Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152598

    Validate with FDA (Live)

    Device Name
    Endo Keeper
    Manufacturer
    Date Cleared
    2016-03-17

    (188 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endo Keeper is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery.

    This device is single use and sterilized.

    Device Description

    The Endo Keeper is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery. This device is single use and sterilized.

    The Endo Keeper is laparoscopic instrument port which retracts a small abdominal incision to allow multiple laparoscopic instruments to pass through to the abdomen.

    The Urethane Forming Sheet (components of the Endo Keeper) function is to cover the incision site of abdominal.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Endo Keeper." This document focuses on demonstrating that the Endo Keeper is substantially equivalent to a legally marketed predicate device, the "Glove Port" (K141715). The information primarily outlines the device's intended use, technological characteristics, and performance testing conducted to support this claim.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical "acceptance criteria" for performance metrics in a table. It states that "Bench testing is performed to demonstrate the functionality and mechanical safety as following items." The reported performance is summarized as demonstrating "functionality and mechanical safety" and being "same as functional and effective as the legally marketed predicate device."

    Performance MetricAcceptance Criteria (Not explicitly stated as numerical targets)Reported Device Performance
    Leak rateFunctionality and Mechanical Safety Comparable to PredicateDemonstrated functionality and mechanical safety; "same as functional and effective as the legally marketed predicate device."
    Tensile strengthFunctionality and Mechanical Safety Comparable to PredicateDemonstrated functionality and mechanical safety; "same as functional and effective as the legally marketed predicate device."
    Insufflation flow rateFunctionality and Mechanical Safety Comparable to PredicateDemonstrated functionality and mechanical safety; "same as functional and effective as the legally marketed predicate device."
    Insertion-withdrawal forces of instrumentFunctionality and Mechanical Safety Comparable to PredicateDemonstrated functionality and mechanical safety; "same as functional and effective as the legally marketed predicate device."
    Minimum size of skin incisionFunctionality and Mechanical Safety Comparable to PredicateDemonstrated functionality and mechanical safety; "same as functional and effective as the legally marketed predicate device."
    Biocompatibility (in-vivo)No irritations and necrosis of the skin and fascia (ISO 10993-2 and ISO 10993-1 compliance implied)"Verified of feasibility and functionality." "No irritations and necrosis of the skin and fascia."

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Bench testing" and "Animal test" as performance data.

    • Bench Testing: The sample sizes for each specific bench test (leak rate, tensile strength, etc.) are not specified in the provided text.
    • Animal Test: The sample size (number of animals) for the in-vivo test is not specified.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective. Given that this is a 510(k) submission for a Korean company (NELIS), it's highly probable the testing was conducted in Korea or at an accredited facility. It is prospective testing, as it was performed to support device approval.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The performance data is based on bench tests and an animal study, not human interpretation requiring experts for ground truth establishment in the traditional sense of diagnostic AI.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies described are bench and animal tests, not studies involving human interpretation that would require an adjudication method.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not performed. The studies described are engineering bench tests and an animal biocompatibility/functionality study. The document explicitly states the comparison is between the "Endo Keeper" and the "Glove Port (K141715)" based on "materials, structure and component" and "function and effectiveness" through bench and animal testing, not human reader performance.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    This concept is not applicable to the "Endo Keeper" device, which is a physical surgical access port, not an algorithm or AI system. The performance studies are for the physical device itself.

    7. Type of Ground Truth Used

    • Bench Testing: The "ground truth" for the bench tests would be the established engineering and ISO standards for properties like leak rate, tensile strength, insertion/withdrawal forces, etc. The comparison is against the predicate device's performance.
    • Animal Test: The "ground truth" for the in-vivo test regarding feasibility, functionality, irritations, and necrosis would be direct observation and pathological examination of the animal tissues, assessed against established biological response criteria (e.g., ISO 10993).

    8. Sample Size for the Training Set

    This information is not applicable as the "Endo Keeper" is a medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as the "Endo Keeper" is a medical device, not an AI model that requires a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1