    K Number
    K233299

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-01-26

    (119 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ExactechGPS® System is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures.

    TKA Pro software application is specifically indicated for Total Knee Arthroplasty.

    Device Description

    The ExactechGPS System proposed in this submission is a modification of the ExactechGPS System cleared per 510(k) #K213877.

    The ExactechGPS System is an image-guided surgery, or navigation, system intended to be used during orthopedic surgical procedures to intraoperatively assist surgeons during arthroplasty. The ExactechGPS System enables surgeons to acquire intraoperative data by computing and displaying information such as distances, angles, and placement of prosthetic components in order to identify and characterize bone cuts necessary to achieve surgical goals.

    The ExactechGPS System works with ExactechGPS hardware trackers that communicate intraoperative data to the ExactechGPS hardware station to provide surgeons with real-time information on the positions of patient anatomical structures and instrumentation used to prepare bone during stereotaxic surgery.

    The proposed modifications do not change the ExactechGPS System general intended use, general design features, or basic fundamental scientific technology.

    AI/ML Overview

    This FDA 510(k) clearance letter and summary primarily address mechanical and electrical modifications to an existing device, the ExactechGPS® System. It focuses on demonstrating that these hardware changes do not significantly alter the performance of the device for its intended use in stereotaxic surgery for total knee arthroplasty (TKA).

    Therefore, the document does not contain the detailed information typically found in acceptance criteria and study designs for Artificial Intelligence/Machine Learning (AI/ML) based medical devices. The performance data section lists general function tests, electrical safety, EMC compatibility, photobiological safety, and sterilization validation, which are standard for hardware modifications, not AI/ML algorithm validation.

    Based on the provided text, I cannot extract the information required for a study proving an AI/ML device meets acceptance criteria. The document pertains to a hardware modification of a surgical navigation system, not an AI/ML component or software that analyzes data for diagnostic or prognostic purposes, which would necessitate the criteria you've outlined.

    Specifically, the document states:

    • "Based on test bench testing, the proposed changes do not alter significantly the performances of the devices and the way it is used."
    • "The proposed ExactechGPS system have the same general features and dimensions, with the same technology and performances."
    • "The redesigned electronic card only resolves obsolescence issues with no impact on user, global performances and safety."

    This indicates that the focus of the submission was to show non-inferiority or equivalence of the modified hardware to its predicate, rather than to prove a new performance claim for an AI/ML component.

    Therefore, I cannot fulfill your request to provide:

    • A table of acceptance criteria and reported device performance (in the context of AI/ML).
    • Sample sizes for a test set or training set (for AI/ML).
    • Details on experts, adjudication methods, MRMC studies, or standalone performance (for AI/ML).
    • Ground truth establishment methods (for AI/ML).

    The provided text deals with the ExactechGPS® System, which is a surgical navigation system (stereotaxic instrument). The submission is for hardware modifications to an existing cleared device, primarily to address obsolescence of sub-components (optical localizer, tablet, wireless feature). The core claim is that these hardware changes do not alter the fundamental performance or intended use of the device.

    The "TKA Pro software application" is mentioned as being compatible with the modified system, but the document does not describe any AI/ML components within this software application or any studies validating AI/ML performance. The performance data listed (electrical safety, EMC, etc.) are typical for hardware evaluations, not AI/ML model validation.

    In summary, this document is a 510(k) clearance for hardware modifications to a surgical navigation system, not an AI/ML device, and thus does not contain the information requested about AI/ML acceptance criteria or an AI/ML study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213877

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-02

    (140 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ExactechGPS® System is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures.

    TKA Pro software application is specifically indicated for Total Knee Arthroplasty.

    Device Description

    The ExactechGPS® System proposed in this submission is a modification of the ExactechGPS® System cleared per 510(k) #K193618.

    The ExactechGPS® System is an image-quided surgery, or navigation, system intended to be used during orthopedic surgical procedures to intraoperatively assist surgeons during arthroplasty. The ExactechGPS® System enables surgeons to acquire intraoperative data by computing and displaying information such as distances, angles, and placement of prosthetic components in order to identify and characterize bone cuts necessary to achieve surgical goals.

    The ExactechGPS® System works with ExactechGPS® hardware trackers that communicate intraoperative data to the ExactechGPS® hardware station to provide surgeons with real-time information on the positions of patient anatomical structures and instrumentation used to prepare bone during stereotaxic surgery.

    The proposed modifications do not change the ExactechGPS® System general intended use, general design features, or basic fundamental scientific technology.

    AI/ML Overview

    This document describes the ExactechGPS® System, a surgical navigation system, and its modifications. While the document mentions general performance tests, electrical safety, EMC compatibility, photobiological safety, biocompatibility, and cleaning/sterilization validation, it does not provide the specific details required to fully address your request regarding acceptance criteria and a study proving the device meets these criteria in the context of an AI/human-in-the-loop system.

    The device described is a "stereotaxic instrument" and mentions a "TKA Pro software application" for Total Knee Arthroplasty, aiding surgeons in locating anatomical structures and aligning endoprostheses. This implies a core functionality related to precise surgical guidance.

    Given the information provided in the document, here's what can be extracted and what is missing:

    Information Provided:

    • Device Name: ExactechGPS® System
    • Intended Use: "intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. TKA Pro software application is specifically indicated for Total Knee Arthroplasty."
    • Modifications: New set of ExactechGPS Trackers V2 (different architecture, updated electronic card, InfraRed bandpass filter) and compatibility with only the TKA Pro clinical software application (whereas the predicate was compatible with TKA Pro and TKA Plus).
    • Claimed Impact of Modifications: "Based on test bench testing, the proposed changes do not alter significantly the performances of the devices and the way it is used: The proposed ExactechGPS Trackers V2 have the same general features and dimensions, with the same technology and performances. The redesigned electronic card only resolves obsolescence issues with no impact on user, performance and safety. The new InfraRed bandpass filter only improves communication reliability."
    • Non-Clinical Testing: General Functions tests, Electrical Safety (IEC 60601-1), EMC Compatibility (IEC 60601-1-2), Photobiological Safety (EN 62471), Biocompatibility (ISO 10993-1:2018), Cleaning and Sterilization Validation (ISO 17664:2017, ISO 17665-1:2006, AAMI TIR12:2020, AAMI TIR30:2011/(R) 2016).

    Information NOT Provided (or not in sufficient detail):

    • Specific Acceptance Criteria and Performance Data for Surgical Accuracy/Clinical Outcome: While "General Functions tests" are mentioned, no quantitative acceptance criteria for surgical accuracy (e.g., angular or translational precision of component placement) or specific performance results (e.g., mean angular deviation, success rate) are presented in a table format as requested. The document states "the proposed changes do not alter significantly the performances," implying a comparison to the predicate, but no specific performance metrics are given.
    • Sample Size for Test Set and Data Provenance: Not specified for any study related to the device's functional performance in surgical guidance. The document mentions "test bench testing" but no details on the size or nature of the test set.
    • Number of Experts and Qualifications for Ground Truth: Not applicable as no reader study or expert-based ground truth establishment is described for the device's performance. The device is a navigation system, so its "ground truth" would presumably be the physical reality of component alignment or anatomical structure location, measured by highly accurate metrology, rather than expert interpretation of images.
    • Adjudication Method for Test Set: Not applicable.
    • MRMC Comparative Effectiveness Study: Not mentioned or implied. This device is a navigation system, not an AI diagnostic tool primarily assisting human readers in interpreting images.
    • Standalone Performance (Algorithm Only): The device functions as a system with hardware and software; standalone algorithm-only performance is not directly relevant in the way it might be for an AI diagnostic algorithm.
    • Type of Ground Truth Used: For a navigation system, ground truth would typically come from highly accurate metrological measurements of physical objects or cadaveric studies, not expert consensus on images or pathology. This exact type of ground truth is not explicitly detailed.
    • Sample Size for Training Set: Not specified. It's unclear if "training" in the machine learning sense applies to this system's development, or if it refers to calibration/validation data for the navigation system.
    • How Ground Truth for Training Set was Established: Not specified.

    Based on the provided document, here's a structured response indicating available information and significant gaps:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriterionReported Device Performance
    Functional Performance (e.g., Surgical Accuracy, Precision)NOT SPECIFIED QUANTITATIVELY. The document states: - "General Functions tests to ensure design is fully verified and validated." - "The proposed changes do not alter significantly the performances of the devices and the way it is used." There are no specific metrics (e.g., angular deviation, translational error) or numerical results provided to define acceptance criteria or demonstrate performance.
    Electrical Safety (IEC 60601-1)Compliant with requirements of the standard.
    EMC Compatibility (IEC 60601-1-2)Compliant with requirements of the standard.
    Photobiological Safety (EN 62471)Compliant with requirements of the standard.
    Biocompatibility (ISO 10993-1:2018)Compliant with requirements of the standard.
    Cleaning and Sterilization Validation (ISO 17664:2017, ISO 17665-1:2006, AAMI TIR12:2020, AAMI TIR30:2011/(R) 2016)Compliant with requirements of the standards.
    Compatibility with Tracker V2 Performance"The proposed ExactechGPS Trackers V2 have the same general features and dimensions, with the same technology and performances."
    Electronic Card Redesign Impact"redesigned electronic card only resolves obsolescence issues with no impact on user, performance and safety."
    Infrared Bandpass Filter Impact"new InfraRed bandpass filter only improves communication reliability."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified. The document only mentions "test bench testing."
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The testing was conducted "in-house and by accredited testing laboratories."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Not applicable as no expert-based ground truth establishment or reader study is described for the performance claims. The device is a navigation system; its performance would likely be evaluated against physical measurements.

    4. Adjudication Method for the Test Set:

    • Not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No evidence of an MRMC study. The device is a surgical navigation system, not a diagnostic AI tool for image interpretation by human readers. Its primary function is to provide real-time guidance during surgery, not to enhance human interpretation of medical images.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The document describes the device as a system of hardware (trackers, station) and software. The listed "performance data" refers to general system functions and regulatory compliance (electrical safety, EMC, etc.). It does not detail a standalone algorithm-only performance study related to its core navigation accuracy. The "TKA Pro software application" is part of the system, not presented as a standalone AI for diagnosis.

    7. The Type of Ground Truth Used:

    • For a surgical navigation system, the ground truth for functional performance (like accuracy of component placement) would typically be established using highly accurate metrological measurements on physical models, cadavers, or potentially intraoperative measurements. This document does not specify the exact type of ground truth used for "General Functions tests" or "test bench testing" related to the navigation's accuracy. It is not expert consensus, pathology, or outcomes data in the traditional sense for diagnostic devices.

    8. The Sample Size for the Training Set:

    • Not specified. It's unclear if "training set" refers to a machine learning training set or calibration data for the navigation system. The document does not describe an AI model that would require a distinct "training set" in the context of deep learning.

    9. How the Ground Truth for the Training Set was Established:

    • Not specified and likely not applicable in the context of an AI training set as described in the prompt, as the device is not presented as a deep learning AI diagnostic tool.