Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K053009

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-17

    (22 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

    Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles

    Device Description

    The 0.35T Signa® Ovation with Excite Magnetic Resonance System is a modification to the 0.35T Signa® MFO/i Magnetic Resonance System (K033504) which utilizes a permanent magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 0.35T Signa® Ovation with Excite Magnetic Resonance System features a permanent magnet operating at 0.35T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the GE 0.35T Signa® Ovation with Excite MR System. However, it does not contain the detailed information necessary to answer all parts of your request regarding acceptance criteria and a study proving device performance.

    Here's what can be extracted and what is missing:


    Description of Acceptance Criteria and Study Proving Device Performance

    The submission focuses primarily on demonstrating substantial equivalence to a predicate device (Signa® 0.35T® Ovation with Excite System, K033504) through engineering modifications and adherence to performance standards, rather than a clinical performance study with defined acceptance criteria for diagnostic efficacy.

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategoryReported Device PerformanceComments
    Safety Standards AdherenceEvaluated to IEC 601-1 (General Safety for Medical Equipment)This indicates compliance with recognized safety standards for electrical and mechanical safety. It's a foundational acceptance criterion for medical devices.
    MR Specific Safety & Performance Standards AdherenceEvaluated to IEC 601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis)This indicates compliance with specific safety and performance standards relevant to MR diagnostic devices, including aspects like magnetic field safety, heating, and acoustic noise.
    General Performance Standards AdherenceEvaluated to appropriate NEMA performance standardsNEMA (National Electrical Manufacturers Association) performance standards for MR systems typically cover aspects like image quality (spatial resolution, signal-to-noise ratio, uniformity), geometric accuracy, and field strength accuracy. Specific values are not provided in this summary.
    Substantial EquivalenceConcluded to be substantially equivalent to the predicate device (K033504)The primary "acceptance criterion" for this 510(k) is demonstrating that the modified device is as safe and effective as the predicate device and does not raise new questions of safety or effectiveness. This is the overarching regulatory acceptance criterion for a 510(k) clearance.
    No New Potential HazardsUsage of the device "does not result in any new potential hazards."This is a key part of demonstrating substantial equivalence and is an acceptance criterion related to safety.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not specified. The submission does not describe a clinical test set or data derived from human subjects for performance evaluation in the way you might expect for a diagnostic AI device. The evaluation appears to be based on technical performance and safety testing against established standards, and comparison to existing device specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable. No clinical ground truth establishment is described for a diagnostic performance study. The "trained physician" is mentioned in the indications for use as interpreting the images, but not in the context of creating a test set for the device's performance.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical test set or adjudication method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is an MR system, not an AI-assisted diagnostic tool. No MRMC study is mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is an MR system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable for diagnostic accuracy. The "ground truth" for this type of submission relates to compliance with engineering and safety standards (e.g., an MR system must correctly measure field strength, produce images with certain resolution, operate safely). No diagnostic accuracy ground truth is discussed.

    8. The sample size for the training set:

    • Not applicable. This device is an MR system, not an AI model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set is relevant for this device.

    Summary of the Study Discussed:

    The actual "study" described in this 510(k) summary is focused on engineering and safety verification of the device's modifications and its adherence to relevant international and national standards.

    • Type of Study: Technical evaluation against established performance and safety standards, and a comparison to a predicate device.
    • Standards Referenced:
      • NEMA performance standards (specific ones not listed, but implied to be relevant to MR system performance).
      • IEC 601-1 (International Medical Equipment Safety standard).
      • IEC 601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis).
    • Purpose: To demonstrate that the GE 0.35T Signa® Ovation with Excite MR System, with the addition of a motorized control circuit for lateral table movement, is substantially equivalent (in terms of safety and effectiveness) to its predicate device (K033504) and meets applicable safety and performance requirements.
    • Conclusion: GE concluded that the device is substantially equivalent to the predicate and introduces no new potential hazards. FDA agreed with this determination.

    This submission is typical for modifications to existing medical imaging hardware where the fundamental imaging principles and diagnostic capability remain unchanged, and the focus is on maintaining safety and performance characteristics in the face of minor design alterations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033504

    Validate with FDA (Live)

    Date Cleared
    2003-11-25

    (20 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

    Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles

    Device Description

    The 0.35T Signa ®Ovation with Excite Magnetic Resonance System is a modification to the 0.35T Signa® MFO/i Magnetic Resonance System (K002611) which utilizes a permanent magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 0.35T Signa ®Ovation with Excite Magnetic Resonance System features a permanent magnet operating at 0.35T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.

    AI/ML Overview

    This document describes the 510(k) summary for the GE 0.35T Signa® Ovation with Excite Magnetic Resonance System. It does not contain information about acceptance criteria, device performance metrics, or a study design to meet specific acceptance criteria as would be typical for an AI/ML device submission. Instead, this is a traditional medical device submission for an MRI system, evaluating its substantial equivalence to a predicate device based on technical specifications and safety standards.

    Therefore, most of the requested information regarding acceptance criteria and study design for an AI/ML device is not applicable to this document.

    Here's an attempt to answer the questions based on the provided text, highlighting what is not applicable:

    1. A table of acceptance criteria and the reported device performance

      • Not Applicable: This document does not specify quantitative performance acceptance criteria in terms of clinical accuracy metrics (e.g., sensitivity, specificity) or report such performance for this MRI system. The evaluation focuses on meeting NEMA performance standards and IEC safety standards, and being "comparable" to a predicate device.
      • The "performance" described relates to imaging capabilities like "improved higher resolution imaging and shorter scan times" and the technical differences from the predicate, such as "addition of four independent receive channels and an update to the performance of the gradient amplifiers." These are not numerical metrics with acceptance criteria provided in the text.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not Applicable: This document does not describe a test set or data provenance for evaluating an AI/ML algorithm. The evaluation methods mentioned are compliance with NEMA and IEC standards, and comparison to a predicate device's technical specifications.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not Applicable: Ground truth for a test set is not discussed as this is not an AI/ML algorithm requiring such a validation. The interpretation of images is mentioned to be "by a trained physician," but this refers to the intended clinical use of the MRI system, not an evaluation method for the system itself as described in this submission.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not Applicable: There is no test set or adjudication method described for an AI/ML algorithm.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not Applicable: No MRMC study or AI assistance is mentioned in this document. This submission is for an MRI hardware system, not an AI/ML diagnostic aid.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not Applicable: This is an MRI system, not an algorithm, so standalone performance is not relevant in the AI/ML sense. The "standalone" performance would be its imaging capabilities, which are assessed against technical standards rather than clinical accuracy metrics.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not Applicable: Ground truth, in the context of validating an AI/ML device, is not mentioned or required for this type of medical device submission.
    8. The sample size for the training set

      • Not Applicable: There is no AI/ML algorithm described in this document, and therefore no training set.
    9. How the ground truth for the training set was established

      • Not Applicable: There is no AI/ML algorithm described in this document, and therefore no ground truth for a training set.

    Summary based on the provided text:

    The GE 0.35T Signa® Ovation with Excite Magnetic Resonance System is a modification of an existing MRI system (Signa® MFO/i MR system K002611). The evaluation described in this 510(k) summary focuses on demonstrating substantial equivalence to the predicate device.

    Key Evaluation Points mentioned:

    • Compliance with Standards: The system was evaluated against "appropriate NEMA performance standards" and "IEC 601-1 International Medical Equipment Safety standard and IEC 601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis."
    • Comparability: It is stated that the device "is comparable to the currently marketed Signa® 0.35T Ovation Magnetic Resonance System."
    • Safety: The conclusion states, "Usage of the 0.35T Signa® Ovation with Excite Magnetic Resonance System does not result in any new potential hazards."

    This submission is a regulatory filing for a hardware medical device, and the evaluation criteria and methods are typical for this type of device, focusing on safety, performance standards (like imaging parameters and resolution), and substantial equivalence to a previously cleared device, rather than diagnostic accuracy metrics or AI/ML algorithm performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1