Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200557

    Validate with FDA (Live)

    Date Cleared
    2020-09-28

    (209 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use as a dispenser, measuring device, and fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by users ranging from clinicians to laypersons (under the supervision of a clinician) to in all age groups.

    Device Description

    The device is an enteral feeding syringe for single use. The devices are piston-style syringes consisting of a syringe barrel, syringe plunger, syringe gasket, and syringe cap. The device incorporates a female ENFit connector designed to be compatible only with enteral access devices or accessories. It has ENFit compliant or compatible male connectors to form a dedicated system that prevents wrong-route administration of fluids.

    The device is a disposable syringe made of the following components:

    1. Plunger: The Plunger is a piston having a longitudinal shaft that is mechanically received into the proximal syringe barrel aperture. It is made of polypropylene and uncoated. The proximal end of the plunger is flat to allow the device user to apply uniform pressure to the plunger shaft, thereby initiating the plunger into the syringe barrel.
    2. Barrel: The Barrel is a single, clear molded plastic hollow barrel made of polypropylene. This is the identical material used to manufacture the barrel of the predicate device. The distal end of the barrel possesses an integrally molded syringe tip. The barrel is marked on the exterior with non-toxic, biocompatible ink to create gradient volume graduation markings. The barrel is translucent to allow visualization of fluid contents and volume. The barrel is designed at an appropriate length to accommodate the nominal volume of each syringe size as labeled.
    3. Gasket: The gasket is attached to the distal end of the syringe plunger in an o-ring orientation. It is attached to the plunger via a mechanical interference fit. The purpose of the gasket is to engage the syringe barrel to form a low-pressure seal, which is impervious to fluids.
    4. Tip cap: The tip cap is an ENFit compatible closure cap for the syringe. The syringe tip cap is designed to be compatible with the ENFit female syringe tip of the syringe. The syringe tip cap is made of polypropylene and purple colorant.
    AI/ML Overview

    This document primarily discusses the substantial equivalence of a medical device (Enteral Feeding Syringes with ENFit connector) to a predicate device, focusing on non-clinical performance testing, biocompatibility, and sterility. It does not present a study proving a device meets acceptance criteria in the context of an AI/ML medical device, clinical efficacy, or human reader performance. The document explicitly states "No clinical study is included in this submission."

    Therefore, based on the provided text, I cannot answer the questions related to clinical studies, human readers, AI assistance, ground truth establishment for AI/ML models, or sample sizes for training sets.

    However, I can extract the acceptance criteria and performance data for the non-clinical aspects of the device.

    Here's the information gleaned from the document regarding the acceptance criteria and device performance for the physical device itself:


    1. A table of acceptance criteria and the reported device performance

    The document defines acceptance criteria by referencing compliance with specific international standards and then states that the device "met all requirements" or that "All of the pre-determined acceptance criteria were met." Specific numerical performance data is provided for sterility residuals and bioburden.

    CategoryAcceptance CriteriaReported Device Performance
    PerformanceCompliance with: - ISO 7886-1: Sterile hypodermic syringes for single use - Part 1: Syringes for manual use - ISO 80369-3: Small-bore connectors for liquids and gases in health care applications Part 3: Connectors for enteral applications - ISO 80369-20: Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods"For the finished device, it meets the defined performance requirements through bench testing." (Implies compliance with the listed ISO standards).
    BiocompatibilityCompliance with ISO 10993-1 for indirectly contacting patients, contact duration ≤24h. Specific tests: - In Vitro Cytotoxicity (ISO10993-5: 2009) - Skin Sensitization (ISO10993-10: 2010) - Intracutaneous Reactivity Test (ISO10993-10: 2010)"The new device complies with the biocompatibility requirement defined in ISO10993-1. The verification test shows that the new devices comply with the biocompatibility requirement defined in ISO10993-1, the same as the predicate device. All of the pre-determined acceptance criteria were met."
    SterilitySterilization conducted per ISO 11135-1, ISO 11737-1, ISO 11737-2, ISO 10993-7, ANSI/AAMI ST72. Sterility Assurance Level (SAL): 10-6. Residuals: - EO Residual: ≤ 4.0 mg/device - ECH Residuals: ≤ 9.0 mg/device Sterility Test Result: 0 Positives Pyrogenicity Test: Temperature rise must meet standard Endotoxin Test (Gel-Clot Limit Test): < 0.25 EU/mlSAL: 10-6 (Met) 10 ml Syringe: - EO Residual: 2.6X10-3 mg/device - ECH Residuals: < 3.0X10-3mg/device 60 ml Control Ring Syringe: - EO Residual: 0.016 mg/device - ECH Residuals: 0.028 mg/device (All residuals well within limits.) Bioburden: - 10ml: 3.4 cfu/device - 60 ml control ring: 3.4 cfu/device Result of Sterility test: 0 Positives (Met) Result of Pyrogenicity test: - 10ml: 0.21°C, 0.00°C, 0.03°C - 60 ml control ring: 0.09℃, 0.12℃, 0.07℃ Result of Endotoxin test: Less than 0.25 EU/ml (Met) "All of the pre-determined acceptance criteria were met."
    Packaging & Shelf LifeCompliance with AAMI/ANSI/ISO 11137-1, AAMI/ANSI/ISO 11737-2, AAMI/ANSI/ISO 11607-1, ASTM F1929-98, ASTM F1980-07, ASTM D3078-02, ASTM F88/F88M-09. Support for a 5-year shelf life."The test result supports the five year shelf life claim for the subject device from the sterilization date. All of the pre-determined acceptance criteria were met."

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the performance, biocompatibility, sterility, or packaging tests. It describes "bench testing" and "verification tests" which are typically done on a defined number of units, but the specific count is not provided. The data provenance is not specified in terms of country of origin, but it is implied to be from the manufacturer's testing facilities (Jiangsu Shenli Medical Production Co., Ltd in China). These tests are typically prospective for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the document describes non-clinical performance and safety testing of a physical device, not an AI/ML system requiring expert interpretation of medical images or data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for tests of a physical medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The document explicitly states: "No clinical study is included in this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the physical device testing:

    • Performance: Ground truth is defined by the technical specifications and requirements outlined in the referenced ISO standards (e.g., specific force required, leak integrity, volume accuracy).
    • Biocompatibility: Ground truth is adherence to the toxicological and biological safety requirements in ISO 10993 series.
    • Sterility: Ground truth is 10^-6 SAL, absence of growth in sterility tests, compliance with residual limits, and pyrogenicity/endotoxin limits.
    • Packaging & Shelf Life: Ground truth is maintaining integrity and sterility over the claimed shelf life as per referenced ASTM and ISO packing standards.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1