Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220329

    Validate with FDA (Live)

    Device Name
    G-CAM
    Date Cleared
    2022-08-16

    (193 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    G-CAM discs are intended to be used for the manufacture of full and partial removable dentures as well as permanent and temporary restorations such as anterior or posterior crowns and bridges, inlays, veneers, copings and substructures.

    Device Description

    G-CAM is a thermoplastic acrylic disc made by a principal base of polymethyl methacrylate (PMMA) resin doped with graphene (allotropic form of carbon), suitable for the creation of dental prostheses using CAD/CAM technology.

    G-CAM discs are intended to be used for the manufacture of full and partial removable dentures, implant overdentures as well as permanent and temporary restorations such as anterior or posterior crowns and bridges, inlays, onlays, veneers, copings and substructures.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a dental device called G-CAM. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness through extensive human studies. Therefore, the information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" will largely relate to non-clinical performance testing against recognized standards and comparison with predicate devices, as clinical human-in-the-loop studies were explicitly not performed.

    Here's a breakdown of the requested information based on the provided document:


    Acceptance Criteria and Device Performance for G-CAM

    The acceptance criteria for G-CAM are primarily based on demonstrating substantial equivalence to predicate devices through conformity with recognized international standards for dental materials.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria" in a traditional sense with numerical targets before the results. Instead, it provides a comparison table (Table 1) that lists the G-CAM's characteristics and its measured performance against the relevant ISO standards, alongside the characteristics of the predicate devices. The "acceptance" is implied by meeting or exceeding the minimum requirements of the specified ISO standards, and by showing substantial equivalence to the predicate devices.

    Table.1 Comparison of G-CAM with Predicate and Reference Devices (with performance data)

    FeatureAcceptance Criteria (from ISO Standard)Reported G-CAM PerformancePredicate (JUVORA Dental Disc) Performance (where stated)Reference (CAD/CAMouflage Milling Block) Performance (where stated)
    Flexural ModulusISO20795 > 2000 MPa3200 MPa +/-7% (ISO 20795)3995 MPa (ISO 20795)Not stated
    Flexural StrengthISO20795 > 65 MPa; ISO10477 ≥ 50 MPa140 MPa +/-7% (ISO 20795); 148 MPa (ISO 10477)165 MPa (ISO 20795); 192 MPa (ISO 10477)Not stated
    Surface HardnessNot specified directly, but comparable to predicates88 ShoreD (ISO 48-4); 19.5 KHN (ASTM E384)Not statedNot stated
    Water AbsorptionISO20795 < 32 µg/mm³; ISO10477 < 40 µg/mm³4 µg/mm³ (ISO 20795)Not stated< 7.5 µg/mm³ (ISO 10477)
    Residual MonomerISO20795 < 2.2%<0.004% (ISO 20795)Not statedNot stated
    Water SolubilityISO20795 < 1.6 µg/mm³; ISO10477 < 7.5 µg/mm³<0.5 µg/mm³ (ISO 20795)Not statedNot stated
    Compressive StrengthNot specified directly, but comparable to predicatesNot stated in this section (but mentioned for predicate as 156 MPa (ISO 5833))156 MPa (ISO 5833)Not stated
    Filler Particle SizeNot specified directly, but comparable to predicates (mentioned for reference device)21.5 µ (ISO 13320)Not statedNot stated
    BiocompatibilityConformance to ISO 7405 (and ISO 10993)Concluded as biocompatible (Cytotoxicity, Sensitization, Acute Systemic Toxicity, Subchronic Toxicity, Material-Mediated Pyrogenicity, Genotoxicity, Carcinogenicity tests were favorable)BiocompatibleBiocompatible
    Manufacturing Method(Comparability to predicate)Powder + Liquid methacrylate-based resins with graphene powder mixed and heat curedVarious plastics processing methodsPowder + Liquid methacrylate-based resins with fillers and pigments mixed and heat cured
    Material Composition(Comparability to predicate)PMMA resin doped with graphenePEEK-OPTIMA™ LT1 (PEEK)Ceramic-filled nanohybrid Polymer resin
    Shelf Life(Comparability to predicate)5 years (real-time testing)10 yearsNot specified
    Performance TestingConformance to ISO 20795-1, ISO 10477, ISO 4049Conforms with ISO 20795-1; Conforms with ISO 10477Conforms with ISO 20795-1; Conforms with ISO 10477Conforms with ISO 4049; Conforms with ISO 10477

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not explicitly state the number of physical samples (e.g., discs, test specimens) used for each specific non-clinical test (e.g., flexural strength, water absorption). The testing was performed according to recognized standards (e.g., ISO 20795-1, ISO 10477), which would dictate the appropriate sample sizes for each test.
    • Data Provenance: The tests were conducted by Graphenano Dental S.L. in Spain. The data is from non-clinical bench testing. The document does not specify if the testing was retrospective or prospective in terms of a specific study design as it's a premarket submission focusing on material properties.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This question is not applicable in the context of this 510(k) submission. The "ground truth" for material properties is established by the specifications and methodologies of the international ISO standards themselves, not by human expert consensus or interpretation of images. The tests measure physical and chemical properties directly.
    • The tests were performed by laboratories or personnel competent in conducting tests according to the specified ISO standards for dental materials.

    4. Adjudication Method for the Test Set

    • Not applicable. This concept typically applies to clinical studies where subjective human assessments (e.g., image interpretation) need to be resolved. For objective bench testing of material properties, results are quantitative and do not require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical testing is not required to demonstrate substantial equivalence as the indications for use are like legally marketed devices and the designs previously cleared under a premarket notification." and "No clinical studies are submitted."
    • Therefore, there is no effect size of how much human readers improve with AI vs. without AI assistance, as this is a dental material and not an AI-assisted diagnostic device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. This device is a dental material (G-CAM discs) intended for the manufacture of dental prostheses using CAD/CAM technology. It is not an algorithm or an AI-based system. Its performance relates to its material properties and manufacturing quality, not autonomous algorithmic outputs.

    7. Type of Ground Truth Used for Performance Testing

    • Non-clinical material properties: The "ground truth" or reference for the performance testing of the G-CAM device are the specifications and minimum requirements outlined in recognized international standards such as ISO 20795-1 (Dentistry — Base polymers — Part 1: Denture base polymers), ISO 10477 (Dentistry — Polymer-based crown and veneering materials), ISO 4049 (Dentistry — Polymer-based restorative materials), ISO 48-4 (Plastics — Determination of hardness — Part 4: Durometer method), ASTM E384 (Standard Test Method for Knoop and Vickers Hardness of Materials), ISO 13320 (Particle size analysis — Laser diffraction methods), ISO 7405 (Dentistry — Evaluation of biocompatibility of medical devices used in dentistry), and ISO 10993 (Biological evaluation of medical devices).
    • For biocompatibility, specific tests (Cytotoxicity, Sensitization, Acute Systemic Toxicity, Subchronic Toxicity by Subcutaneous Implantation, Material-Mediated Pyrogenicity, Genotoxicity, Carcinogenicity) were performed to establish the ground truth of biological safety.

    8. Sample Size for the Training Set

    • Not applicable. This device is a physical dental material, not an AI or machine learning model that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As stated above, there is no training set for this type of medical device submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1