Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K221395

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-01

    (49 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The implants of the Footmotion Plating System are intended for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults.

    Device Description

    The Footmotion Plating System consists of plates and screws, designed for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults.

    The implants of the Footmotion Plating System will be provided non sterile for sterilization by health care professionals prior to use, or provided sterile by gamma sterilization.

    The instruments of the Footmotion Plating System will be provided non sterile for sterilization by health care professionals prior to use.

    AI/ML Overview

    This FDA 510(k) summary is for a medical device called the Footmotion Plating System, which is a metallic bone fixation device. As such, the "acceptance criteria" and "device performance" are typically related to mechanical and material properties rather than diagnostic accuracy or clinical outcomes. The study described is a non-clinical, benchtop study.

    Here's the breakdown of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    Mechanical Performance4-point-bending (plates)Performance comparable to predicates (ASTM F382 standards)Subject device plates performance found to be comparable to predicates.
    Mechanical PerformanceTorsional testing (screws)Performance aligned with predicate standards (ASTM F543)Performed on subject screws (details not quantified but implies acceptable performance vs. predicates).
    Mechanical PerformanceAxial pull-out strength (screws)Performance aligned with predicate standards (Equation described by Chapman, FDA-Guidance: Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway-2020)Performed on subject screws (details not quantified but implies acceptable performance vs. predicates).
    BiocompatibilityEndotoxin testingAcceptable endotoxin levels (LAL quantitative kinetic chromogenic method)Testing performed and results considered acceptable.
    Material EquivalenceMaterial CompositionConform to predicate materials (CP Titanium per ASTM F67/ISO 5832-2; Titanium alloy Ti-6Al-4V ELI per ASTM F136/ISO 5832-3)Uses the same materials as the predicate devices.
    Design EquivalenceShape, Material, Operating PrinciplesEquivalent to predicate devicesEquivalent in terms of shape, material, and operating principles.
    Dimensional EquivalenceSizesEquivalent to predicate devicesEquivalent in their dimensions.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not explicitly state the number of plates or screws tested. However, mechanical testing typically involves a sufficient number of samples to achieve statistical significance.
    • Data Provenance: The studies are non-clinical (benchtop) tests performed by the manufacturer, Newclip Technics, as part of their 510(k) submission. There is no mention of country of origin for the data or if it's retrospective/prospective as these terms are not applicable to non-clinical bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. For mechanical and material testing of this nature, "ground truth" is established by adherence to recognized international and national standards (e.g., ASTM, ISO, FDA guidance documents) and the results of the physical tests themselves. Human experts are involved in designing and conducting the tests, interpreting the raw data, and comparing it to the performance of predicate devices, but they are not establishing a "ground truth" in the diagnostic sense.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods like "2+1" or "3+1" are relevant for clinical studies or studies involving human assessment of clinical data, especially in diagnostic accuracy claims. For benchtop mechanical testing, the results are typically quantitative and objective, directly compared against established standards or predicate device performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance

    • Not Applicable. This device is a bone fixation system, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study and the concept of "human readers improving with AI vs. without AI assistance" are not relevant to this submission.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not Applicable. This device is a physical medical implant (plating system), not a software algorithm. Therefore, "standalone algorithm performance" is not applicable.

    7. The Type of Ground Truth Used

    • The "ground truth" for the mechanical performance of the device is established by:
      • Adherence to recognized industry standards: ASTM F382 (for plate bending), ASTM F543 (for screw torsion), and FDA Guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway-2020" (for screw axial pull-out).
      • Comparison to predicate devices: The subject device's mechanical and material properties are demonstrated to be "substantially equivalent" to legally marketed predicate devices.
      • Biocompatibility testing: Endotoxin levels assessed by LAL quantitative kinetic chromogenic method.

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" in the context of this device because it is a physical hardware device, not a machine learning or AI algorithm that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set for this type of device, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171510

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-09-08

    (107 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Footmotion Plating System is intended for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults.

    Device Description

    The Footmotion Plating System consists of plates and screws, designed for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults. The implants of the Footmotion Plating System will be provided non sterile for sterilization by health care professionals prior to use or provided sterile by gamma sterilization. The instruments of the Footmotion Plating System will be provided non sterile for sterilization by health care professionals prior to use. Single use kits (Initial F MTP) contain implants and instruments or instruments only provided sterile by gamma sterilization.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Footmotion Plating System. As such, it does not contain the detailed information typically found in a study proving a device meets specific acceptance criteria in the context of AI/ML performance evaluation. This device is a medical implant system, not a software or AI-driven diagnostic tool.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size, ground truth for training set) are not applicable to this type of submission. The document focuses on demonstrating substantial equivalence to predicate devices based on design, materials, and mechanical safety, rather than evaluating performance against a specific clinical metric.

    Here's a breakdown of the information that is applicable or discernible from the provided text, along with an explanation for why other points are not present:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present acceptance criteria for performance in the typical sense of accuracy, sensitivity, or specificity that would apply to an AI/ML device. Instead, "acceptance criteria" are implied through the demonstration of substantial equivalence to legally marketed predicate devices. The performance is assessed through engineering analyses and material testing to ensure the device is safe and effective for its intended use, similar to its predicates.

    Acceptance Criteria (Implied)Reported Device Performance
    Intended Use: Similar to predicate devices."The Footmotion Plating System is substantially equivalent to the predicate devices in terms of intended use..." "The Footmotion Plating System is intended for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults." (Matches predicate's likely intended use implicitly)
    Design: Similar to predicate devices."The Footmotion Plating System is substantially equivalent to the predicate devices in terms of... design..." "The Footmotion Plating System consists of plates and screws, designed for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults." (Implies design is appropriate for stated use and therefore comparable to predicates)
    Materials: Biocompatible and mechanically suitable."Titanium alloy Ti-6Al-4V (conform to ASTM F136 and ISO 5832-3)." (Demonstrates use of a known, biocompatible, and mechanically sound material, equivalent to predicates.)
    Mechanical Safety/Performance: Equivalent to predicate devices."Engineering analyses have been conducted in comparison with the predicate devices." (Implied acceptance is that these analyses demonstrate comparable mechanical performance and safety to predicates.)
    Biocompatibility/Sterility: Meets relevant standards."LAL testing was performed to demonstrate that the subject device meets the endotoxin limit of <20 EU/device." (Demonstrates biocompatibility regarding endotoxins.) "The implants... will be provided non sterile for sterilization... or provided sterile by gamma sterilization." (Describes sterilization practices which must meet standards to be deemed safe and effective, and thus 'acceptable').

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable in the traditional sense for an AI/ML device. This is a physical implant system. The "test set" would refer to samples of the device components undergoing various engineering and material tests.
    • The document mentions "Engineering analyses," which would involve testing physical samples (e.g., fatigue testing, tensile strength testing). The specific number of samples for each type of engineering analysis is not provided in this summary.
    • Data Provenance: The analyses and testing appear to be conducted by or for Newclip Technics, a French company. The nature of the analyses (e.g., bench testing, material characterization) means the "data provenance" is derived from these tests. These are prospective tests conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not Applicable. For a physical device like this, there isn't a "ground truth" to be established by clinical experts in the same way there would be for an AI diagnostic. The "ground truth" is derived from engineering principles, material science standards, and established surgical practices for similar devices.
    • The "experts" involved would be materials scientists, biomechanical engineers, and regulatory specialists designing and evaluating the device and its testing protocols. Their specific number and qualifications are not detailed in this summary.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As there's no "ground truth" requiring expert consensus adjudication for clinical interpretations, no such method is described. The evaluation relies on objective engineering measurements against predefined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI/ML device, nor is it a device that is 'read' by human readers in the context of diagnostics. Therefore, an MRMC study is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device (implants and instruments), not an algorithm.

    7. The type of ground truth used:

    • For this device, the "ground truth" is based on:
      • Engineering Standards: Conformance to material standards (e.g., ASTM F136, ISO 5832-3 for Titanium alloy).
      • Biomechanical Principles: The "engineering analyses" would establish that the device's mechanical properties (e.g., strength, fatigue resistance for fixation of osteotomies, fractures, arthrodeses) are appropriate for its intended use, likely by comparison to predicate devices' known performance.
      • Regulatory Compliance: Meeting endotoxin limits through LAL testing.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is a physical device and does not have a "training set" in the context of AI/ML.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable.

    In summary, this 510(k) submission for the Footmotion Plating System demonstrates substantial equivalence to predicate devices through engineering analyses, material testing, and confirmation of intended use rather than through clinical performance metrics and studies typical of AI/ML software.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161448

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-31

    (159 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Footmotion Plating System is intended for arthrodeses, fractures and osteotomies fixation surgeries of the foot in adults.

    Device Description

    The Footmotion Plating System consists of a plate with as many screws as holes (slots) designed for arthrodeses, fractures and osteotomies fixation and revision surgeries of the foot in adults. The plates and screws are manufactured from titanium alloy and color anodized. The Footmotion Plating System will be provided non sterile for sterilization by health care professionals prior to use or provided sterile by gamma sterilization.

    AI/ML Overview

    This document is a 510(k) summary for the Footmotion Plating System, a metallic bone fixation device. As such, it does not describe a study proving software performance against well-defined acceptance criteria for an AI/ML medical device. It's for a physical implantable device.

    Therefore, many of the requested categories related to AI/ML device studies, such as sample size for test set and training set, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and how ground truth was established, are not applicable to this document.

    However, I can extract the information relevant to this device's non-clinical testing and acceptance criteria as described:

    1. A table of acceptance criteria and the reported device performance:

    Test TypeAcceptance Criteria (Standard)Reported Device Performance
    Comparative Fatigue Compression TestsASTM F382Subject devices met the standard (implied by conclusion of substantial equivalence to predicates that meet this standard)
    Comparative Static TestsASTM F382Subject devices met the standard
    Torsional TestsASTM F543Subject devices met the standard
    Endotoxin TestingLAL quantitative kinetic chromogenic method; max 20 EU per deviceDemonstrated devices meet the recommended maximum endotoxin level of 20 EU per device
    Water Endotoxin Level (for cleaning)Max limit 0.25 EU/mLNever reached the maximum limit of 0.25 EU/mL
    Bioburden, TOC, THC (for water)Not explicitly statedTesting performed, values not specified but implied to be acceptable for cleaning water

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    Not applicable. These are non-clinical mechanical and material tests, not a clinical study or AI/ML evaluation. The tests are performed on representative samples of the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    Not applicable. Ground truth as typically understood for AI/ML devices is not relevant for these types of mechanical and material tests. Compliance with established ASTM standards and endotoxin limits forms the 'ground truth' or acceptance criteria for these tests. Engineering analysis is performed by qualified engineers, though their specific number and qualifications are not detailed here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. This is not a study involving human interpretation or subjective assessment that would require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No. This document specifies "No clinical studies were performed." This is not an AI/ML device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    No. This is not an algorithm or AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    The "ground truth" for this device's non-clinical testing is based on:

    • Compliance with recognized industry standards (ASTM F382, ASTM F543).
    • Quantitative measurements against established limits (e.g., endotoxin levels).
    • Engineering analysis.

    8. The sample size for the training set:
    Not applicable. No AI/ML component; therefore, no training set.

    9. How the ground truth for the training set was established:
    Not applicable. No AI/ML component; therefore, no training set or ground truth for one.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1