510(k) Data Aggregation

    K Number
    K063664

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-25

    (48 days)

    Product Code
    Regulation Number
    866.3390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GEN-PROBE APTIMA® Assay for Neisseria gonorrhoeae is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection of ribosomal RNA (rRNA) from Neisseria gonorrhoeae (GC) to aid in the diagnosis of gonococcal urogenital disease using the TIGRIS® DTS® Automated Analyzer or semiautomated instrumentation as specified. The assay may be used to test the following specimens from symptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens; and female and male urine specimens. The assay may be used to test the following specimens from asymptomatic individuals: clinician-collected endocervical, vaginal swab specimens; patient-collected 'vaginal swab specimens; and female and male urine. The assay is also intended for use with the testing of gynecological specimens, from both symptomatic and asymptomatic patients collected in the PreservCyt® Solution and processed with the Cytyc ThinPrep® 2000 System.

    Device Description

    Clearance of this premarket notification extends the clinical performance claims of the commercially available GEN-PROBE®APTIMA® Assay for Neisseria gonorrhoeae with the testing of gynecological specimens collected in the PreservCyt® Solution and processed with the Cytyc ThinPrep® 2000 System, for use on the TIGRIS® DTS® System. The ancillary kit for this application is commercially available as the GEN-PROBE APTIMA Specimen Transfer Kit. The components of the APTIMA Specimen Transfer Kit include: (1) a transport tube containing transport media with a penetrable cap and (2) specific instructions for use regarding decontamination and specimen processing procedures. The APTIMA Transfer Kit may only be used in conjunction with the APTIMA Assays.

    AI/ML Overview

    This document describes the GEN-PROBE APTIMA Assay for Neisseria gonorrhoeae on the TIGRIS DTS System. The goal of the submission appears to be to expand the clinical performance claims of the existing assay to include gynecological specimens collected in PreservCyt Solution and processed with the Cytyc ThinPrep 2000 System, for use on the TIGRIS DTS System.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria provided in the document are primarily related to agreement studies and analytical performance targets rather than explicit numerical thresholds for clinical sensitivity/specificity against a gold standard for the new use case (PreservCyt specimens on the TIGRIS DTS System). The clinical studies focus on demonstrating equivalence between the TIGRIS DTS System and previously validated DTS Systems.

    Acceptance Criteria (Implied)Reported Device Performance (for PreservCyt specimens on TIGRIS DTS)
    Analytical Sensitivity (Limit of Detection)
    100% positivity at 50 CFU/assay (250 fg of total GC rRNA)100% positive (95.1-100% CI) for N. gonorrhoeae rRNA spiked into post-processed PreservCyt liquid Pap specimen pool at 50 CFU/assay (250 fg). (N=60)
    Analytical Specificity
    No cross-reactivity with closely related organisms and common floraAll 24 tested culture isolates (including 17 phylogenetically related to N. gonorrhoeae) showed no cross-reactivity when tested at 1 x 10^6 cells/mL in PreservCyt liquid Pap media and Swab Transport Media on three different TIGRIS DTS Systems. The document does not explicitly state an "acceptance criteria" but the presented data indicates 100% specificity for the tested organisms. Some Neisseria species known to cross-react in other amplification assays were noted as potential cross-reactors, which is a disclaimer, not a failure.
    Specimen-Caused Inhibition
    < 5% inhibition rate for post-processed PreservCyt specimens0% inhibition (0/240) detected in 240 negative clinical post-processed PreservCyt liquid Pap specimens.
    Interference by Whole Blood
    No interference up to 10% (v/v) blood in PreservCyt specimensBackground signals remained below the assay cut-off for negative PreservCyt liquid Pap specimens with up to 10% (v/v) blood. For spiked specimens, the presence of up to 10% (v/v) blood did not interfere with the recovery of a positive signal.
    Clinical Equivalence (TIGRIS DTS vs. DTS Systems)
    Overall agreement ≥ 90% (with acceptable CI)100% Overall Agreement (93.0-100% CI) for all 51 PreservCyt specimens (40 positive, 11 negative) between DTS Systems and TIGRIS DTS System.
    Positive agreement ≥ 90% (with acceptable CI)100% Positive Agreement (91.2-100% CI) between DTS Systems and TIGRIS DTS System.
    Negative agreement ≥ 90% (with acceptable CI)100% Negative Agreement (71.5-100% CI) between DTS Systems and TIGRIS DTS System.
    Clinical Panel Equivalence
    100% agreement between TIGRIS and DTS Systems for all panel members100% agreement for all five panel members (0 fg, 25 fg, 250 fg, 2,500 fg, 25,000 fg rRNA/Assay) between TIGRIS and DTS Systems. Overall % Agreement between TIGRIS and DTS was 100% (97.2-100% CI) for the 250 fg (Low) panel member.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Analytical Sensitivity (LOD): N=60 for post-processed PreservCyt liquid Pap specimen. The data provenance is from laboratory spiking experiments.
    • Analytical Specificity: 24 culture isolates. The data provenance is from laboratory testing of known organisms.
    • Specimen-Caused Inhibition: N=240 negative clinical post-processed PreservCyt liquid Pap specimens. Data provenance is from clinical specimens presumably collected in the US (not explicitly stated for this part, but the clinical study below is multi-center in the US).
    • Interference by Whole Blood: Not specified as a precise "test set" size, but involved three negative PreservCyt liquid Pap specimen pools, tested in the absence and presence of N. gonorrhoeae and varying blood concentrations. Data provenance is from laboratory spiking experiments.
    • Clinical Specimen Study: N=51 PreservCyt specimens (34 symptomatic, 17 asymptomatic female subjects).
      • Data Provenance: Prospective, multi-center clinical study. Patients were enrolled from family planning, OB/GYN, public health, and STD clinics. While not explicitly stated, multi-center studies for FDA submissions are typically conducted in the USA.
    • Clinical Panel Study: 132 replicates (30 aliquots for each of 4 positive panel members, and 12 aliquots for the negative panel member). These were created from pooled residual negative PreservCyt specimens from female subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of "experts" in the context of establishing ground truth for the test sets in this submission.

    • Clinical Specimen Study: The "ground truth" or reference method for comparison was the AGC Assay performed on the previously validated DTS Systems, following an initial screening with FDA-cleared APTIMA COMBO 2 (AC2) Assay. The AC2 Assay results determined whether specimens were selected for the clinical specimen or panel study. This is a comparison between two automated systems rather than expert interpretation of a gold standard.
    • Clinical Panel Study: The "ground truth" for the panel members was their known N. gonorrhoeae ribosomal RNA (rRNA) concentration (spiked or not spiked).
    • Analytical Studies: Ground truth was based on known concentrations of spiked organisms or rRNA, or confirmed negative samples.

    4. Adjudication Method for the Test Set

    Not applicable in the typical sense of expert adjudication of imaging or clinical findings.

    • For the Clinical Specimen Study, the comparison was made between the results of the AGC Assay on the DTS Systems (predicate) and the TIGRIS DTS System (new system). No explicit adjudication process between discordant results from these two systems is described; rather, the agreement between them was calculated.
    • For the Analytical Studies and Clinical Panel Study, the "ground truth" was established by experimental design (e.g., spiking known quantities of analyte, using confirmed negative samples), not through adjudication of expert opinions.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This submission is for an in vitro diagnostic (IVD) assay, specifically a nucleic acid amplification test (NAAT) for Neisseria gonorrhoeae. It is a standalone diagnostic device with no "human reader" component in the interpretation of results in the way an imaging AI device would have. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable to this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, the studies presented are essentially standalone performance evaluations of the GEN-PROBE APTIMA Assay on the TIGRIS DTS System. The assay generates a qualitative (positive/negative) result without human interpretation of raw data beyond confirming valid assay run parameters. The purpose of this submission was to demonstrate equivalence of this assay on a new automated platform (TIGRIS DTS) for a new specimen type (PreservCyt).

    7. The Type of Ground Truth Used

    • Analytical Sensitivity: Known spiked concentrations of N. gonorrhoeae rRNA.
    • Analytical Specificity: Known culture isolates.
    • Specimen-Caused Inhibition and Interference: Known negative samples and known spiked concentrations (for inhibition/recovery).
    • Clinical Specimen Study: The results of the AGC Assay on the predicate DTS Systems, following initial screening with the FDA-cleared APTIMA COMBO 2 (AC2) Assay. This acts as a reference method for the comparison between the old and new platforms.
    • Clinical Panel Study: Known spiked concentrations of N. gonorrhoeae ribosomal RNA (rRNA) into confirmed negative PreservCyt specimens.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of device development or algorithm training. This is a molecular diagnostic assay, not a machine learning or AI algorithm in the conventional sense that would require a dedicated training set. The assay's parameters (e.g., cut-offs) would have been established during its initial development and validation, which is not detailed in this specific 510(k) summary, as this submission is an extension of claims for an already cleared assay.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" for an AI/ML algorithm is mentioned or applicable here, this question is not relevant to the provided documentation. The assay relies on target amplification and detection of rRNA, with predetermined analytical characteristics, rather than learning from a training dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062440

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-11-07

    (78 days)

    Product Code
    Regulation Number
    866.3390
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APTIMA® Assay for Neisseria gonorrhoeae is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection of ribosomal RNA (rRNA) from Neisseria gonorrhoeae (GC) to aid in the diagnosis of gonococcal urogenital disease. The assay may be used to test the following specimens from symptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens; and female and male urine specimens. The assay may be used to test the following specimens from asymptomatic individuals: cliniciancollected endocervical and vaginal swab specimens; and patient-collected vaginal swab specimens and female and male urine specimens. The assay is also intended for use with the testing of gynecological specimens, from both symptomatic and asymptomatic patients, collected in the PreservCyt Solution and processed with the Cytyc ThinPrep 2000 System.

    Device Description

    Clearance of this premarket notification extends the clinical performance claims of the commercially available GEN-PROBE APTIMA Assay for Neisseria gonorrhoeae to include PreservCyt liquid Pap specimens (collected and processed by the Cytyc ThinPrep 2000 Processor) as acceptable testing specimens. The ancillary kit formulated for this specific application is the commercially available GEN-PROBE APTIMA Specimen Transfer Kit. The components of the APTIMA Specimen Transfer Kit include: (1) a transport tube containing transport media with a penetrable cap and (2) specific instructions for use regarding decontamination and specimen processing procedures. The APTIMA Specimen Transfer Kit may only be used in conjunction with GEN-PROBE APTIMA Assays for the detection of Chlamydia trachomatis and/or Neisseria gonorrhoeae.

    AI/ML Overview

    The provided document describes the GEN-PROBE APTIMA Assay for Neisseria gonorrhoeae, specifically its expanded indication for use with ThinPrep Specimens. The acceptance criteria and the study proving the device meets these criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for sensitivity and specificity are not explicitly stated as numerical targets in the document. However, the study results are presented as the "Summary of Clinical Performance Data" and indicate the device's performance attributes. The general expectation for such a diagnostic assay is high sensitivity and specificity. The reported performance is based on the clinical study.

    MetricAcceptance Criteria (Implied / Not Explicitly Stated)Reported Device Performance (Overall)
    SensitivityHigh sensitivity expected92.3% (12/13)
    SpecificityHigh specificity expected99.8% (1630/1633)

    2. Sample Sizes and Data Provenance

    • Test Set Sample Size: 1,646 symptomatic and asymptomatic female subjects were evaluated in the clinical study. (Page 8)
    • Data Provenance: The data was collected from a prospective multi-center clinical study (Page 8). The study subjects were enrolled from sites with GC prevalence ranging from 0.0% to 5.0% (Page 8). While the exact country of origin is not explicitly stated, the context of an FDA submission (K062440) for a device from "San Diego, California" implies the study was likely conducted in the United States.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish ground truth for the clinical study.

    4. Adjudication Method for the Test Set

    The adjudication method used to establish the "patient infected status" (ground truth) for the clinical study was based on a reference standard involving two other tests:

    • The APTIMA Combo 2 Assay (for Chlamydia trachomatis and Neisseria gonorrhoeae).
    • The APTIMA GC Assay (performed on endocervical swab specimens). (Page 8)

    The criteria were:

    • Infected Patient Status: Both reference NAATs (APTIMA Combo 2 Assay and APTIMA GC Assay on endocervical swab) were required to be positive.
    • Non-Infected Patient Status: At least one reference NAAT was required to be negative.
    • Inconclusive: If an equivocal result was obtained from any one of the reference NAATs, the patient infected status was categorized as inconclusive, and these specimens were not included in sensitivity and specificity calculations. (Page 8)

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned. The study focuses on the performance of the device itself against a defined ground truth, not on human readers with or without AI assistance.

    6. Standalone Performance

    Yes, a standalone (algorithm only) performance study was performed. The clinical study directly evaluated the performance of the APTIMA GC Assay in PreservCyt liquid Pap specimens against a defined patient infected status (ground truth) derived from other laboratory tests, without human-in-the-loop performance described. (Page 8)

    7. Type of Ground Truth Used

    The ground truth used was based on a composite reference standard derived from the results of two other Nucleic Acid Amplification Tests (NAATs) – the APTIMA Combo 2 Assay and the APTIMA GC Assay – performed on endocervical swab specimens. Specifically, two positive reference NAATs defined an infected patient, and at least one negative defined a non-infected patient. (Page 8)

    8. Sample Size for the Training Set

    The document does not explicitly mention a separate "training set" or its sample size for model development. Diagnostic assays like this typically undergo extensive analytical validation to determine performance characteristics (e.g., limit of detection, specificity) and then clinical validation using a distinct clinical sample set. The analytical studies describe testing with various organisms and dilutions, which might be considered part of the development and refinement process, but not a formally delineated "training set" in the context of machine learning.

    9. How Ground Truth for the Training Set Was Established

    Since a distinct "training set" is not described, the method for establishing its ground truth is not provided. For the analytical studies (e.g., Limit of Detection, Analytical Specificity), organism concentrations were precisely controlled by direct comparison/dilution of clinical isolates or by spiking known concentrations of organisms/cells into samples. (Page 3-4)

    Ask a Question

    Ask a specific question about this device

    K Number
    K043144

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-03-14

    (119 days)

    Product Code
    Regulation Number
    866.3390
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview