Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121561

    Validate with FDA (Live)

    Device Name
    FLOW2, SWING
    Date Cleared
    2012-10-26

    (150 days)

    Product Code
    Regulation Number
    890.5150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ThyssenKrupp Accessibility's Battery Operated Stair Lift FLOW2 is intended to mechanically transport one mobility impaired person in a fold-down seat, up and down curved stairs, indoors.

    Device Description

    Flow2 mono rail chairlift brings one person safely and comfortably up and down the stairs. In the meantime one saves ones energy for long outdoor walks, playing with the children or the daily tasks in house.

    Flow2 is capable to swivel during the ride up and down the stairs. This unique and patented feature makes that Flow2 chairlift fits most staircases, straight, around corners and even spiral. The angle of the staircase can vary from -70 degrees up to 70 degrees.

    The footrest is mounted directly to the chair. This provides a comfortable and ergonomic seating position of the user with less knee flexion.

    Flow2 is easy to use by a simple and light folding operation and the intuitive to use joystick control. Features such as anti squeeze, curved armrest and safety belt provide additional safety for the user and other inhabitants of the home.

    The design can be adapted to the home decor by the application of a variety of different seat fabrics and colors. Flow2 is the chairlift on the market which combines the best of 2 worlds; capable to fit on the smallest stairs while transporting the user in the most comfortable and safe way.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ThyssenKrupp Accessibility FLOW2 stair lift:

    This 510(k) summary does not contain a traditional performance study comparing the device to acceptance criteria with detailed metrics like sensitivity, specificity, accuracy, or confidence intervals. Instead, it relies on a substantial equivalence approach by comparing the FLOW2 to a predicate device (Bruno CRE-2110 Electra-Ride) and demonstrating compliance with relevant international standards.

    For medical devices, especially those that are electromechanical and not diagnostic in nature, "acceptance criteria" often refer to meeting established safety and performance standards, and demonstrating that the new device is as safe and effective as a legally marketed predicate.

    Here's a breakdown of the requested information based on the provided document:


    1. A table of acceptance criteria and the reported device performance

    For devices like a stair lift, acceptance criteria often relate to meeting specific safety standards, load capacities, speed limits, and other functional aspects. The "reported device performance" is essentially that the device meets these standards and is comparable to the predicate.

    Acceptance Criteria / StandardReported Device Performance (FLOW2)Predicate Device (Bruno CRE-2110 Electra-Ride)
    General Safety and Performance StandardsCompliant with ASME 18.1, IEC 60601-1-2:2007 (Certified by ETL and Lift Institute)Compliant with ASME 18.1, IEC 60601-1-2:2007
    Intended UseMechanically transport one mobility impaired person in a fold-down seat, up and down curved stairs, indoors.Mechanically transport one mobility impaired person in a fold-down seat, up and down stairs, indoors. (Implied for curved stairs as well given "Straight and Curved stairs" application)
    Rated Load (Maximum)275 lb (125 kg)400 lb (181 kg)
    Passengers11
    Power Supply2 pieces 7AH 12 Battery2 pieces 7AH 12 Battery
    Charger24VDC/ 2A24VDC/ 2A
    Drive MechanismPlanetary motor with worm gearDirect drive worm gear motor
    Final DriveLasered rack and pinion8dp gear rack with spur gear
    Braking SystemDynamic, worm gear, el/mech brakeDynamic, worm gear, el/mech brake
    Call & Send FunctionalityRFRF
    Support/MountingAnchored to stair threadAnchored to stair thread
    Maximum Stair Angle70 degrees50 degrees
    Speed0.15 m/s0.13 m/s
    Track Length (Maximum)34 m15 m
    Safety DevicesMultipleMultiple
    Footrest TypeFolding typeFolding type
    Seat TypeFolding typeFolding type
    Control Mechanism3 position switched RF unit3 position momentary rocker switch
    Construction MaterialBended round steel tube 80mmSteel welded shape
    Leveling MechanismPatented electronic leveling (multiple sensor controlled)Mechanically forced
    Swiveling MechanismPatented "en route" swivelingEnd of rail

    Note on Acceptance Criteria: For this type of device and submission, the primary "acceptance criteria" are the successful demonstration of substantial equivalence to a predicate device and compliance with recognized consensus standards (ASME 18.1, IEC 60601-1-2:2007). The "reported device performance" is essentially that the device meets these standards and its characteristics are within an acceptable range compared to the predicate, even where there are non-equivalent technological characteristics. The non-equivalent characteristics were discussed to show they do not raise new questions of safety or effectiveness.


    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not applicable in the context of a clinical trial or performance study with a test set of data. The submission relies on engineering testing and comparison to standards and a predicate device.
    • Data Provenance: Not applicable for clinical data. The data provenance is from engineering and safety standard testing conducted by organizations like ETL and Lift Institute. The country of origin of the manufacturer is The Netherlands.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth establishment. Third-party testing and certification bodies (ETL, Lift Institute) verified compliance with safety and performance standards. Their "expertise" for this type of device lies in their accreditation to test against these specific standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. There was no "test set" of clinical cases requiring adjudication. Compliance was determined by accredited testing laboratories against engineering standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. This device is a mechanical stair lift, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a mechanical device with human users, not an algorithm. Its operation involves human interaction (joystick control) for its intended function. The "standalone" performance would be its mechanical and electrical function independent of a specific user, verified by engineering tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this submission is established by compliance with international consensus safety and performance standards (ASME 18.1, IEC 60601-1-2:2007) as verified by accredited testing bodies (ETL, Lift Institute). For mechanical devices, this often involves physical load testing, endurance testing, electrical safety testing, and functional assessments without human subjects.

    8. The sample size for the training set

    • Not applicable. There is no AI component or "training set" for this device.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1