrating substantial equivalence to a previously cleared predicate device (Cutera GenesisPlus under K103626) by comparing technical specifications and intended uses. The clinical study mentioned did not involve AI/ML performance metrics but rather evaluated the safety and effectiveness of updated treatment parameters for the podiatry indication concerning onychomycosis.

    Here's the information that is applicable from your request:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for substantial equivalence are based on demonstrating that the revised Cutera GenesisPlus Laser System is as safe and effective as the predicate device (K103626) and shares similar technological characteristics and indications for use. No specific numerical performance metrics or thresholds are provided as "acceptance criteria" in the context of an algorithmic performance study, as this is a laser device modification, not an AI system.

    However, the comparison tables demonstrate that the revised device either matches or has slightly modified parameters (e.g., fluence, spot size, pulse duration, repetition rate for the clear nail indication) that were deemed acceptable based on a clinical study for safety and effectiveness.

    Comparison of Technical Specifications for Dermatology Indications for Use:

    FeaturesCutera GenesisPlus Laser (K103626)Cutera GenesisPlus Laser (revised)
    Wavelength1064nm Nd:YAG1064nm Nd:YAG
    Aiming Beam630-680nm (≤ 2.5mW)630-680nm (≤ 2.5mW)
    Energy per Pulse20-3500 mJ20-3500 mJ
    Fluence25.5J/cm2 (with 1mm spot)25.5J/cm2 (with 1mm spot)
    Power≤ 100W≤ 100W
    Spot SizeUp to 13mmUp to 13mm
    Pulse Duration100-3000μs100-3000μs
    Output modePulsedPulsed
    Repetition Rate5-100 Hz5-100 Hz
    Laser MediaFlashlamp pumped solid state rodFlashlamp pumped solid state rod
    User InterfaceLCD color touchscreenLCD color touchscreen

    Comparison of Technical Specifications for Temporary Increase of Clear Nail Indication for Use:

    FeaturesCutera GenesisPlus Laser (K103626)Cutera GenesisPlus Laser (revised)
    Wavelength1064nm Nd:YAG1064nm Nd:YAG
    Aiming Beam630-680nm (≤ 2.5mW)630-680nm (≤ 2.5mW)
    Fluence25.5J/cm2 (with 1mm spot)15-18 J/cm2
    Power≤ 100W≤ 100W
    Spot Size1.0-1.5mm5mm
    Pulse Duration100μs0.3ms
    Output modePulsedPulsed
    Repetition Rate20 Hz2-3Hz
    Burst Mode Sequence½ second on, ½ second offNot applicable
    Laser MediaFlashlamp pumped solid state rodFlashlamp pumped solid state rod
    User InterfaceLCD color touchscreenLCD color touchscreen
    Laser activationFootswitchFootswitch
    Delivery Devices (How Supplied)Non-sterile, reusable, cleanableNon-sterile, reusable, cleanable

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states, "A clinical study was conducted to evidence that the change to the treatment parameters for the podiatry indication is safe and effective." However, it does not specify the sample size, data provenance, or whether the study was retrospective or prospective. This information would typically be detailed in the full clinical study report, which is not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this is a laser device and not an AI/ML system requiring expert-established ground truth for algorithmic performance. The clinical study would have involved physicians treating patients and evaluating outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is a laser device and not an AI/ML system. Adjudication methods refer to how disagreements among independent readers are resolved to establish a definitive ground truth for AI model evaluation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is a laser device and not an AI/ML system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a laser device and not an AI/ML system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the clinical study mentioned for the temporary increase of clear nail in patients with onychomycosis, the "ground truth" would be related to clinical outcomes data, specifically the assessment of "temporary increase of clear nail." This would be determined by a clinician based on visual inspection and possibly other diagnostic methods (e.g., microscopy, culture) if the study protocol included them, though the summary does not detail the specific assessment methods.

    8. The sample size for the training set

    Not applicable, as this is a laser device and not an AI/ML system.

    9. How the ground truth for the training set was established

    Not applicable, as this is a laser device and not an AI/ML system.