510(k) Data Aggregation

    K Number
    K132923

    Validate with FDA (Live)

    Date Cleared
    2014-01-03

    (107 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bonding all classes of direct composite restorations as well as for indirect procedures involving metal, porcelain, or composite crowns, inlays or onlays. The Angelus products multi-purpose systems also bond amalgam and self-cure composite can be used to bond orthodontic brackets to enamel.

    Device Description

    The Fusion Single Link by Angelus is a versatile system for dental bonding. It is a onebottle adhesive system, indicated for restorations, adhesive cementation of indirect restorations (cast metal, porcelain and composite resin veneers) and intraradicular posts (cast metal and prefabricated in metal or fiber). It can also be used for the bonding of orthodontic brackets, and for the repair of porcelain restorations.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Fusion Single Link device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Predicate Device K962785 3M Dent System)Reported Device Performance (Fusion Single Link)
    Film thickness
    Enamel adhesive layer6.40 (2.80) µm8.2 (2.40) µm
    Dentin adhesive layer4.22 (1.25) µm8.4 (2.6) µm
    Application Time57 seconds60 seconds
    Light cure time10 seconds12 seconds
    Bond to Dentin45 MPa37 MPa
    Indications for UseBonding all classes of direct composite restorations, indirect procedures (metal, porcelain, composite crowns, inlays, onlays), amalgam, self-cure composite, orthodontic brackets to enamel, root desensitization.Bonding all classes of direct composite restorations, indirect procedures (metal, porcelain, composite crowns, inlays, onlays), amalgam, self-cure composite, orthodontic brackets to enamel.

    Note: The document states that "Based on performance data according the ISO/TS 11405: 2003, it was found that the average strength and average resistance of the two devices are substantially equivalent." This implies that despite some numerical differences, the overall performance is considered comparable for the purpose of substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set. It mentions "performance data" according to ISO/TS 11405: 2003, but not specific numbers of samples or cases.

    The data provenance is not specified beyond being "performance data according the ISO/TS 11405: 2003". It doesn't indicate country of origin or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The study relies on standardized testing methods (ISO/TS 11405: 2003) to measure bond strength and other properties, rather than expert-established ground truth in a clinical or imaging context.

    4. Adjudication Method for the Test Set

    Not applicable as the study uses objective measurements based on ISO/TS 11405: 2003, not expert review or consensus for ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study compares the performance of the new device to a predicate device based on material properties and bond strength, not on clinical interpretation or reading by human experts.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This concept is not applicable to the device described. The device is a dental adhesive, not an algorithm or AI system. The performance data presented refers to the physical and chemical properties of the adhesive itself.

    7. Type of Ground Truth Used

    The "ground truth" for this study is based on objective measurements of material properties and bond strength as defined by the ISO/TS 11405: 2003 standard. This standard provides guidelines for testing adhesion to tooth structure, including tensile and shear bond strength measurements.

    8. Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no "training set" in that sense. The study involves laboratory testing of material properties.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as points 8 explains.