Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K221798

    Validate with FDA (Live)

    Date Cleared
    2023-05-04

    (317 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fingertip Pulse Oximeter are intended for measuring function of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as not checking in home and professional caring environment. It is designed for fingers between 7mm and 25.4mm in diameter (0.28 inches to 1.0inches). And it is not intended to be used under motion or low perfusion scenarios.

    Device Description

    The Fingertip Pulse Oximeter X1609P is a fingertip device, which can measure the arterial SpO2 and pulse rate value and can display the results to the user. Moreover, the X1609P also has the function of low battery voltage indicate and automatically power off. The power source is 2xAAA batteries. Fingertip Pulse Oximeter X1609P is small in volume, light in weight and convenient in carrying. It is composed of light signal driving circuit, preamplification circuit, low pass filtering circuit, DC bias circuit, gain control circuit, display control circuit, and main control circuit. The device is for prescription. It is neither for life-supporting nor for implanting. It does not contain any drug or biological product and it does not need to be sterile.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) summary for the Finger Pulse Oximeter, Model: X1906P, organized to address your specific points about acceptance criteria and study proof.

    It's important to note that this document is a summary for a 510(k) submission, and therefore provides high-level information. Full details of the protocols, raw data, and statistical analyses would be found in the complete submission documents, which are not provided here.


    Acceptance Criteria and Device Performance

    The core of the performance evaluation for a pulse oximeter revolves around its accuracy in measuring oxygen saturation (SpO2) and pulse rate (PR). The key standard referenced is ISO 80601-2-61: 2017 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of Fingertip Pulse Oximeter Equipment. The acceptance criteria are typically derived from this standard.

    1. Table of Acceptance Criteria and Reported Device Performance:

    ParameterAcceptance Criteria (from ISO 80601-2-61)Reported Device Performance (Model: X1906P)
    SpO2 Accuracy70% to 99% range: $\pm$ 2% ARMS (as per ISO 80601-2-61)"The measure result between the SpO2 measured by subject device and the SaO2 measured by the blood gas analyzer... were conformed with the requirements of this clinical trial and related standards." (Implicitly meets $\pm$ 2% within 70-99% range as per standard)
    Pulse Rate AccuracyAs per ISO 80601-2-61, Clause 201.12.1.104 (typically $\pm$ 3 bpm or $\pm$ 2% for specific ranges)"Pulse Rate Accuracy meets the requirements defined in ISO 80601-2-61, Clause 201.12.1.104." (Implicitly meets the standard's criteria)

    Note on Acceptance Criteria Interpretation: The document states that "Bench testing was conducted, and the results show that the subject device complies with the ISO 80601-2-61: 2017... standard. And Pulse Rate Accuracy meets the requirements defined in ISO 80601-2-61, Clause 201.12.1.104." This implies the acceptance criteria were the specific performance requirements outlined in that ISO standard for accuracy, which is typically Average Root Mean Square (ARMS) difference for SpO2. The specific numerical ARMS values are not explicitly stated for the device's performance, but rather that it "conformed with the requirements."


    Study Details

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 13 adult volunteers.
      • Breakdown: 5 males, 8 females.
      • Fitzpatrick skin types included: 3 subjects with Fitzpatrick VI, 3 subjects with Fitzpatrick II, and 7 subjects with Fitzpatrick IV. This inclusion of diverse skin tones is crucial for pulse oximetry accuracy.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study. Given the manufacturer (Shenzhen Changkun Technology Co., Ltd. from China) and consultant (Shenzhen Reanny Medical Devices Management Consulting Co., Ltd. from China), it is highly probable the study was conducted in China.
    • Retrospective or Prospective: The study description "Clinical studies were conducted to verify the accuracy of proposed device" and "There were 13 adult volunteers to validated the accuracy of finger pulse oximeter" indicates a prospective clinical study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not mention the use of experts to establish ground truth in the traditional sense of image adjudication or clinical interpretation, as this is a pulse oximeter accuracy study. The "ground truth" for SpO2 and pulse rate is established through direct, highly accurate physiological measurements:

    • SpO2 Ground Truth: Measured by a blood gas analyzer (SaO2).
    • Pulse Rate Ground Truth: Measured by a patient monitor (HR).

    These are considered gold standards for these physiological parameters. Therefore, the concept of "experts establishing ground truth" as it applies to subjective interpretation (like radiology reads) is not directly applicable here. The expertise lies in the certified laboratory/clinical personnel performing the blood gas analysis and operating the patient monitor, ensuring proper calibration and technique. Specific qualifications of these operators are not detailed but are assumed to be standard for clinical laboratory and patient monitoring practices.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of this study. This study directly compares the device's numerical output against gold-standard physiological measurements (blood gas analyzer for SaO2, patient monitor for HR). There is no "adjudication" of the device's readings by human experts in the sense of consensus reading or interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging devices where human readers interpret medical images with and without AI assistance. For a pulse oximeter, the study design is centered on direct numerical accuracy against a gold standard, not human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done:

    Yes, the clinical accuracy study described for the pulse oximeter is effectively a standalone study. The device's measurement (SpO2 and PR) is directly compared to the reference measurement (blood gas analyzer and patient monitor) without an explicit "human-in-the-loop" component influencing the device's output or the ground truth measurement. The user of the oximeter reads the displayed value, but the study validates the intrinsic accuracy of the measurement.

    7. The Type of Ground Truth Used:

    • SpO2 Ground Truth: Physiological measurement via arterial blood gas analysis (SaO2). This is considered a gold standard for true arterial oxygen saturation.
    • Pulse Rate Ground Truth: Physiological measurement via a patient monitor (HR). This is also considered a gold standard for heart rate.

    8. The Sample Size for the Training Set:

    The document does not provide information on a training set size. For a traditional medical device like a pulse oximeter, "training set" is usually not a relevant concept in the same way it is for AI/ML algorithms. The device's underlying principles (Lambert Beer Law, photo-plethysmography) are based on established physics and physiology, not on learning from a dataset. Any internal calibration or algorithm refinement would be part of the engineering design process, not typically referred to as a "training set" in the context of FDA submissions for such devices.

    9. How the Ground Truth for the Training Set was Established:

    As there is no mention of a "training set" in the context of AI/ML, this question is not applicable. The device's design and calibration are based on established scientific principles and engineering practices, not on a machine learning training dataset. The clinical study validates the final, produced device's performance against established physiological ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163135

    Validate with FDA (Live)

    Date Cleared
    2017-09-19

    (315 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).

    Device Description

    The Fingertip Pulse Oximeteris intended for spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of adult and pediatric patients in the home and hospital.

    The fingertip pulse oximeter features a small size, low power consumption, a convenient operation, and portability.It is only necessary for a patient to put one of his/her fingers into the fingertip clips for measurement.

    Principle of the fingertip pulse oximeter as follows:

    A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and infrared zones.

    Operation principle of the instrument:

    Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. Relevant data is shown on the Oximeter's display through electronic circuits and a microprocessor.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The core acceptance criteria for the Fingertip Pulse Oximeter relate to its accuracy in measuring SpO2 and Pulse Rate.

    Acceptance CriterionReported Device Performance
    SpO2 Accuracy±2%: (70%~100%)
    Pulse Rate±2 bpm

    Note: For SpO2, no definition is provided for the 0%~69% range, implying the device is not specified to be accurate in this range.

    The device's performance is stated to meet the criteria specified in ISO 80601-2-61.

    Study Details

    The assessment of the device's accuracy was conducted through a clinical study described as a "Clinical hypoxia accuracy testing (controlled desaturation study)."

    1. Sample Size used for the test set and Data Provenance:

      • Sample Size: 12 subjects were enrolled.
      • Total Data Sets Collected: 294
      • Valid Data Sets: 288 (6 data sets less than 70% SpO2 were excluded).
      • Data Provenance: The study was conducted during "induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned subjects in an independent research laboratory." This indicates a prospective study design. The country of origin is not explicitly stated, but the manufacturer and consultant are based in China.
    2. Number of experts used to establish the ground truth for the test set and their qualifications:
      The text does not explicitly state the number of experts used to establish ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience"). However, the ground truth was established by a "CO-oximeter (control device)" from arterial blood samples. This implies that medical professionals (e.g., lab technicians, clinicians) would have been involved in collecting and analyzing these samples, but their specific roles or qualifications are not detailed.

    3. Adjudication method for the test set:
      The text does not mention an adjudication method (like 2+1, 3+1). The "ground truth" was directly compared from the CO-oximeter.

    4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:
      No, an MRMC comparative effectiveness study was not done. The study focused on the standalone performance of the device against a ground truth, not on how human readers' performance improved with or without AI assistance.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
      Yes, a standalone performance study was done. The "measured arterial hemoglobin saturation value (SpO2) of the subject device" was directly compared to the arterial blood samples, indicating an assessment of the device's accuracy on its own.

    6. The type of ground truth used:
      The ground truth was established using arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter (control device). This is considered a highly accurate and direct measurement suitable for establishing ground truth for oxygen saturation.

    7. The sample size for the training set:
      The document does not provide information about a specific training set or its size. This submission is for the device's performance evaluation, implying the device (or its underlying algorithms) would have been developed and "trained" prior to this evaluation.

    8. How the ground truth for the training set was established:
      As no information about a training set is provided, how its ground truth was established is also not available in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123801

    Validate with FDA (Live)

    Date Cleared
    2013-05-28

    (169 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PM601 Finger pulse oximeter is intended for spot-checking oxygen saturation in blood (SpO2) and pulse rate. The pulse oximeter is used on adults at hospital, clinics, and/or home. Not for continuously monitoring.

    Device Description

    The proposed device, Finger Pulse Oximeters PM601 is a fingertip device, which can display % SpO2. pulse rate value and pulse strength. It is based on digital blood oxygen. Power consumption of the proposed device is low and the two originally-equipped AAA alkaline batteries can be operated continuously for 24 hours. It will automatically turned off when no signal is detected for more than 8 seconds. And low voltage warning will be displayed and battery symbol flash when battery voltage is low.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Predicate Device)Reported Device Performance (PM601)
    SpO2 measurement range70%-100%70%-100%
    SpO2 accuracy70%~100%, ±3%< 3% (from clinical study), ±3% (from comparison)
    PR measurement rangeSimilar (Predicate: 30-235 bpm)30-199 bpm
    PR accuracy±2 bpm or ±2% MAX±2 bpm or ±2% MAX

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: 10 healthy volunteers
    • Data Provenance: Not explicitly stated, but the study was conducted according to a European standard (ISO9919:2005 Annex EE.2), which might suggest it was conducted outside the US, possibly in China where the manufacturer is located, or by a contract research organization adhering to international standards. The text does not specify if it was retrospective or prospective, but the description of "performed according to Annex EE.2 Procedure for invasive laboratory testing" suggests a prospective study.

    3. Number of Experts and Their Qualifications for Ground Truth

    • The document does not mention the use of experts to establish ground truth for the test set. For a pulse oximeter, "ground truth" for SpO2 accuracy is typically established through direct measurement of arterial blood oxygen saturation (SaO2) via co-oximetry, which is considered the gold standard. The text mentions "invasive laboratory testing," which is consistent with this method.

    4. Adjudication Method

    • Not applicable. The ground truth for pulse oximetry accuracy is established by a co-oximeter directly measuring SaO2, not by expert consensus or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This device is a directly measuring medical device, not an AI-assisted diagnostic tool that would typically involve human readers.

    6. Standalone Performance Study (Algorithm Only)

    • Yes, a standalone performance study was done. The clinical trial focused solely on the accuracy of the PM601 Finger Pulse Oximeter in measuring SpO2 and pulse rate against a 'gold standard' (implied by "invasive laboratory testing" for SaO2), without human interpretation in the loop.

    7. Type of Ground Truth Used

    • The ground truth used was established through "invasive laboratory testing" according to ISO9919:2005 Annex EE.2. In the context of pulse oximetry, this procedure involves inducing varying levels of hypoxia in volunteers and simultaneously measuring arterial blood oxygen saturation (SaO2) using a co-oximeter (the gold standard) and the device under test.

    8. Sample Size for the Training Set

    • Not applicable. This device is a hardware-based medical instrument that directly measures physiological parameters, not a machine learning model that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K081285

    Validate with FDA (Live)

    Date Cleared
    2008-06-05

    (30 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nonin® Onyx II® Model 9560 Finger Pulse Oximeter is a small, lightweight, portable, device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of patients who are well or poorly perfused. It is intended for spot-checking of adult and pediatric patients on fingers (other than the thumb) between 0.3 - 1.0 inch (0.8 -- 2.5 cm) thick. The index finger is the recommended site.

    Device Description

    The Model 9560 is a small, lightweight, portable finger pulse oximeter with Bluetooth® communication which can be integrated into a telemedicine system or other health data collection system through the wireless connection. The device displays numerical values for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate by measuring the absorption of red and infrared (IR) light passing through perfused tissue. Changes in the absorption caused by the pulsation of blood in the vascular bed are used to determine oxygen saturation and pulse rate. Light emitting diodes (LEDs) are contained within the sensor along with the photo detector, which is on the opposite side of the probe from the LEDs. The SpO2 and heart rate are displayed on the LED digital displays contained within the finger clip sensor. A tricolor LED display provides a visual indication of the pulse quality signal, while blinking at the corresponding pulse rate. This display changes colors to alert you to changes in pulse quality that may affect the readings: green indicates a good pulse quality signal, yellow indicates a marginal pulse quality, and red indicates as inadequate pulse signal. All associated electronics and the microprocessor are within the sensor, which is activated by inserting a patient's finger. This allows the power to be applied to all the internal circuitry upon activation. The Model 9560 uses 2 "AAA" disposable alkaline batteries for power, which shall provide 600 spot-checks. The 9560 requires no routine calibration or maintenance other than replacement of the alkaline batteries.

    AI/ML Overview

    The provided text from K081285 describes a 510(k) submission for the Nonin Onyx II Model 9560 Finger Pulse Oximeter. While it states that the device "has successfully undergone both bench and human testing to support the determination of substantial equivalence," it does not provide explicit acceptance criteria or detailed study results and methodologies to definitively prove the device meets specific acceptance criteria. The document focuses on establishing substantial equivalence to predicate devices based on accuracy, functional design, and operating principles.

    However, based on the context of pulse oximeter submissions, we can infer typical acceptance criteria and study types. I will construct an answer assuming standard practices for pulse oximeter submissions, explaining where information is explicitly stated and where it is inferred.

    Inferred Acceptance Criteria for Pulse Oximeters based on FDA Guidance:

    For pulse oximeters, key performance metrics typically involve accuracy for oxygen saturation (SpO2) and pulse rate (PR).

    • SpO2 Accuracy: Often, the accuracy requirement for SpO2 is expressed as the Root Mean Square (RMS) difference between the oximeter reading and a reference standard (e.g., arterial blood gas co-oximetry). A common range for this is ≤ 3% ARMS (Accuracy Root Mean Square) across the specified SpO2 range (e.g., 70-100%).
    • Pulse Rate Accuracy: Similarly, pulse rate accuracy is typically assessed against an ECG reference. A common requirement is ± 3 bpm (beats per minute).

    Study and Performance Details from the Provided Text (and inferred standards):

    Given the limitations of the provided document, the table below will combine explicitly stated information with common expectations for pulse oximeter 510(k) submissions.


    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Inferred from industry standards/FDA guidance)Reported Device Performance (Inferred from "substantially equivalent" claim)
    SpO2 Accuracy≤ 3% ARMS (Accuracy Root Mean Square) for SpO2 70-100%Performance comparable to predicate devices (K053130, K043359), implying it meets or exceeds this criterion. (Not explicitly stated in numbers).
    Pulse Rate± 3 bpmPerformance comparable to predicate devices, implying it meets or exceeds this criterion. (Not explicitly stated in numbers).
    Functional EquivalenceTo operate safely and effectively, providing SpO2 and PR readings for well or poorly perfused patients."substantially equivalent to the predicate devices in terms of accuracy, functional design and principles of operation."
    SafetyNo new questions of safety raised compared to predicate devices."Performance test results do not raise new questions of safety and effectiveness when compared to the legally marketed devices."

    Note: The document explicitly states "Nonin's Model 9560 Finger Pulse Oximeter has successfully undergone both bench and human testing to support the determination of substantial equivalence." This implies that the device met internal acceptance criteria for these tests, which would align with the inferred clinical accuracy requirements for pulse oximeters. However, the specific numerical acceptance criteria and the quantitative results are not disclosed in this summary.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. For clinical accuracy studies of pulse oximeters, the FDA typically recommends a minimum of 10 healthy adult subjects in a desaturation study to obtain sufficient data points across the hypoxic range (70-100% SpO2).
    • Data Provenance: The document does not specify the country of origin of the data. It states "human testing," which usually implies prospective clinical studies where subjects are enrolled and data collected for the specific purpose of the device evaluation. There is no indication of retrospective data use.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable/Not explicitly stated. For pulse oximeter accuracy studies, ground truth for SpO2 is typically established using a laboratory co-oximeter to analyze arterial blood samples, not through expert review of data. For pulse rate, an ECG machine provides the ground truth.
    • Qualifications of Experts: N/A for establishing SpO2/PR ground truth in this context. Medical professionals (e.g., anesthesiologists, nurses) would oversee the subject desaturation protocol and blood draws, but they are not "experts" establishing a consensus ground truth like in image interpretation.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Adjudication methods (like 2+1 or 3+1) are typically used when subjective interpretations are involved, such as in clinical trials requiring event confirmation or in diagnostic imaging studies where multiple readers interpret images to establish a consensus ground truth. For pulse oximeter accuracy, the ground truth is objective (co-oximetry and ECG), so no adjudication of device readings or reference measurements is needed in the same way.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • MRMC Study: No, not applicable. MRMC studies are designed for diagnostic devices where human readers (e.g., radiologists, pathologists) interpret cases (e.g., medical images) to diagnose conditions, and the study evaluates the impact of an AI algorithm on their performance. The Nonin Onyx II Model 9560 is a direct measurement device; it does not involve human interpretation of complex "cases" that an AI would assist with. Therefore, an MRMC study is not relevant for this type of device.
    • Effect Size of Human Readers with/without AI assistance: N/A, as no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, implicitly. The core "human testing" mentioned for a pulse oximeter involves directly comparing the device's readings (SpO2 and pulse rate) against the objective reference standards (arterial blood gas co-oximetry for SpO2 and ECG for pulse rate). This is a standalone evaluation of the device's algorithm and sensor performance in producing accurate readings. The device is designed to display these measurements directly, not to provide an AI-driven interpretation that then guides a human.

    7. The Type of Ground Truth Used

    • Type of Ground Truth:
      • For SpO2: Arterial blood gas co-oximetry. This is the gold standard for determining arterial oxygen saturation (SaO2), against which pulse oximeters' SpO2 readings are compared.
      • For Pulse Rate: Electrocardiogram (ECG). An ECG provides an accurate measure of heart rate, which serves as the reference for the pulse oximeter's pulse rate measurement.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable/Not explicitly stated. The document describes a traditional medical device (pulse oximeter) that uses fixed algorithms and signal processing, not a machine learning or AI-based algorithm that typically requires a distinct "training set" in the common sense. Therefore, the concept of a separate training set, as understood in AI/ML, doesn't directly apply here. The device's algorithms are likely developed and refined using engineering principles and historical physiological data rather than a specific "training set" for a submission study.

    9. How the Ground Truth for the Training Set was Established

    • How Ground Truth for Training Set was Established: Not applicable. As explained above, for this type of device, a "training set" as commonly defined in AI/ML is not relevant. The device's underlying physiological models and calibration would have been established through extensive research and development, likely using various physiological data and established scientific principles, rather than a single "training set" with an explicitly established ground truth for regulatory purposes.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1