510(k) Data Aggregation

    K Number
    K143659

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-09

    (107 days)

    Product Code
    Regulation Number
    888.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Equinoxe Mega Prosthesis System is intended for use in hemi or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of:

    • Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
    • · Rheumatoid arthritis
    • · Revision where other devices or treatments have failed
    • · Correction of functional deformity
    • Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment
    • · Traumatic arthritis
    • · Oncology applications including bone loss due to tumor resection.
    • · Significant humeral resection which are unmanageable using other treatment methods

    The Equinoxe Mega Prosthesis System can be used in either primary or revision arthroplasty procedures.

    The Equinoxe Mega Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreparable and grossly deficient, and a functional deltoid muscle is present.

    The Equinoxe Mega Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications.

    The Equinoxe Mega Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Mega Prosthesis Distal Fixation Rings are only for uncemented, press-fit use.

    Device Description

    The Equinoxe Mega Prosthesis differs from conventional Total Shoulder Arthroplasty implants in that it is intended to replace both the articulating surface (humeral head) as well as the resected humerus. The modular design allows for the surgeon to construct the middle and proximal segments of the prosthesis in various lengths in order to match the length of resected bone, and to maintain the patient's natural arm length.

    The prosthesis is manufactured from Ti-6A1-4V and has both Titanium Plasma Spray and HA coated regions. The proximal bodies are available in four sizes (Small, Medium, Large, Extra-Large) and two lengths (+0mm, +12.5mm). The middle segments are available in 3 lengths (25mm, 50mm, 75mm) with one diameter (20mm). The distal stems are available in 3 lengths (80mm, 120mm, 200mm) and 6 diameters (6mm, 7mm, 8mm, 9mm, 11mm, 13mm). The distal fixation rings are available in 17 diameters (17.5mm-33.5mm, 1mm increments). The distal fixation rings are intended for press-fit fixation. The proximal bodies and middle segments have suture through holes to facilitate soft tissue attachment.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Exactech Equinoxe Mega Prosthesis, which is a medical device. The primary purpose of this type of submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, rather than proving its effectiveness or safety through extensive clinical trials as would be required for a PMA. Therefore, the information typically requested for acceptance criteria and studies (especially those involving human subjects, ground truth, experts, and statistical analysis with sample sizes) for AI/ML devices is not explicitly detailed in this type of submission for a traditional medical implant.

    However, I can extract the relevant information regarding the non-clinical testing performed to support substantial equivalence.

    Here's an attempt to fill in the requested table and information based on the provided document, acknowledging the limitations for a traditional medical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance Criteria (Implied/General)Reported Device Performance
    Indications for UseSimilar to predicate device; new indications do not create a new intended use.The proposed Exactech Equinoxe Mega Prosthesis has similar indications for use to the predicate device. Added hemi- and reverse shoulder arthroplasty indications, but these "do not create a new intended use" as stability is achieved comparably (dynamic muscle balancing for hemi/total; inverted articular concavities for reverse without rotator cuff dependence).
    Materials/Surface Finish/CoatingsComposed of similar, biocompatible materials and similar surface finish/coatings to predicate.Composed of Ti-6A1-4V with Titanium Plasma Spray and HA coated regions, similar to the predicate device, and are biocompatible.
    Design FeaturesSimilar design features to predicate device; design differences addressed by bench testing.Shares similar design features. Design differences (distal fixation ring, different sizes of distal stems and proximal bodies, hemi/reverse shoulder indications) were considered and addressed via bench testing.
    DimensionsDimensionally comparable to predicate device.Dimensionally comparable to the predicate device.
    SterilizationProvided sterile for single use, consistent with predicate.Provided sterile for single use only.
    Performance RequirementsConform to recognized performance standards for shoulder replacement devices."The proposed Exactech Equinoxe Mega Prosthesis and the predicate device conform to recognized performance standards for shoulder replacement devices." This is a general statement rather than specific quantitative criteria, common for a 510(k) where equivalence to existing standards is key.
    Plasma Spray Coating CharacterizationCoating properties meet specifications for permanent implants.Performed. Implied performance met industry standards for coating integrity and adhesion. (No specific quantitative data provided in the summary).
    Fatigue TestingDevice withstands cyclic loading without failure for expected lifespan.Performed. Implied performance met industry standards for fatigue strength and durability under simulated physiological loads. (No specific quantitative data provided in the summary).
    Torsion TestingDevice withstands torsional forces without failure.Performed. Implied performance met industry standards for torsional strength. (No specific quantitative data provided in the summary).
    Taper AnalysisTaper connections maintain integrity.Performed. Implied performance met design specifications for taper security. (No specific quantitative data provided in the summary).
    Axial Pull-Off TestingComponents remain securely connected under axial pull-off forces.Performed. Implied performance met design specifications for resistance to axial decoupling. (No specific quantitative data provided in the summary).
    Suture Abrasion TestingSuture through-holes do not significantly abrade sutures.Performed. Implied performance ensured that the suture through-holes did not cause excessive abrasion to sutures used for soft tissue attachment. (No specific quantitative data provided in the summary).
    Cadaveric StudyConfirms proper surgical technique, fit, and biomechanical interaction in a simulated environment.Performed. The study was part of validating the device's performance in a simulated surgical setting. (Details on specific outcomes or quantitative results are not provided in the summary). This would typically assess surgical handling, fit, range of motion, and stability characteristics.

    Detailed Study Information (Based on provided text, acknowledging this is a mechanical device, not AI/ML):

    The document describes a submission for a traditional medical device (Exactech Equinoxe Mega Prosthesis), not an AI/ML device. Therefore, many of the requested categories like "test set," "training set," "ground truth," "experts," and "MRMC comparative effectiveness study" are not applicable in the context of an AI/ML diagnostic system. The evaluation here is based on non-clinical (bench and cadaveric) testing and comparison to a predicate device for substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size (Test Set): Not explicitly stated in the summary for each test. For mechanical tests like fatigue, torsion, pull-off, etc., sample sizes would typically be determined by relevant ASTM or ISO standards but are not reported in this summary. For the cadaveric study, the sample size (number of cadavers or specimens) is not specified.
    • Data Provenance: The studies are non-clinical (bench and cadaveric). Country of origin is not specified, but the applicant (Exactech, Inc.) is based in Gainesville, Florida, USA. The studies are prospective in the sense that they were conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts/Qualifications: Not applicable in the context of this traditional device submission. For a cadaveric study, surgeons might be involved in evaluating surgical technique or fit, but they are not "establishing ground truth" in the diagnostic sense. The "ground truth" for mechanical testing is derived from engineering principles and standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable. This concept belongs to expert review of diagnostic cases, which is not the nature of these non-clinical tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This is a traditional medical device, not an AI/ML diagnostic.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Study: Not applicable. This is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth:
      • For mechanical tests (Fatigue, Torsion, Taper, Axial Pull-Off, Suture Abrasion, Plasma Spray Characterization): The "ground truth" is adherence to established engineering standards (e.g., ASTM, ISO) and successful demonstration of mechanical integrity and performance under simulated physiological conditions.
      • For the Cadaveric Study: The "ground truth" would be the observed biomechanical performance, surgical fit, and handling characteristics within the cadaveric specimen, assessed against expected anatomical and surgical outcomes.
      • For Clinical Literature Review: The "ground truth" is established medical knowledge and previous clinical outcomes reported in peer-reviewed literature for similar devices or surgical approaches.

    8. The sample size for the training set

    • Sample Size (Training Set): Not applicable. There is no AI/ML model being trained. The design of the device is based on engineering principles and existing knowledge.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment (Training Set): Not applicable.