Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160210

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-13

    (320 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EnSite Precision™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.

    The EnSite™ Precision System interfaces to either the MediGuide™ Technology system or the EnSite Precision Module to combine and display magnetic processed patient positioning and navigation mapping information.

    When used with the EnSite™ Array™ Catheter, the EnSite Precision™ Cardiac Mapping System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

    OR

    When used with an EnSite Precision™ Surface Electrode Kit, the EnSite Precision™ Cardiac Mapping System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.

    Device Description

    The EnSite Precision™ Cardiac Mapping System is a catheter navigation and mapping system capable of displaying the three-dimensional (3D) position of conventional electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as dynamic 3-D isopotential maps of the cardiac chamber. The contoured surfaces of these three-dimensional maps are based on the anatomy of the patient's own cardiac chamber.

    The EnSite Precision™ Cardiac Mapping System with version 2.0 software is a new module that is designed to allow the system to interface to the EnSite Precision™ Module, Sensor Enabled, which is an alternative magnetic field-based medical positioning hardware kit for magnetic field scaling functionality.

    AI/ML Overview

    This document does not contain the detailed acceptance criteria and study results in the format requested. While it outlines the device, its intended use, and that verification/validation activities were performed, it does not provide specific metrics or performance tables.

    Here's what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    • Missing. The document states "Design verification/validation activities were performed with their respective acceptance criteria to ensure that the modifications do not affect the safety or effectiveness of the device. All testing performed met the established performance specifications." However, it does not specify what those acceptance criteria were or what the measured performance results were against those criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Partially Missing. The document mentions "Dynamic Wetlab Bench Comparative and Preclinical Animal testing."
      • Sample Size: Not specified.
      • Data Provenance: "Wetlab Bench" and "Preclinical Animal testing" suggest laboratory/animal studies, which are prospective. Country of origin not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable / Missing. This device is a cardiac mapping system for electrophysiology studies. The "ground truth" in this context would likely be precise catheter location measurements, which are typically established through highly accurate measurement systems, not expert consensus in the same way an image diagnosis would be. The document describes "magnetic field-based catheter localization" and "impedance locations" as mechanisms for determining position, suggesting instrumental ground truth rather than expert interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable / Missing. As the ground truth is instrumental (catheter position) rather than interpretive (e.g., diagnosing an image), adjudication methods by human readers are not relevant here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This document describes a K160210 submission for a cardiac mapping system, focusing on its technical performance and equivalence to predicate devices, not on AI assistance for human readers in diagnostic interpretation. The testing involved comparisons of model rendering and catheter electrode localization between the subject and predicate devices, and workflow performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The testing described ("Electrical, Mechanical, Accuracy and Repeatability, Packaging, shelf life, Biocompatibility, Preclinical Animal Studies") focuses on the device's technical performance. The "accuracy testing for use with the magnetic field scaling feature" and comparisons of "model rendering and location of catheter electrodes" are examples of standalone performance evaluation of the system's ability to accurately map and localize.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth would be based on highly accurate physical measurements for catheter positions in the wetlab/animal studies. The document states "magnetic field-based catheter localization used for field scaling during model creation involves sensing of the magnetic fields generated through small coil sensors embedded inside the enabled device." This implies a physical, instrumental ground truth for position validation.

    8. The sample size for the training set:

    • Not Applicable / Missing. This document describes a device approval, not a machine learning algorithm that requires a "training set" in the conventional sense. The "training" here refers to the device's development and validation, which uses various engineering and animal studies.

    9. How the ground truth for the training set was established:

    • Not Applicable / Missing. See point 8. The ground truth for validation testing (such as catheter position) would be established through precise physical measurement techniques in the wetlab or animal models.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1