Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220531

    Validate with FDA (Live)

    Device Name
    FOX 810, FOX 980
    Manufacturer
    Date Cleared
    2022-09-01

    (189 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FOX 810:
    Indicated for surgical applications requiring the ablation, vaporization, hemostasis, or coagulation of soft tissues in medical specialties including dermatology, dentistry, gastroenterology, general surgery, otolaryngology, and pulmonology.

    FOX 980:
    Indicated for surgical applications requiring the ablation, vaporization, hemostasis, or coagulation of soft tissues in medical specialties including dermatology, dentistry, gastroenterology, general surgery, genitourinary, gynecology, neurosurgery, otolaryngology, orthopedics, pulmonology, and thoracic surgery.

    Device Description

    The FOX Lasers are a family of products with a laser diode as the beam source. Dependent on the chosen diode, the laser system can radiate one factory set wavelength with either 810nm or 980mm. The FOX is a compact diode laser with a high-resolution color touchscreen for user control.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA for the FOX 810 and FOX 980 Diode Laser Systems. It establishes substantial equivalence to predicate devices for surgical applications involving soft tissue. There is no information regarding AI/ML in this document. Therefore, I will answer the questions based on the provided text, primarily focusing on general device performance criteria where applicable.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in terms of specific performance metrics for the device (e.g., incision depth, coagulation time) that were then "reported" as meeting those criteria. Instead, it details that the device has been tested against various IEC and EN standards to ensure safety and essential performance, and it performs comparably to its predicate devices in terms of power output and tissue interaction mechanisms.

    Performance CharacteristicAcceptance Criteria (Implicit from Predicate Equivalence)Reported Device Performance
    Device SafetyCompliance with recognized international standards for medical electrical equipment, laser safety, electromagnetic compatibility, usability, and software lifecycle processes to ensure basic safety and essential performance. (IEC 60601-1, IEC 60825-1, EN 60601-1-2, IEC 60601-2-22, IEC 60601-1-6, IEC 62304)The FOX systems have been tested against IEC 60601-1:2005+ COR1:2006, COR2:2007, AMD1:2012; IEC 60825-1:2014; EN 60601-1-2:2014; IEC 60601-2-22:2007+A1:2012; IEC 60601-1-6:2010/AMD1:2013; and IEC 62304:2006 + A1:2015. The submission concludes that the new devices are "as safe and effective as the predicate devices."
    Tissue InteractionSimilar wavelengths (810nm and 980nm) and power outputs to the predicate devices, ensuring comparable tissue interactions (ablation, vaporization, hemostasis, coagulation) based on absorption in corresponding chromophores. The predicate devices' performance in terms of incision depth and thermal tissue damage sets the implicit benchmark for acceptable tissue interaction.The device under evaluation uses 810nm and 980nm wavelengths, sharing the same wavelengths as the primary predicates. The power outputs are: FOX 810: Up to 8 W, FOX 980: Up to 12 W. This is compared to primary predicate Fox Q-810 (Up to 7W) and Fox Q-980 (Up to 9W), and secondary predicate XPulse Pro (810nm: Up to 8W, 980nm: Up to 12W). The document states that the power difference "does not absolutely affect the extent of incision depth and thermal tissue damage" and that "Power alone or pulse length alone do not alter the tissue effect because for every pulse the power is set with a pulse length. These parameters, power and pulse length, equal the pulse energy when multiplied." The submission concludes that the subject devices are substantially equivalent to the predicate devices in terms of safety and efficacy.
    Operational ModesCapability for continuous wave, pulsed, and single pulse operation, along with comparable pulse length/duration and frequency ranges to predicate devices.The subject device operates in Continuous wave, pulsed, single pulse mode. Pulse length/duration: 100μs – 45s. Pulse frequency: 0.01 – 5000 Hz. These are comparable to the predicate devices which have similar modes, and ranges of 2ms - 30s for pulse length/duration and 0.016 – 250 Hz for pulse frequency (primary predicate) or 100 μs to 60 s for pulse length/duration and 0.008 Hz to 5000 Hz for pulse frequency (secondary predicate).
    Physical and Functional CharacteristicsSimilar design, user interface (touchscreen/buttons), aiming beam specifications, cooling method (air), delivery devices (optical fibers), main power supply, dimensions, and weight to predicate devices, ensuring equivalent functionality and usability.The subject device features a touchscreen and buttons for user interface, air cooling, 532nm or 650nm @ <1mW aiming beam, and uses 300μm, 400μm and 600μm optical fibers. These are consistent with or an improvement upon the predicate devices' specifications, which had either a display with a rotary knob or a touch screen control, similar aiming beam wavelengths/power, air cooling, and similar fiber sizes. Dimensions and weight are also comparable.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document makes no mention of a "test set" in the context of clinical or observational data. The device's performance was evaluated through bench testing against recognized standards and by comparison to predicate device specifications. Therefore, there is no sample size of patients/cases or data provenance information.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. As there was no clinical "test set" or data analysis requiring ground truth establishment by experts, this information is not provided. The substantial equivalence relies on technical specifications and performance against industry standards rather than expert consensus on clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser surgical instrument, not an AI/ML-driven diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is irrelevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a hardware laser system. There is no mention of an algorithm or AI component that would have standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance and safety is established by adherence to recognized international standards (e.g., IEC 60601 series, IEC 60825-1, IEC 62304, EN 60601-1-2) which define accepted levels of electrical safety, laser safety, EMC, usability, and software quality. Additionally, the technical specifications and established performance of legally marketed predicate devices serve as a benchmark for equivalence in tissue interaction.

    8. The sample size for the training set

    Not applicable. This document pertains to a hardware medical device (laser system) and does not describe any AI/ML model for which a training set would be required.

    9. How the ground truth for the training set was established

    Not applicable. As there is no AI/ML model or training set, this question is not relevant to the provided documentation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1