510(k) Data Aggregation

    K Number
    K250515

    Validate with FDA (Live)

    Device Name
    EpiMonitor
    Manufacturer
    Date Cleared
    2025-06-19

    (118 days)

    Product Code
    Regulation Number
    882.1580
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EpiMonitor is a prescription only medical device system composed of a wearable device "EmbracePlus" and paired mobile software application "EpiMonitor" intended as an adjunct to seizure monitoring in adults and children aged 6 and up in a home environment or healthcare facilities. The device is worn on the wrist and senses Electrodermal Activity (EDA) and motion data to detect patterns that may be associated with either primary or secondary generalized tonic clonic seizures in patients with epilepsy or at risk of having epilepsy. When a seizure event is detected, the wearable device component of EpiMonitor sends a command to a paired mobile device where the EpiMonitor App is programmed to initiate an alert to a designated caregiver. The EpiMonitor app incorporates additional detection sensitivity modes, "high" for use during periods of rest or sleeping or "low" for use during periods of low-intensity activity, in order to reduce false alarm incidents.

    EpiMonitor records, stores and transmits accelerometer, EDA, peripheral skin temperature and activity data for subsequent retrospective review by a trained healthcare professional via a cloud-based software.

    Device Description

    The EpiMonitor system consists of a wearable device and mobile application:

    • A wearable medical device called EmbracePlus,
    • A mobile application running on smartphones called "EpiMonitor"

    The EmbracePlus is worn on the user's wrist and continuously collects raw data via specific sensors, these data are continuously analyzed by an on-board algorithm (EpiAlgo 2.1), which assesses the physiological data and determines if the user may be undergoing a generalized tonic-clonic seizure (GTCS). The EpiAlgo has been validated through testing, using the gold-standard video-Electroencephalogram (EEG) methodology designed by a group of epileptologists at a top level 4 epilepsy center, from epilepsy patients experiencing GTCSs in hospital Epilepsy Monitoring Units.

    When a likely GTCS is detected, EmbracePlus sends, via Bluetooth Low Energy, a message to the EpiMonitor app. The EpiMonitor app communicates to the Empatica Cloud which initiates, through the external provider a voice call and SMS text message is sent to summon the attention of user-designated caregiver(s).

    In addition to initiating alerts, the EpiMonitor app also continuously receives all the raw sensor data collected by the EmbracePlus. These data are analyzed by one of the EpiMonitor app software modules, EmpaDSP (paragraph 2.3.2), which computes the additional physiological parameters, such as activity during sleep and peripheral skin temperature.

    The EpiMonitor App is also responsible for transmitting, over a cellular data plan or Wi-Fi connection the sensors' raw data, device information, and computed physiological parameters to the Empatica Cloud. On the Empatica Cloud, these data are stored, and made available to healthcare providers via a specific cloud-based software called Care Monitoring Portal.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for EpiMonitor, based on the provided FDA clearance letter:


    Acceptance Criteria and Device Performance for EpiMonitor

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for PPA and FAR in a table format. Instead, it presents the device's performance for these metrics, implying that these results were deemed acceptable by the FDA for clearance. For the purpose of this response, I'm interpreting the "reported device performance" as the achieved PPA and FAR values and will frame the "acceptance criteria" as the expectation for these metrics to be within reasonable clinical utility.

    MetricAcceptance Criteria (Implicit)Reported Device Performance (Low-Sensitivity Mode)
    Positive Percent Agreement (PPA) - During Non-Rest Activities (Epilepsy Monitoring Unit Data)Clinically acceptable detection of GTCS6-21 years: 0.895 (corrected PPA: 0.791, CI: 0.619-0.925)
    >21 years: 1.000 (corrected PPA: 0.905, CI: 0.891-0.917)
    False Alarm Rate (FAR) per 24 hours - During Non-Rest Activities (Epilepsy Monitoring Unit Data)Clinically acceptable false alarm rate6-21 years: Overall FAR: 0.70 (CI: 0.41-1.06), Mean FAR: 0.91 (CI: 0.44-1.57)
    >21 years: Overall FAR: 0.28 (CI: 0.15-0.46), Mean FAR: 0.33 (CI: 0.17-0.53)
    Positive Percent Agreement (PPA) - During Non-Rest Activities (Real-World Data)Clinically acceptable detection of GTCS6-21 years: 0.87 (corrected PPA: 0.86, CI: 0.78-0.92)
    >21 years: 0.8 (corrected PPA: 0.77, CI: 0.64-0.87)
    False Alarm Rate (FAR) per 24 hours - During Non-Rest Activities (Real-World Data)Clinically acceptable false alarm rate6-21 years: Overall FAR: 0.34 (CI: 0.23-0.50), Mean FAR: 0.35 (CI: 0.28-0.45)
    >21 years: Overall FAR: 0.25 (CI: 0.22-0.30), Mean FAR: 0.29 (CI: 0.26-0.33)

    2. Sample Size for the Test Set and Data Provenance:

    • Epilepsy Monitoring Unit (EMU) Data (Retrospective Analysis):

      • Patients for PPA: 12 patients (6-21 years old) and 12 patients (>21 years old).
      • GTCS events for PPA: 19 GTCS events (6-21 years old) and 17 GTCS events (>21 years old).
      • Patients for FAR: 80 patients (6-21 years old) and 61 patients (>21 years old).
      • Data Provenance: Retrospective analysis of previously collected clinical data from patients observed in Epilepsy Monitoring Units. The document mentions data from "epilepsy patients experiencing GTCSs in hospital Epilepsy Monitoring Units" for the validation of the algorithm (EpiAlgo 2.1).
    • Real-World Data (Longitudinal Analysis) for Low-Sensitivity Mode:

      • Patients for PPA/FAR: 601 patients (6-21 years old) and 843 patients (>21 years old).
      • GTCS events for PPA: 1157 GTCS events (6-21 years old) and 3625 GTCS events (>21 years old).
      • Observation days for FAR: 37594.2 days (6-21 years old) and 56389.1 days (>21 years old).
      • Data Provenance: Longitudinal analysis of real-world data, based on sensor data captured using the Embrace2 wearable device. This suggests the data was collected prospectively in a real-world setting, but its analysis for this specific submission was retrospective.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications:

    • For the initial validation of EpiAlgo 2.1 (which supports the predicate device and is used in the subject device), the ground truth was "designed by a group of epileptologists at a top level 4 epilepsy center." The exact number of epileptologists and their specific years of experience are not provided. The method mentioned is "gold-standard video-Electroencephalogram (EEG) methodology."
    • For the retrospective analyses presented, "adjudicated tonic-clonic seizure data" was used, implying expert review to establish the ground truth of GTCS events. The number and qualifications of the experts performing this adjudication for the analyses presented in Tables 1-4 are not explicitly stated.

    4. Adjudication Method for the Test Set:

    • The document implies clinical adjudication was performed to establish "adjudicated tonic-clonic seizure data" and the "gold-standard video-Electroencephalogram (EEG) methodology." However, it does not specify a particular adjudication method such as 2+1 or 3+1 for the test set data used in these retrospective analyses. It only mentions that the data was "adjudicated."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was done.
    • The document describes a standalone algorithm performance without human assistance for seizure detection.

    6. Standalone (Algorithm Only) Performance:

    • Yes, a standalone performance evaluation of the algorithm (EpiAlgo ver 2.1) was conducted. The PPA and FAR metrics presented (Tables 1-4) reflect the performance of the algorithm without human-in-the-loop assistance for seizure detection and alerting.

    7. Type of Ground Truth Used:

    • Expert Consensus / Clinical Diagnosis (Video-EEG): For the initial validation of EpiAlgo 2.1, the ground truth was established using "gold-standard video-Electroencephalogram (EEG) methodology designed by a group of epileptologists." This indicates a high standard of clinical diagnosis and expert consensus.
    • Adjudicated Data: For the retrospective analyses of EMU and real-world data, "adjudicated tonic-clonic seizure data" were used, implying expert review and decision-making on seizure events.

    8. Sample Size for the Training Set:

    • The document does not explicitly state the sample size for the training set of EpiAlgo ver 2.1. It mentions that EpiAlgo 2.1 was validated using data from epilepsy patients in EMUs, but this typically refers to validation/test sets, not specifically the training data.

    9. How the Ground Truth for the Training Set Was Established:

    • The method for establishing the ground truth for the training set is not detailed in this document. It only states that the EpiAlgo "has been validated through testing, using the gold-standard video-Electroencephalogram (EEG) methodology designed by a group of epileptologists at a top level 4 epilepsy center, from epilepsy patients experiencing GTCSs in hospital Epilepsy Monitoring Units." This description primarily refers to the validation data, not the data used for initial training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232915

    Validate with FDA (Live)

    Device Name
    EpiMonitor
    Manufacturer
    Date Cleared
    2024-02-15

    (149 days)

    Product Code
    Regulation Number
    882.1580
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EpiMonitor is a prescription only medical device system composed of a wearable device "EmbracePlus" and paired mobile software application "EpiMonitor" intended as an adjunct to seizure monitoring in adults and children aged 6 and up in a home environment or healthcare facilities. The device is worn on the wrist and senses Electrodermal Activity (EDA) and motion data to detect patterns that may be associated with either primary or secondary generalized tonic clonic clonic clonic seizures in patients with epilepsy or at risk of having epilepsy. When a seizure event is detected, the wearable device component of EpiMonitor sends a command to a paired mobile device where the EpiMonitor App is programmed to initiate an alert to a designated caregiver. The EpiMonitor app incorporates additional detection sensitivity modes, "high" for use during periods of rest or sleeping or "low" for use during periods of low-intensity activity, in order to reduce false alarm incidents.

    EpiMonitor records, stores and transmits accelerometer. EDA, peripheral skin temperature and activity data for subsequent retrospective review by a trained healthcare professional via a cloud-based software.

    Device Description

    The EpiMonitor system consists of a wearable device and mobile application:

    • A wearable medical device called EmbracePlus,
    • A mobile application running on smartphones called "EpiMonitor"

    The EmbracePlus is worn on the user's wrist and continuously collects raw data via specific sensors, these data are continuously analyzed by an on-board algorithm (EpiAlgo 2.1), which assesses the physiological data and determines if the user may be undergoing a generalized tonic-clonic seizure (GTCS). The EpiAlgo has been validated through testing, using the gold-standard video-Electroencephalogram (EEG) methodology designed by a group of epileptologists at a top level 4 epilepsy center, from epilepsy patients experiencing GTCSs in hospital Epilepsy Monitoring Units. When a likely GTCS is detected, EmbracePlus sends, via Bluetooth Low Energy, a message to the EpiMonitor app. The EpiMonitor app communicates to the Empatica Cloud which initiates, through the external provider a voice call and SMS text message is sent to summon the attention of userdesignated caregiver(s).

    In addition to initiating alerts, the EpiMonitor app also continuously receives all the raw sensor data collected by the EmbracePlus. These data are analyzed by one of the EpiMonitor app software modules, EmpaDSP (paragraph 2.3.2), which computes the additional physiological parameters, such as activity during sleep and peripheral skin temperature.

    The EpiMonitor App is also responsible for transmitting, over a cellular data plan or Wi-Fi connection the sensors' raw data, device information, and computed physiological parameters to the Empatica Cloud. On the Empatica Cloud, these data are stored, and made available to healthcare providers via a specific cloud-based software called Care Monitoring Portal.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the EpiMonitor device, derived from the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the EpiMonitor device's Low-Sensitivity mode were evaluated based on Positive Percent Agreement (PPA) for seizure detection and False Alarm Rate (FAR) for both Epilepsy Monitoring Unit (EMU) data and real-world data.

    Metric (Low-Sensitivity Mode)Acceptance Criteria (Implicit from "Acceptable")Reported Device Performance (EMU Data)Reported Device Performance (Real-World Data)
    Positive Percent Agreement (PPA)Acceptable seizure detection accuracyAge 6-21: 0.895 (corrected 0.791, CI: 0.619-0.925) Age >21: 1.000 (corrected 0.905, CI: 0.891-0.917)Age 6-21: 0.87 (corrected 0.86, CI: 0.78-0.92) Age >21: 0.80 (corrected 0.77, CI: 0.64-0.87)
    False Alarm Rate (FAR) per 24 hoursReduced rate of false alertsAge 6-21: 0.70 (Overall), 0.91 (Mean) Age >21: 0.28 (Overall), 0.33 (Mean)Age 6-21: 0.34 (Overall), 0.35 (Mean) Age >21: 0.25 (Overall), 0.29 (Mean)

    Note: The document explicitly states "Analysis of performance for the Low-Sensitivity alerting mode in the EpiMonitor system demonstrated acceptable seizure detection accuracy and a reduced rate of false alerts." This implies that the reported performance met the sponsor's internal acceptance criteria for these metrics. Specific numerical thresholds for "acceptable" are not explicitly stated within the provided text.

    2. Sample Sizes Used for the Test Set and Data Provenance

    For Epilepsy Monitoring Unit (EMU) Data (Retrospective Analysis):

    • Seizure Detection (PPA):
      • Patients: 24 (12 for age 6-21, 12 for age >21)
      • GTCS events: 36 (19 for age 6-21, 17 for age >21)
    • False Alarm Rate (FAR):
      • Patients: 141 (80 for age 6-21, 61 for age >21)
      • Days of monitoring: 241.62 (88.94 for age 6-21, 152.68 for age >21)
    • Data Provenance: The data was collected from patients observed in Epilepsy Monitoring Units. The exact geographic origin (country) is not specified, but the data was from "a top level 4 epilepsy center" (mentioned in device description for original EpiAlgo validation). This was a retrospective analysis of previously collected clinical data.

    For Real-World Data (Longitudinal Analysis) - based on Embrace2 wearable device:

    • Seizure Detection (PPA):
      • Patients: 1444 (601 for age 6-21, 843 for age >21)
      • GTCS events: 4782 (1157 for age 6-21, 3625 for age >21)
    • False Alarm Rate (FAR):
      • Patients: 1444 (601 for age 6-21, 843 for age >21)
      • Days of monitoring: 93983.3 (37594.2 for age 6-21, 56389.1 for age >21)
    • Data Provenance: "real-world data" captured using the Embrace2 wearable device, likely from home settings. The exact geographic origin (country) is not specified. This was a retrospective longitudinal analysis of real-world data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document states that the EpiAlgo was validated "using the gold-standard video-Electroencephalogram (EEG) methodology designed by a group of epileptologists at a top level 4 epilepsy center". It also refers to "adjudicated tonic-clonic seizure data" for the EMU data. This implies that epileptologists were involved in establishing the ground truth.

    • Number of experts: Not explicitly stated, but referred to as "a group of epileptologists."
    • Qualifications of experts: "epileptologists at a top level 4 epilepsy center." No specific experience (e.g., 10 years of experience) is detailed.

    4. Adjudication Method for the Test Set

    The document mentions "adjudicated tonic-clonic seizure data" for the EMU study. However, the specific adjudication method (e.g., 2+1, 3+1) is not explicitly described in the provided text.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No.
    • The study focuses on the performance of the algorithm itself (standalone and with different sensitivity modes), not on how human readers improve with or without AI assistance.
    • The effect size of human readers improving with AI vs. without AI assistance is not applicable as this type of study was not performed.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone study was done. The entire performance analysis for PPA and FAR presented in Tables 1-4 reflects the algorithm's performance (EpiAlgo ver 2.1) using the Low-Sensitivity mode, without human intervention in the detection process. The device detects an event, and the app initiates an alert; there's no mention of a human-in-the-loop directly influencing the detection sensitivity.

    7. Type of Ground Truth Used

    • Expert Consensus / Gold Standard (Video-EEG): The ground truth for seizure events was primarily established using gold-standard video-Electroencephalogram (EEG) methodology and "adjudicated tonic-clonic seizure data." This indicates expert consensus based on clinical and physiological evidence.

    8. Sample Size for the Training Set

    The provided text does not specify the sample size used for the training set of the EpiAlgo. It only describes the validation phases for the Low-Sensitivity mode.

    9. How the Ground Truth for the Training Set Was Established

    The provided text states: "The EpiAlgo has been validated through testing, using the gold-standard video-Electroencephalogram (EEG) methodology designed by a group of epileptologists at a top level 4 epilepsy center, from epilepsy patients experiencing GTCSs in hospital Epilepsy Monitoring Units."

    This implies that the training data's ground truth was established by epileptologists using video-EEG data from patients with generalized tonic-clonic seizures (GTCSs) in hospital Epilepsy Monitoring Units. This is consistent with clinical gold standards for seizure identification.