    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FLOWRUNNER Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.

    Device Description

    The FLOWRUNNER Aspiration System is a peripheral thrombectomy system consisting of several components:

    • FLOWRUNNER Aspiration Catheter
    • FLOWRUNNER Aspiration Sheath
    • Vacuum Fitting
    • High Flow Stopcock Connector
    • Y-Connector with Hemostatic Valve
    • Hand Actuator Clip

    The FLOWRUNNER Aspiration System family is designed for the minimally invasive removal of thrombus from the peripheral vasculature using aspiration. The extension of the device family includes smaller-diameter sizes of the FLOWRUNNER Aspiration System. The system is a single-use catheter-based device with the capability to infuse fluids.

    FLOWRUNNER SP Aspiration System is over the wire system.

    FLOWRUNNER Aspiration System consists of one aspiration catheter, one aspiration sheath, one vacuum fitting, one high flow stopcock connector, one Y-connector with hemostatic valve, and a hand actuator clip.

    The FLOWRUNNER Aspiration System is introduced to the site of the primary occlusion. The Aspiration Catheter is advanced through the Aspiration Sheath and targets aspiration directly to the thrombus. The Aspiration Catheter is then retracted back into the Aspiration Sheath. This process of extension and retraction of the Aspiration Catheter is then repeated to fully aspirate the clot. Suction is applied directly to the Aspiration Catheter from an external vacuum source to aspirate thrombus from an occluded vessel (maximum pressure -27.6 in Hg and minimum pressure of -8.0 in Hg).

    Sterile saline flows through the Aspiration Sheath and into the Aspiration Catheter when connected proximally. The Aspiration Catheter and Sheath are visible under fluoroscopy. The hand actuator is an optional, proximal clip attached to the Aspiration Catheter to assist the user.

    The FLOWRUNNER Aspiration System comes in 7F, SP 7F, 12F, and 14F compatible diameters and 60cm, 80cm, 100cm, 120 and 130cm lengths.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251488

    Validate with FDA (Live)

    Date Cleared
    2025-07-14

    (61 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Expanse FLOWRUNNER Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.

    Device Description

    The Expanse FLOWRUNNER Aspiration System is a peripheral thrombectomy system consisting of several components:

    • FLOWRUNNER Aspiration Catheter
    • FLOWRUNNER Aspiration Sheath
    • Vacuum Fitting
    • High Flow Stopcock Connector
    • Y-Connector with Hemostatic Valve
    • Hand Actuator Clip

    The FLOWRUNNER Aspiration System is designed for the minimally invasive removal of thrombus from the peripheral vasculature using aspiration. The system has demonstrated compatibility with standard vascular implants, including stents and filters. It is a single use, over-the-wire, catheter-based system with the ability to infuse fluid. The FLOWRUNNER Aspiration System consists of one aspiration catheter, one aspiration sheath, one vacuum fitting, one high flow stopcock connector, one Y-connector with hemostatic valve, and a hand actuator clip.

    The FLOWRUNNER Aspiration System is introduced to the site of the primary occlusion. The Aspiration Catheter is advanced through the Aspiration Sheath and targets aspiration directly to the thrombus. The Aspiration Catheter is then retracted back into the Aspiration Sheath. This process of extension and retraction of the Aspiration Catheter is then repeated to fully aspirate the clot. Suction is applied directly to the Aspiration Catheter from an external vacuum source to aspirate thrombus from an occluded vessel (maximum pressure -27.6 in Hg and minimum pressure of -8.0 in Hg).

    Sterile saline flows through the Aspiration Sheath and into the Aspiration Catheter when connected proximally. No saline is injected into the patient during aspiration. The Aspiration Catheter and Sheath are visible under fluoroscopy. The hand actuator is an optional, proximal clip attached to the Aspiration Catheter to assist the user.

    The FLOWRUNNER Aspiration System comes in 12F, 14F compatible diameters and 60cm, 80cm, 100cm and 120cm lengths.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the FLOWRUNNER Aspiration System are for a medical device (an embolectomy catheter), not an AI/ML powered device. As such, the document does not contain information typically found in submissions for AI/ML devices regarding acceptance criteria, study design for AI performance, ground truth, or expert involvement in classification.

    The information provided focuses on the FLOWRUNNER Aspiration System's equivalence to a predicate device (ICE Aspiration System) for its intended use in removing emboli and thrombi from peripheral vessels. The "performance data" section refers to bench testing, which is standard for hardware medical devices, not AI/ML performance metrics.

    Therefore, I cannot extract the requested information about acceptance criteria, study details, human reader improvement with AI, standalone performance, ground truth establishment, or training set size. This document does not describe a study involving AI/ML.

    Here's a summary of what information can be extracted from the document, related to its non-AI performance claims:

    1. A table of acceptance criteria and the reported device performance:

    The document summarizes the performance testing generically as "Bench testing was performed to support a determination of substantial equivalence to the predicate." It lists the types of bench tests conducted but does not provide specific acceptance criteria values or detailed reported performance results against those criteria.

    Acceptance Criteria (Implied)Reported Device Performance
    In-Stent and IVC Filter compatibility"demonstrated compatibility with standard vascular implants, including stents and filters"
    Thrombus CapturePerformed (specific results not detailed)
    Conformance to intended use requirements"designed and tested to assure conformance to the requirements for its intended use"
    Reliability (general)"Performance testing demonstrated that the FLOWRUNNER Aspiration System reliably raises no new questions of safety and effectiveness."

    2. Sample size used for the test set and the data provenance:

    Not applicable, as this refers to bench testing on physical components, not a clinical test set with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications:

    Not applicable. Ground truth for clinical data is irrelevant to bench testing of a hardware device.

    4. Adjudication method for the test set:

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This device is not AI-powered, so no such study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    No. This device is not an algorithm. Bench testing evaluates the physical device's performance.

    7. The type of ground truth used:

    Not applicable in the context of AI/ML. For the bench testing, the "ground truth" would be engineering specifications and validated test methods to assess the physical device's capabilities (e.g., ability to capture thrombus in a model, compatibility with stents).

    8. The sample size for the training set:

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established:

    Not applicable. This is not an AI/ML device.