Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K110268

    Validate with FDA (Live)

    Date Cleared
    2011-05-31

    (123 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ForceTriad is a full-featured electrosurgical generator intended for open and laparoscopic surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating), coagulation (hemostasis, coagulating, or desiccating), or vessel sealing or fusing.). The generator is intended for use in general, laparoscopic, and gynecologic surgical procedures where ligation of vessels, pulmonary vasculature, or lymph vessels, is desired. The system creates a vessel ligation (seal) by the application of bipolar electrosurgical RF energy (coagulation) to vessels interposed between the jaws of the device. The generator can also be used with standard bipolar devices where bipolar cutting or coagulation is desired or with compatible resectoscopes for endoscopically controlled removal (resection) or coagulation of tissue using 0.9% NaCl solution (saline) as the irrigation medium.

    The indications for use include general (including urologic, thoracic, plastic, and reconstructive), laparoscopic, and gynecological procedures where electrosurgical cutting and coagulation of tissue, and sealing (fusion) of vessels, including pulmonary vessels, and tissue bundles is performed, including such procedures as bowel resections, hysterectomies (both vaginal and abdominal), laparoscopic cholesystectomies, laparoscopically assisted vaginal hysterectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectorny, etc. The devices can be used on vessels (arteries, veins. pulmonary arteries, pulmonary veins, lymph) up to 7mm and bundles as large as will fit in the jaws of the instruments.

    The LigaSure tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.

    Device Description

    The ForceTriad™ Electrosurgical Generator is a full-featured electrosurgical generator with monopolar, bipolar, and LigaSure™ vessel sealing output modes. TThe generator is an electrically isolated, microcontroller-based device, incorporating closed-loop control for all output modes implemented in the microcontroller firmware. The generator incorporates Instant Response™ technology to constantly measure the electrical impedance of the tissue and instantaneously adjust the generator output to maintain the desired power.

    The generator is used with a selection of electrosurgical instruments designed for use with the ForceTriad and the LigaSure Vessel Sealing generator. All of the LigaSure instruments are capable of sealing vessels up to, and including, 7mm, and tissue bundles as large as can fit in the jaws of each instrument. When a LigaSure instrument is applied to a vessel or tissue bundle and RF energy is applied, the collagen and elastin in the tissues are reformed by heat and pressure to fuse vessel walls, thereby forming a permanent seal. The microprocessor in the generator monitors the tissue properties, stops the application of energy, and allows a brief period of cooling before indicating that the seal cycle is complete.

    No changes are being made to the design, operation, or intended use of any of the current system except the software modifications to create the Bipolar Resection modes and to allow the ForceTriad to output system data to a computer. The subject of this 510(k) notification is in regards to the addition of specific definitions for "cut" and "coagulating" and the addition to the indications statement of a bipolar resection function for use with resectoscopes "for endoscopically controlled removal of tissue using 0.9% NaCl solution (saline) as the irrigation medium" and the hardware and software associated with the activation of this feature.

    AI/ML Overview

    Here's an analysis of the provided text regarding the ForceTriad™ Electrosurgical Generator, focusing on acceptance criteria and supporting studies:

    It's important to note that the provided document is a 510(k) premarket notification, which concerns substantial equivalence to predicate devices, rather than a detailed report of a new device's efficacy studies in the same way one might find for novel drug or medical device. Therefore, the information regarding specific acceptance criteria and detailed study outcomes for meeting those criteria is presented at a high level.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Software ValidationSoftware validation was conducted to qualify the software changes related to the bipolar resection feature.
    Bench TestingBench testing using a pork model (to simulate prostate) was conducted.
    Live Porcine ModelLive porcine model testing using kidney (for hemostasis) was conducted.
    Safety StandardsTesting required by relevant safety standards was conducted.
    Overall Conclusion"Results were acceptable." "Testing found that the software met acceptance criteria." "The testing results at the system and individual accessories levels have met the pass-fail criteria." Conclusion: "Based on the information provided, Covidien concludes that the modified device is substantially equivalent to the predicate devices. The software modifications made to the ForceTriad™ detailed in this Traditional 510(k) have not introduced any new questions of Safety and Efficacy. The new software was developed and tested in accordance with the Design Control regulations and internal procedures."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes (i.e., number of pork models or live porcine models) used for the bench and live animal testing. It also doesn't specify the country of origin for the data or whether the studies were retrospective or prospective, though animal model testing is inherently prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The document refers to "bench testing" and "live porcine model" studies, but it doesn't detail how the "ground truth" or "acceptable results" were determined, nor does it mention the involvement or qualifications of experts in assessing the outcomes of these tests. For electrosurgical devices, ground truth might relate to tissue effects, seal integrity, or lack of collateral damage, which would traditionally be assessed by veterinary pathologists or surgeons.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given the nature of bench and animal testing for an electrosurgical device, it's less likely to involve the same type of "adjudication" seen in image-based diagnostic studies. Performance would likely be assessed against predefined engineering specifications (e.g., power output, tissue cut/coagulation characteristics).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. The document explicitly states: "This premarket notification report does not rely on the assessment of clinical performance data to demonstrate substantial equivalence." This means there was no human reader study comparing effectiveness with and without AI assistance (as AI is not the primary focus here, rather software updates for device functionality).

    6. Standalone (Algorithm Only) Performance Study

    • No standalone algorithm performance study in the typical sense was performed or reported. This submission focuses on software updates to an existing electrosurgical generator to enable a bipolar resection feature. The "algorithm" here refers to the closed-loop control and Instant Response™ technology within the electrosurgical generator, which is designed to adjust output based on tissue impedance. The testing confirmed the device's functionality with these software updates, not an independent algorithm's diagnostic or predictive performance.

    7. Type of Ground Truth Used

    Based on the description of "bench testing using a pork model (to simulate prostate) and live porcine model using kidney (for hemostasis)," the ground truth for these tests would likely involve direct observation and measurement of tissue effects. This could include:

    • Pathology/Histology: Microscopic examination of treated tissue to assess cut quality, depth of coagulation, cell viability, and collateral damage.
    • Physiological Measurements: For hemostasis, this might involve quantitative measures of blood loss or time to achieve cessation of bleeding.
    • Engineering Specifications: Measurement of power output, impedance feedback, and response times against predefined technical specifications.

    The document states "Results were acceptable," implying that these assessments met predetermined performance standards.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI algorithm development. The software updates relate to activating a specific electrosurgical mode and recognizing accessories, not to a learning algorithm requiring a distinct training phase. The "software validation" and "testing" refer to verification and validation of the software's functionality and safety, not training an AI model.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a "training set" for an AI algorithm, this question is not applicable to the provided document. The ground truth for the testing that was performed (bench and live animal) would have been established as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102913

    Validate with FDA (Live)

    Device Name
    FORCETRIAD
    Date Cleared
    2011-05-06

    (217 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indications for use include general (including urologic, thoracic, plastic and reconstructive, arthroscopic), laparoscopic, and gynecologic procedures where electrosurgical cutting and coagulation of tissue, and sealing (fusion) of vessels and tissue bundles is performed, including such procedures as bowel resections, hysterectomies (both vaginal and abdominal). laparoscopic cholesystectomies, laparoscopically assisted vaqinal hysterectomies, gall bladder procedures, Nissen fundoplication, adhesilysis, oophorectomy, etc. Vessels (arteries, veins, pulmonary arteries, pulmonary veins, lymph) 7mm and smaller in diameter, and bundles as large as will fit in the jaws of the devices can be sealed with the LigaSure™ vessel sealing (tissue fusion) output.

    Device Description

    The ForceTriad™ generator is a full-featured electrosurgical generator with monopolar, bipolar, and LigaSure™ vessel sealing outputs. The generator is an electrically isolated, microcontroller-based device, incorporating closed-loop control for all output modes implemented in the microcontroller firmware. The generator incorporates Instant Response™ technology to constantly measure the electrical impedance of the tissue and instantaneously adjust the generator output to maintain the desired power.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ForceTriad™ Electrosurgical Generator:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Safety and Efficacy: No new questions of Safety and Efficacy due to modifications.The software modifications that allow communication of the ForceTriad with the Olympus EndoAlpha endosurgery system add no new questions of Safety and Efficacy.
    Performance, Function, and Intended Use: No impact on the performance, function, or intended use of the ForceTriad.The software changes have no impact on the performance, function or intended use of the ForceTriad.
    Software Validation: Extensive software validation conducted.Extensive software validation was conducted to qualify the software changes.
    Bench Testing: Bench testing with a simulated and an actual Integrated OR System.Bench testing with a simulated Integrated OR System was conducted as well as testing with an actual system, the Olympus Medical Systems EndoAlpha system.
    Compliance with Safety Standards: New testing required by relevant safety standards conducted.Any new testing required by the new revisions of the relevant safety standards was also conducted.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states: "No clinical testing was conducted." Therefore, there is no test set in the traditional sense of patient data. The evaluation was based on non-clinical testing, including software validation and bench testing. The provenance of this non-clinical test data is not specified in terms of country of origin, but it is implied to be internal testing by the manufacturer (Covidien). The testing itself would be considered prospective for the specific software change being evaluated.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    As no clinical testing was performed and the evaluation focused on software and bench testing, there were no "experts" in the sense of clinical professionals establishing a ground truth on patient data. The ground truth for the non-clinical tests would have been established by engineering and quality assurance personnel, likely using predefined specifications and expected outputs for the software and hardware interfaces. Their specific qualifications are not detailed in this summary.

    4. Adjudication Method for the Test Set:

    Not applicable, as no clinical test set with human assessment was utilized. The "ground truth" for the non-clinical tests would have been determined through objective measurements and comparisons against technical specifications, not through a consensus-based adjudication process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states, "No clinical testing was conducted." Therefore, there is no information about the effect size of human readers with vs. without AI assistance. This submission focuses on a software change for integration, not on a new diagnostic or interpretative AI feature.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This device is an electrosurgical generator, which is inherently a human-in-the-loop device. However, the specific software change being evaluated (allowing the generator to interface with an Integrated OR system) was likely tested in a standalone manner relative to its function of communication. The bench testing and software validation would assess the algorithm's ability to correctly interface and transmit data without direct human intervention once configured, but the overall device always requires human operation. The document doesn't explicitly describe a "standalone performance" study in the context of an AI algorithm making decisions, as this is not an AI-driven diagnostic device.

    7. The Type of Ground Truth Used:

    The ground truth used for this submission was primarily technical specifications and expected system behavior. For software validation, this would involve comparing actual software outputs against designed functional requirements. For bench testing, it would involve verifying that the device communicated correctly with the simulated and actual Olympus EndoAlpha system according to predefined communication protocols and interface standards.

    8. The Sample Size for the Training Set:

    Not applicable. This submission is for a software modification to an existing device for interoperability. There is no mention of a machine learning or artificial intelligence component that would require a "training set" in the traditional sense. The software development process would involve iterative testing and debugging, but not a distinct training set for an AI model.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there was no training set in the context of machine learning. The "ground truth" for the software development and validation would have been based on the functional specifications and design documents for the integration with the Olympus EndoAlpha endosurgery system, established by Covidien's engineering team.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070162

    Validate with FDA (Live)

    Date Cleared
    2007-03-06

    (48 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ForceTriad™ is a full-featured electrosurgical generator intended for open and laparoscopic surgical procedures where the surgeon requires electrosurgical cutting, coagulation, or vessel sealing (tissue fusion). The generator is intended for use in general, laparoscopic and gynecologic surgical procedures where ligation of vessels, pulmonary vasculature, or lymph vessels, is desired. The system creates a vessel ligation (seal) by the application of bipolar electrosurgical RF energy (coagulation) to vessels interposed between the jaws of the device. The generator can also be used with standard bipolar devices where bipolar cutting or coagulation is desired.

    The indications for use include general (including urologic, thoracic, plastic and reconstructive), laparoscopic, and gynecological procedures where electrosurgical cutting and coagulation of tissue, and sealing (fusion) of vessels, including pulmonary vessels, and tissue bundles is performed. including such procedures as bowel resections, hysterectomies (both vaginal and abdominal), laparoscopic cholesystectomies, laparoscopically assisted vaginal hysterectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomy, etc. The devices can be used on vessels (arteries, veins, pulmonary arteries, pulmonary veins, lymph) up to 7mm and bundles as large as will fit in the jaws of the instruments.

    The LigaSure tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.

    Device Description

    The ForceTriad™ generator is a full-featured electrosurgical generator with monopolar, bipolar, and Valleylab LigaSure™ vessel sealing outputs. This 510(k) only applies to the LigaSure™ Vessel sealing portion of the generator. The generator is electrically isolated, microcontroller-based device, incorporating closed-loop control for all output modes implemented in the microcontroller firmware. The generator incorporates Instant Response™ technology to constantly measure the electrical impedance of the tissue and instantaneously adjust the generator output to maintain the desired power.

    The generator is used with a selection of instruments designed for use with the ForceTriad and the LigaSure Vessel Sealing generator. All of the instruments are capable of sealing vessels up to, and including, 7mm, and tissue bundles as large as can fit in the jaws of each instrument. When a LigaSure™ instrument is applied to a vessel or tissue bundle and RF energy is applied, the collagen and elastin in the tissues are reformed by heat and pressure to fuse vessel walls, thereby forming a permanent seal. The microprocessor in the generator monitors the tissue properties, stops the application of energy, and allows a brief period of cooling before indicating that cycle is complete.

    No changes are being made to the design or operation of any of the devices within the current system. The change as proposed in this 510(k) notification is to the intended use as described above and the resulting labeling changes.

    AI/ML Overview

    This document describes the ForceTriad™ Electrosurgical Generator, specifically its LigaSure™ Vessel Sealing System component. The submission is a 510(k) for a change in intended use and labeling, as the technological characteristics of the device have not been modified.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance

      The document does not explicitly state quantitative acceptance criteria in a table format. However, it implicitly defines successful performance for the LigaSure™ vessel sealing function:

      Acceptance Criteria (Implied)Reported Device Performance
      Effective sealing of pulmonary vasculaturePre-clinical studies (acute and chronic) and bench testing have shown that the ForceTriad Electrosurgical Generator effectively seals pulmonary vasculature.
      Seals with burst pressures substantially greater than physiologic pressures in the vesselsPre-clinical studies (acute and chronic) and bench testing have shown that the ForceTriad Electrosurgical Generator produces seals with burst pressures substantially greater than the physiologic pressures in the vessels.
      Capability to seal vessels up to and including 7mmAll of the instruments are capable of sealing vessels up to, and including, 7mm. The devices can be used on vessels (arteries, veins, pulmonary arteries, pulmonary veins, lymph) up to 7mm.
      Capability to seal tissue bundles as large as can fit in the jaws of each instrumentAll of the instruments are capable of sealing tissue bundles as large as can fit in the jaws of each instrument. The devices can be used on... bundles as large as will fit in the jaws of the instruments.
      Formation of a permanent seal to fuse vessel wallsWhen a LigaSure™ instrument is applied to a vessel or tissue bundle and RF energy is applied, the collagen and elastin in the tissues are reformed by heat and pressure to fuse vessel walls, thereby forming a permanent seal. The microprocessor in the generator monitors the tissue properties, stops the application of energy, and allows a brief period of cooling before indicating that cycle is complete.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      The document states: "Pre-clinical studies (acute and chronic) and bench testing have shown that the ForceTriad Electrosurgical Generator effectively seals pulmonary vasculature..."

      • Sample size: Not specified.
      • Data provenance: Not specified, but "pre-clinical studies" generally implies laboratory or animal studies, and "bench testing" is laboratory-based. No information on country of origin is provided. Both would be considered prospective studies in their design.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      Not applicable. The performance is assessed through objective biophysical measurements (e.g., burst pressure) and physical observation of sealing, not through expert human interpretation of images or clinical outcomes that require expert consensus for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      Not applicable. Adjudication methods like 2+1 are typically used for establishing ground truth from multiple human readers, often in image-based diagnostic studies. This study relies on objective physical measurements and observations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      Not applicable. This is a medical device performance study, not a diagnostic AI study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      Yes, the performance data presented is for the device operating in a standalone manner. The device itself (the ForceTriad™ Electrosurgical Generator with its LigaSure™ functionality) is assessed based on its ability to achieve effective vessel sealing and high burst pressures, without human intervention in the sealing process once initiated. The human's role is in operating the device, not in interpreting its output for the technical performance measurements.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      The ground truth is established through objective biophysical measurements (e.g., burst pressure testing) and direct observation of physical outcomes (e.g., effective sealing/fusion of vessels). This type of ground truth is based on engineering and physiological principles rather than expert interpretation of medical data.

    8. The sample size for the training set

      Not applicable. This device is not an AI/machine learning model in the sense that it requires a "training set" of data to develop its algorithm. The device's operation is based on established electrosurgical principles and a microcontroller-based closed-loop control system. Its parameters are likely determined through engineering design and validation, not a data-driven training process.

    9. How the ground truth for the training set was established

      Not applicable, as there is no "training set" in the context of this electrosurgical generator's design and validation described here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1