Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K233419

    Validate with FDA (Live)

    Device Name
    GBR System
    Date Cleared
    2025-01-03

    (451 days)

    Product Code
    Regulation Number
    872.4880
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Guided Bone Regeneration (GBR) System is used to stabilize and fix bone filling materials and/or barrier membranes used for bone regeneration in the oral cavity.

    Device Description

    The Guided Bone Regeneration (GBR) System is used to stabilize and fix bone filling materials and/or barrier membranes used for bone regeneration in the oral cavity.
    MATERIALS: Only materials approved for medical purposes are used in the manufacture of products: Pure Titanium Grade 4 according to standard ISO 5832-2 and ASTM F67-13.

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) K233419 FDA clearance letter and related summary for a dental device called "GBR System". This document focuses on demonstrating substantial equivalence to existing predicate devices, primarily through non-clinical testing and comparison of technological characteristics.

    It does not contain information about acceptance criteria for a study proving device performance, nor details about a study that would meet such criteria. Therefore, I cannot extract the requested information from this document.

    The document discusses:

    • Non-clinical tests: Biocompatibility, Insertion Torque, Fracture Torque, Pull-out tests, sterilization validation, and MRI compatibility. These are primarily engineering and material characterization tests, not clinical studies with acceptance criteria for device performance in patients.
    • Comparison to predicate devices: This is done to establish substantial equivalence, not to prove that the device meets specific performance acceptance criteria from a clinical study.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170697

    Validate with FDA (Live)

    Device Name
    GBR system
    Manufacturer
    Date Cleared
    2018-05-24

    (443 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in stabilizing and fixating bone grafts, bone filling material and/or barrier membranes used for guided bone/tissue regeneration in the oral cavity. Single patient use only.

    Device Description

    The GBR System is comprised of bone screws and bone plates that are used for bone fixation surgery. The bone screws and plates are not reusable. The bone screws are composed of the titanium alloy (ASTM F136) and the bone plates are composed of the pure titanium (ASTM F67 Grade 1). The screws are offered in following range; diameter (0.85, 1.4, 2.0 mm) and length (3.0, 3.5, 4.0, 5.0, 6.0, 8.0, 10.0, 12.0 mm). Also, the tip of screw has single type; the self-drilling and self-tapping available type and the type that require the pre-drilling (Surgeon's option). The bone plate is offered in 0.1 mm thickness and in length 22 mm x 15 mm, 36 mm x 22 mm and 44 mm x 44 mm. Each device is provided non-sterile. Therefore, this device must be sterilized prior to use. Steam sterilization is recommended.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "GBR System" (K170697), a bone fixation system. The testing outlined assesses the device's substantial equivalence to predicate devices, rather than establishing acceptance criteria for a new, standalone device's performance in a diagnostic or clinical efficacy context. Therefore, direct acceptance criteria in terms of clinical performance metrics (like sensitivity, specificity, AUC) and a study proving a device meets these in the way typically seen for AI/diagnostic devices are not applicable or present in this document.

    However, I can extract the non-clinical performance testing conducted and its reported outcomes, which serve as the "acceptance criteria" for demonstrating substantial equivalence based on physical and biological properties.

    Here's the information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance Criteria (Implied)Reported Device Performance
    Mechanical PropertiesDimension Test (per ASTM F543-17)Equivalent to predicate deviceFound substantially equivalent to predicate device
    Pull Out Test (per ASTM F543-17)Equivalent to predicate deviceFound substantially equivalent to predicate device
    Torsion Test (per ASTM F543-17)Equivalent to predicate deviceFound substantially equivalent to predicate device
    Tensile Test (per ASTM F543-17)Equivalent to predicate deviceFound substantially equivalent to predicate device
    Bending Test (per ASTM F543-17)Equivalent to predicate deviceFound substantially equivalent to predicate device
    BiocompatibilityISO 10993-3, 10993-6, 10993-10, 10993-11BiocompatibleTest results and analysis found the device biocompatible
    SterilizationISO 17665-1:2006 (for gravity displacement steam sterilization)Sterility Assurance Level (SAL) of $10^{-6}$Demonstrated SAL of $10^{-6}$
    ISO 17665-1:2006 (for pre-vacuum steam sterilization)Sterility Assurance Level (SAL) of $10^{-6}$Demonstrated SAL of $10^{-6}$

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The text mentions "bench testing" and "test results and analysis" but does not specify sample sizes for these tests, nor the data provenance (e.g., country of origin, retrospective/prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable and not provided. The "acceptance criteria" here relate to physical, mechanical, and biological properties of the device, not a diagnostic or clinical performance where expert-established ground truth would be relevant.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. As above, adjudication methods apply to expert-based ground truth establishment, which is not relevant to the non-clinical bench and biocompatibility testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not mentioned and is not applicable to this type of device (bone fixation system) or the non-clinical testing described. This device is a physical implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided. This device is a physical medical device, not an algorithm, and therefore "standalone" performance in this context is irrelevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance testing, the "ground truth" is defined by the standards and specifications outlined in the referenced ASTM and ISO standards. For example, for mechanical tests, the ground truth is the expected mechanical behavior according to the standard and comparison to the predicate device. For biocompatibility, it's compliance with ISO 10993 standards. For sterilization, it's achieving the specified SAL.

    8. The sample size for the training set

    This information is not applicable and not provided. This device is a physical medical device and does not involve AI/machine learning, so there is no training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. As there is no training set for this type of device, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143730

    Validate with FDA (Live)

    Device Name
    GBR System
    Date Cleared
    2015-06-10

    (163 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GBR System is intended for use in stabilizing and fixating bone filling material and/or barrier membranes used for guided bone/tissue regeneration in the oral cavity.

    Device Description

    The GBR System is rigid fixation consisting of plates and screws in various configurations, shapes and sizes. The GBR System is made of Unalloyed Titanium and Titanium Alloy (Ti-6AL-4V). The device screw is available in three different types, Auto Screw, Tenting Screw and GBR Screw. The device plate is provided in two different types, mesh plate and GBR mesh plate. The device system also includes various manual surgical instruments, such as mesh puncher, screwdriver handle, screw block, driver shaft and drill bit. The GBR System is not provided sterile. It is required to be sterilized via autoclave method to reach a SAL of 10^-6 by the hospital prior to surgery. The GBR System is provided as silver color which is casted according to the anodizing technique while apply 5 V ~ 10 V electric energy to the titanium.

    AI/ML Overview

    The provided text describes a 510(k) summary for the "GBR System" medical device, which consists of bone plates and screws. The submission aims to demonstrate substantial equivalence to a predicate device, the "Jeil Bone Fixation System" (K050669).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a quantitative manner (e.g., minimum tensile strength), but rather implies that the device is acceptable if its performance is "equivalent" to the predicate device. The performance is assessed through non-clinical testing based on ASTM standards.

    Acceptance Criteria (Implied)Reported Device Performance
    Equivalent Driving torque per ASTM F543-13 (Screws)Testing indicates GBR System is equivalent to predicate device.
    Equivalent Axial pull-out test per ASTM F543-13 (Screws)Testing indicates GBR System is equivalent to predicate device.
    Equivalent Torsion test per ASTM F543-13 (Screws)Testing indicates GBR System is equivalent to predicate device.
    Equivalent Tensile Strength (Plates)Testing indicates GBR System is equivalent to predicate device.
    Equivalent material (Plates)Unalloyed Titanium (ASTM F67) - Same as predicate.
    Equivalent material (Screws)Titanium Alloy (Ti-6Al-4V) (ASTM F136) - Same as predicate.
    Equivalent sterilization methodNon-sterile, Steam sterilization prior to use - Same as predicate.
    Equivalent single useYes - Same as predicate.
    Equivalent intended useStabilizing and fixating bone grafts, bone filling material and/or barrier membranes used for guided bone/tissue regeneration in the oral cavity - Same as predicate.
    No significant differences that would adversely affect use"No significant differences between the Model GBR System and the predicate device that would adversely affect the use of the proposed device."

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each non-clinical test (e.g., how many screws were tested for driving torque). It states that the evaluation was based on ASTM F543-13. The data provenance is presumed to be from laboratory testing conducted by Jeil Medical Corporation. There is no mention of country of origin for the data or whether it was retrospective or prospective, as it's non-clinical lab testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this submission. The "ground truth" for the non-clinical tests is established by objective engineering measurements against ASTM standards, not by expert interpretation.

    4. Adjudication method for the test set

    This information is not applicable. Since the testing is non-clinical and objective, there is no need for adjudication by experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study done. This device is a physical medical implant (bone plates and screws) and not an AI-assisted diagnostic tool or software, so such a study would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The GBR System is a physical medical device, not an algorithm or software. Therefore, standalone algorithm performance is not a relevant concept for this submission.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" is defined by the objective measurements against the established ASTM standards for mechanical properties (driving torque, axial pull-out, torsion, tensile strength). For material, it's compliance with ASTM F67 and ASTM F136 specifications.

    8. The sample size for the training set

    This information is not applicable. This is a 510(k) submission for a physical medical device, not a machine learning model. There is no "training set."

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1