Search Results
Found 1 results
510(k) Data Aggregation
(77 days)
The Fungitell® STAT assay is a protease zymogen-based colorimetric assay for the qualitative detection of (1->3)-B-D-glucan in the serum of patients with symptoms of, or medical conditions predisposing the patient to, invasive fungal infection. The serum concentration of (1->3)-B-glucan, a major cell-wall component of various medically important fungis, can be used as an aid in the diagnosis of deep-seated mycoses and fungemias . A positive result does not indicate which genus of fungi may be causing infection.
(1->3)-B-D-glucan index values should be used in conjunction with other diagnostic procedures, such as microbiological culture, histological examination of biopsy samples and radiological examination.
The Fungitell® STAT assay provides a qualitative measurement of (1-3)-D-Dglucan. The assay is based upon a modification of the Limulus Amebocyte Lysate (LAL) pathway. The Fungitell® STAT Reagent is modified to eliminate bacterial endotoxin reactivity and, thus, to only react to (1->3)-B-glucan, through the Factor Gmediated side of the pathway. (1->3)-3-D-glucan activates Factor G, a serine protease zymogen. The activated Factor G converts the inactive pro-clotting enzyme to the active clotting enzyme, which in turn cleaves the para-nitroanilide substrate, Boc-Leu-Gly-Arg-pNA, creating a chromophore, para-nitroaniline (pNA), that absorbs at 405 nm. The Fungitell® STAT kinetic assay is based upon the determination of the rate of optical density increase produced by a sample. This rate is interpreted against the rate of optical density increase of the Fungitell® STAT Standard to produce an index. This patient sample index value is qualitatively interpreted as a Negative, Indeterminate or Positive result according to the index value ranges provided in Table 1 below. The Fungitell® STAT Standard is calibrated at 80 +/- 8 pg/mL which is the Positive cut-off for the Fungitell® predicate.
Here's a summary of the acceptance criteria and study details for the Fungitell® STAT device, based on the provided document:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Acceptance Criteria (from predicate Fungitell®) | Reported Device Performance (Fungitell® STAT) |
|---|---|---|
| PPA (Positive Percent Agreement) | Not explicitly stated, inferred from predicate's established performance | 99.2% (95% CI: 95.4%, 99.9%) |
| NPA (Negative Percent Agreement) | Not explicitly stated, inferred from predicate's established performance | 97.6% (95% CI: 95.4%, 99.9%) |
| Precision/Reproducibility | Consistent with predicate Fungitell® assay | Intra-assay % CV: 0.4% to 26.8%; Inter-assay % CV: 11% to 20.44% (considered consistent) |
| Index Cut-off Values | Not explicitly stated as "acceptance criteria," but established based on predicate performance. | Negative: ≤ 0.74; Indeterminate: 0.75 - 1.1; Positive: ≥ 1.2 |
Study Details
-
Sample sizes used for the test set and the data provenance:
- Assay Cut-off study: 93 de-identified patient serum samples.
- 44 Positive, 15 Indeterminate, and 34 Negative based on the predicate device Fungitell®.
- Method Comparison study: 488 de-identified patient serum samples.
- 309 Negative, 36 Indeterminate, and 143 Positive based on the predicate device Fungitell®.
- Data Provenance: De-identified patient serum samples collected for routine clinical care of the intended population. Retrospective, as these were pre-existing samples. Country of origin not specified, but the study was performed at a CLIA laboratory, suggesting a US context.
- Assay Cut-off study: 93 de-identified patient serum samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The ground truth for the test set (both cut-off and method comparison studies) was established by comparing to the predicate device, Fungitell®. There is no mention of human experts directly establishing ground truth for these studies.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- No adjudication method involving human readers is mentioned for the test set. All comparisons were made against the predicate device.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was performed or mentioned. This is an in vitro diagnostic device, not an AI-assisted diagnostic.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance studies (Assay Cut-off, Method Comparison, Precision/Reproducibility) represent standalone performance of the Fungitell® STAT device. Its measurements and interpretations are independent of human-in-the-loop analysis beyond the initial sample collection and processing.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth used for performance evaluation was the results from the legally marketed predicate device, Fungitell®. The (1->3)-B-D-Glucan concentrations in the samples were distributed over the full measuring range of the predicate device's standard curve.
-
The sample size for the training set:
- The document does not describe a separate "training set" in the context of machine learning. The Fungitell® STAT is a modified biochemical assay, not an AI/ML algorithm requiring a distinct training phase.
-
How the ground truth for the training set was established:
- As there's no mention of an ML/AI model with a training set, this question is not applicable in the context of this device's development as described. The "calibration" of the Fungitell® STAT Standard (using Saccharomyces cerevisiae) was done against the predicate assay's glucan standard, which could be considered an analogous process to establishing a baseline reference.
Ask a specific question about this device
Page 1 of 1