510(k) Data Aggregation

    K Number
    K172205

    Validate with FDA (Live)

    Device Name
    ExoToe Staple
    Manufacturer
    Date Cleared
    2018-01-11

    (174 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ExoToe Staple is intended for fracture and osteotomy fixation and joint arthrodesis of the foot in conjunction with k-wire fixation.

    Device Description

    The ExoToe Staple is an implant that is crimped onto bone with forceps after a k-wire has been placed to achieve fixation. The ExoToe Staple is provided in a sterile package. A kit containing four sizing templates and forceps is provided non-sterile. The staple is comprised of stainless steel and is pre-formed during manufacturing. The sizing templates and pliers are comprised of stainless steel. The sizing templates are used to select staple size. The forceps are used by a surgeon to conduct the final deformation required to anchor the staple to the bone.

    AI/ML Overview

    This document describes the premarket notification for the ExoToe Staple, a bone fixation device. The information provided does not pertain to an AI/ML medical device and therefore there is no discussion of acceptance criteria for an algorithm, test sets, training sets, ground truth establishment, or human reader studies.

    The document discusses the substantial equivalence of the ExoToe Staple to previously marketed predicate devices (Memodyn Staple, Z-Staple, and Compressyn Staple) based on non-clinical performance data.

    Here's the relevant information from the provided document regarding the device's performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for non-clinical performance were based on demonstrating "substantial equivalence" to predicate devices. The reported device performance is presented as conclusions of substantial equivalence for various biomechanical tests.

    EvaluationAcceptance Criteria (Implied)Reported Device Performance Conclusion
    Dorsal pull-offSubstantially EquivalentSubstantially Equivalent
    Static 4-point bendSubstantially EquivalentSubstantially Equivalent
    Dynamic 4-point bendSubstantially EquivalentSubstantially Equivalent
    ProtrusionSubstantially EquivalentSubstantially Equivalent

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes (number of staples tested) for each biomechanical test. It also does not explicitly state the provenance of the data, but it is implied to be from laboratory testing conducted as part of the device development and submission process. This would be prospective testing, as it's performed specifically to support the regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The "ground truth" for a physical medical device like a bone staple is established through objective biomechanical testing according to established standards or validated test methods, not through expert consensus on interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) to resolve discrepancies. For biomechanical testing, the results are quantitative measurements, and the "truth" is the measured value against a pre-defined acceptance threshold or comparison to predicate device performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI/ML devices where human readers' diagnostic performance is evaluated with and without AI assistance. The ExoToe Staple is a physical implant, not an AI/ML device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The ExoToe Staple is not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance of the ExoToe Staple was established through non-clinical performance testing (biomechanical testing). This involves objective measurements of physical properties (e.g., pull-off strength, bend strength, protrusion) against defined criteria or in comparison to predicate devices, rather than expert consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set

    This information is not applicable because the ExoToe Staple is not an AI/ML device and thus does not have a "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable, as there is no training set for a physical device.