510(k) Data Aggregation

    K Number
    K223683

    Validate with FDA (Live)

    Device Name
    EnteraLoc Flow
    Manufacturer
    Date Cleared
    2023-07-10

    (214 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vonco Products EnteraLoc Flow spouted pouch with ENFit® Connector is indicated for use as a dispenser of enteral nutrition by way of direct connection to a feeding tube or extension set. It is intended to deliver nutrition into the gastrointestinal system of a patient. The pouch, once filled, is intended to be used in clinical or home care settings by users ranging from laypersons to clinicians, in all age groups.

    Device Description

    The Vonco Products EnteraLoc Flow spouted pouch with ENFit® Connector consists of a foil pouch with an integrated female ENFit® connector at one end. The EnteraLoc Flow pouch is provided with a tamper evident cap. The pouch is designed to be pre-filled with liquid nutrition formula for tube feeding. EnteraLoc Flow is a single use device.

    EnteraLoc Flow spouted pouch with ENFit® Connector is designed as a single use, over-the-counter device. The pouches will be available in a range of sizes, with a range of nutritional values (based upon the formula which the pouch is filled). Based upon the required nutritional intake requirements, the user will select the proper size pouch and quantity for feeding.

    AI/ML Overview

    The provided document is a 510(k) premarket notification decision letter from the FDA regarding the "EnteraLoc Flow" device, which is an enteral nutrition dispensing pouch. The document details the device's indications for use, comparison to predicate devices, and a list of performance tests conducted.

    However, the document does not contain the level of detail requested for the acceptance criteria and study that proves the device meets those criteria, particularly within the context of an AI/ML-based medical device. Specifically, it lacks information on:

    1. A table of acceptance criteria and reported device performance with specific quantitative metrics. The document lists types of tests (e.g., Flow Rate Analysis, Biocompatibility Testing, Performance Testing per ISO standards) but does not provide the pass/fail criteria or the actual results obtained.
    2. Sample size used for the test set and data provenance.
    3. Number of experts used to establish ground truth and their qualifications.
    4. Adjudication method.
    5. Whether a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, or the effect size of human readers' improvement with AI vs. without AI assistance. This is explicitly an AI/ML-specific question, and the device described is a physical medical device, not an AI/ML diagnostic tool.
    6. Whether a standalone performance study was done (algorithm only without human-in-the-loop performance). Again, this is an AI/ML-specific question.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices through technical characteristics, intended use, and various physical and biological performance tests relevant to a non-AI/ML medical device.

    Therefore, it is impossible to extract the requested information from the provided text as it pertains to an AI/ML product's acceptance criteria and study details. The device in question, the "EnteraLoc Flow" spouted pouch with ENFit® Connector, is a physical device for dispensing enteral nutrition, not an AI/ML diagnostic or assistive tool. The questions are tailored for evaluating an AI/ML-based medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210971

    Validate with FDA (Live)

    Device Name
    EnteraLoc Flow
    Manufacturer
    Date Cleared
    2021-08-18

    (140 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EnteraLoc Flow spouted pouch with ENFit® Connector is indicated for use as a dispenser of enteral nutrition by way of direct connection to a feeding tube or extension set. It is intended to deliver nutrition into the gastrointestinal system of a patient. The pouch, once filled, is intended to be used in clinical or home care settings by users ranging from laypersons to clinicians, in all age groups.

    Device Description

    The Vonco Products EnteraLoc Flow spouted pouch with ENFit Connector consists of a foil pouch with an integrated female ENFit connector at one end. The EnteraLoc Flow pouch is provided with a tamper evident cap. The pouch is designed to be pre-filled with liquid nutrition formula for tube feeding.

    The EnteraLoc Flow spouted pouch with ENFit Connector will be available in sizes ranging from 30mL - 1500mL dependent upon market and customer need.

    AI/ML Overview

    The provided text is a 510(k) summary for the Vonco Products EnteraLoc Flow device, which is an enteral nutrition dispensing pouch. This document details the device's technical specifications and the testing performed to demonstrate its substantial equivalence to predicate devices, but it does not describe an AI/ML medical device or a study involving human readers and AI assistance.

    Therefore, I cannot provide the detailed information requested in the prompt regarding:

    • A table of acceptance criteria and reported device performance related to diagnostic accuracy or clinical outcomes of an AI/ML model.
    • Sample sizes for test sets, data provenance, ground truth establishment for AI/ML models.
    • Number of experts, qualifications, and adjudication methods for AI/ML ground truth.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies or standalone algorithm performance for AI/ML.

    The performance data listed in Section 5.8 of the document are for the physical device (a pouch with a connector), not an AI/ML algorithm. These tests include:

    • Flow Rate Analysis
    • Biocompatibility Testing (Cytotoxicity, Irritation/Intracutaneous Reactivity, Sensitization)
    • Performance Testing per ISO 80369-3:2016 (Dimensioning, Falling Drop Positive Pressure Liquid Leakage, Stress Cracking, Resistance to Separation from Axial Load, Resistance to Separation from Unscrewing, Resistance to Overriding, Disconnection by Unscrewing)
    • ISO 80369-1:2018 Misconnection Analysis
    • Spouted Pouch Functional Testing
    • Retort Suitability

    The document states, "The results of the performance testing listed above show that the subject device meets its specifications." However, it does not provide the specific acceptance criteria or the numerical performance results for these tests. It only generally states that the device met its specifications.

    In summary, this document describes the regulatory approval of a physical medical device (an enteral nutrition pouch), not an AI/ML medical device. Therefore, the information needed to answer your prompt about AI/ML device acceptance criteria and study design is not present in this text.