Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230407

    Validate with FDA (Live)

    Date Cleared
    2023-07-17

    (152 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Upon intravenous administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence angiography.

    The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after: vascular, gastrointestinal, organ transplant, and plastic, micro- and reconstructive surgical procedures.

    Upon interstitial administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    The FloNavi Open Field Fluorescence Imaging System is an imaging system used in hospitals for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue transfer circulation for use in imaging during various surgical procedures.

    The FloNavi Open Field Fluorescence Imaging System has two system models: HD system and 4K system. Each system consists of the following main components: Imaging Head, Image Processing Unit, power supply cord and video cables.

    The Imaging Head may be either handheld or attached to a mechanical arm and provides illumination of the regions of a patient's body to be observed with near infrared light to excite ICG fluorescence. Alternatively, the Imaging Head provides white light illumination of the regions of a patient's body to be observed for color imaging. The cameras in the Imaging Head capture the fluorescent image under near infrared illumination or a color image under white light illumination. The Image Processing Unit receives the video signal from the Imaging Head and processes and outputs the video image to a medical grade video monitor and/or video recorder. Adjustments to the operation of the FloNavi Open Field Fluorescence Imaging System are possible through switches at either the Imaging Head or the Image Processing Unit.

    The FloNavi Open Field Fluorescence Imaging System is a reusable device and provided nonsterile. Its components should be cleaned and low-level disinfected prior to the first use and after every subsequent use.

    AI/ML Overview

    The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion and for visualization of the lymphatic system. It is a Class II device.

    Here's an analysis of the acceptance criteria and supporting study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the device as being "Substantially Equivalent (SE) to the predicate device" (SPY Portable Handheld Imaging (SPY-PHI) System, K200737). The acceptance criteria are implicitly linked to the predicate device's performance and the subject device's ability to meet its own design specifications and relevant safety standards.

    Acceptance Criteria CategorySpecific Criteria (Implicit for SE, based on predicate)Reported Device Performance
    SafetyCompliance with relevant electrical, EMC, and photobiological safety standards.The device complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, and IEC 62471.
    PerformanceAbility to capture and display fluorescence and white light images; functional operation as intended.Performance testing demonstrated that the proposed system performs according to specifications and functions as intended.
    Software ValidationSoftware meets "Moderate level of concern" requirements per FDA guidance.Software was validated as "Moderate level of concern" in accordance with FDA guidance documents.
    Biocompatibility(Not explicitly mentioned, but typically required for devices with patient contact)(Not explicitly mentioned in the provided text.)
    Sterility(Not explicitly mentioned, but the device is stated as reusable and non-sterile, implying reprocessing procedures are defined.)The device is provided nonsterile. Components should be cleaned and low-level disinfected prior to first use and after every subsequent use.

    2. Sample Size and Data Provenance for Test Set

    The document mentions "Performance testing and animal testing" were conducted.

    • Sample Size:
      • Performance Testing: Not specified.
      • Animal Testing: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth - Test Set

    Not applicable/Not specified. The document does not describe a clinical study involving human readers or expert-established ground truth for the "test set" in the context of diagnostic accuracy. The testing primarily focuses on the device's functional and safety performance, as well as animal studies.

    4. Adjudication Method for Test Set

    Not applicable/Not specified. No such adjudication method is mentioned as the studies described appear to be non-clinical performance and animal studies, not reader studies for diagnostic accuracy.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention a MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device appears to be an imaging system, not an AI-powered diagnostic algorithm enhancing human reading.

    6. Standalone (Algorithm Only) Performance Study

    No. The document presents the FloNavi Open Field Fluorescence Imaging System as a standalone imaging device, not an AI algorithm. The performance testing and animal testing evaluate the device itself.

    7. Type of Ground Truth Used

    For the performance and animal testing mentioned, the "ground truth" would likely be based on:

    • Engineering Specifications: For the technical performance aspects (e.g., image resolution, light output).
    • Physiological Response: In animal testing, the "ground truth" for fluorescence imaging would be the actual blood flow, tissue perfusion, or lymphatic system visualization as observed and validated within the animal model, potentially through direct observation or comparison with established methods.

    8. Sample Size for Training Set

    Not applicable. This device is presented as an imaging system, not an AI/machine learning algorithm requiring a separate training set.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no mention of an AI/machine learning component requiring a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1